Skip to content

HUDSON RCI mBrace, ET Tube Holder with Bite Block and 4… recallZ-0698-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Nov 18, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Medline Industries issued an URGENT MEDICAL DEVICE RECALL to its consignees of 11/18/2024 via USPS and email. The notice explained the issue, risk to user, and requested the destruction of the affected product. For questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-0698-2025
Status
Open, Classified
Initiated by the company
Nov 18, 2024
Posted by the FDA
Dec 18, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
768 units
Product code
CBH
Distribution
US and Canada

Product description

HUDSON RCI mBrace, ET Tube Holder with Bite Block and 4 Point Head Strap, REF DYNJMBRC4B; tracheal tube fixation device

Code information (models, lots, serial numbers)

UDI/DI 20193489005339 (case), 10193489005332 (each); All Lots

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95821).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page