Oec 9900 Ref recallZ-0705-2025 · GE OEC Medical Systems, Inc · initiated Nov 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE OEC Medical Systems, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
FDA root cause category: Process control
What the company is doing
On 11/21/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that GE HealthCare has become aware that recently manufactured OEC Elite and OEC 3D mobile C-arms, as well as recently replaced X-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/OEC 9800 mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability. Customers are instructed to: 1.Continue to use their OEC Elite, OEC 3D, OEC 9900 and/or OEC9800 mobile C-arm. 2.If oil is found leaking from the X-ray tube on their OEC Elite, OEC 3D, OEC9900 and/or OEC 9800 mobile C-arm, contact the local GE HealthCareService representative. 3.Ensure all potential users in their facility are made aware of this safety notification and the recommended actions. 4.Please complete and return the attached acknowledgement form [email protected]. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to plan and schedule for the inspection and correction if needed. for any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-0705-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 21, 2024
- Posted by the FDA
- Dec 18, 2024
- Recalling firm
- GE OEC Medical Systems, Inc
- Quantity
- 78 systems
- Product code
- OXO
- FDA submissions
- K120613, K122234, K073543, K082781
- Distribution
- U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
Product description
OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
Product Name: OEC 9900 GTIN code: 00840682114301 Lot/Serial Numbers / UDI codes: E2-0345 Not Applicable-MFG date prior to UDI compliance requirement E2-7031-MH Not Applicable-MFG date prior to UDI compliance requirement E2-1862 Not Applicable-MFG date prior to UDI compliance requirement ES-0942 Not Applicable-MFG date prior to UDI compliance requirement E9-1101 Not Applicable-MFG date prior to UDI compliance requirement E9-2173 Not Applicable-MFG date prior to UDI compliance requirement E9-2706 Not Applicable-MFG date prior to UDI compliance requirement E2-5146 Not Applicable-MFG date prior to UDI compliance requirement E2-1777 Not Applicable-MFG date prior to UDI compliance requirement E2XXXX05533 (01)00840682114301(11)160600(21)E2XXXX05533 ES-3271 Not Applicable-MFG date prior to UDI compliance requirement E2XXXX05700 (01)00840682114301(11)160900(21)E2XXXX05700 E2XXXX06210 (01)00840682114301(11)170900(21)E2XXXX06210 E2MHXX07986 (01)00840682114301(11)170500(21)E2MHXX07986 E9-2586 Not Applicable-MFG date prior to UDI compliance requirement ES-1901 Not Applicable-MFG date prior to UDI compliance requirement E9-2999 Not Applicable-MFG date prior to UDI compliance requirement ESXXXX04466 (01)00840682114301(11)190100(21)ESXXXX04466 ES-2359 Not Applicable-MFG date prior to UDI compliance requirement E9-1824 Not Applicable-MFG date prior to UDI compliance requirement E2-4112 Not Applicable-MFG date prior to UDI compliance requirement E2-0633 Not Applicable-MFG date prior to UDI compliance requirement E9-0156 Not Applicable-MFG date prior to UDI compliance requirement ES-3537 Not Applicable-MFG date prior to UDI compliance requirement ES-0348 Not Applicable-MFG date prior to UDI compliance requirement ES-1269 Not Applicable-MFG date prior to UDI compliance requirement E9-2634 Not Applicable-MFG date prior to UDI compliance requirement E9-0763 Not Applicable-MFG date prior to UDI compliance requirement E2MHXX07958 (01)00840682114301(11)160900(21)E2MHXX07958 E2-4224 Not Applicable-MFG date prior to UDI compliance requirement E2XXXX05753 (01)00840682114301(11)161000(21)E2XXXX05753 ES-0186 Not Applicable-MFG date prior to UDI compliance requirement ES-1648 Not Applicable-MFG date prior to UDI compliance requirement ESXXXX04227 (01)00840682114301(11)170800(21)ESXXXX04227 ES-2800 Not Applicable-MFG date prior to UDI compliance requirement E2-2074 Not Applicable-MFG date prior to UDI compliance requirement E9XXXX03158 (01)00840682114301(11)170100(21)E9XXXX03158 ES-1390 Not Applicable-MFG date prior to UDI compliance requirement ESXXXX03869 (01)00840682114301(11)160800(21)ESXXXX03869 E9XXXX03157 (01)00840682114301(11)170100(21)E9XXXX03157 E2XXXX06067 (01)00840682114301(11)170400(21)E2XXXX06067 ES-1476 Not Applicable-MFG date prior to UDI compliance requirement ES-0317 Not Applicable-MFG date prior to UDI compliance requirement ES-1011 Not Applicable-MFG date prior to UDI compliance requirement E2-3459 Not Applicable-MFG date prior to UDI compliance requirement ES-0144 Not Applicable-MFG date prior to UDI compliance requirement ES-0336 Not Applicable-MFG date prior to UDI compliance requirement E9-2976 Not Applicable-MFG date prior to UDI compliance requirement E2-7081-MH Not Applicable-MFG date prior to UDI compliance requirement E2-4029 Not Applicable-MFG date prior to UDI compliance requirement E2-4133 Not Applicable-MFG date prior to UDI compliance requirement E2-5039 Not Applicable-MFG date prior to UDI compliance requirement E2-7161-MH Not Applicable-MFG date prior to UDI compliance requirement E2-4550 Not Applicable-MFG date prior to UDI compliance requirement E9-1488 Not Applicable-MFG date prior to UDI compliance requirement E2-4026 Not Applicable-MFG date prior to UDI compliance requirement E2XXXX05597 (01)00840682114301(11)160600(21)E2XXXX05597 E2-4617 Not Applicable-MFG date prior to UDI compliance requirement ES-0515 Not Applicable-MFG date prior to UDI compliance requirement E2-2621 Not Applicable-MFG date prior to UDI compliance requirement E9-1434 Not Applicable-MFG date prior to UDI compliance requirement E2-2531 Not Applicable-MFG date prior to UDI compliance requirement E2-1272 Not Applicable-MFG date prior to UDI compliance requirement ES-1439 Not Applicable-MFG date prior to UDI compliance requirement E9-2364 Not Applicable-MFG date prior to UDI compliance requirement ES-2273 Not Applicable-MFG date prior to UDI compliance requirement ES-2603 Not Applicable-MFG date prior to UDI compliance requirement E2-0371 Not Applicable-MFG date prior to UDI compliance requirement E2XXXX05573 (01)00840682114301(11)160600(21)E2XXXX05573 E9-2096 Not Applicable-MFG date prior to UDI compliance requirement ES-0717 Not Applicable-MFG date prior to UDI compliance requirement E2XXXX05451 (01)00840682114301(11)160500(21)E2XXXX05451 E9XXXX03164 (01)00840682114301(11)170100(21)E9XXXX03164 E2-4265 Not Applicable-MFG date prior to UDI compliance requirement E9-2471 Not Applicable-MFG date prior to UDI compliance requirement E2XXXX06639 (01)00840682114301(11)181000(21)E2XXXX06639 E9-2720 Not Applicable-MFG date prior to UDI compliance requirement E9-2710 Not Applicable-MFG date prior to UDI compliance requirement
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- OEC 9900 ELITEOec Medical Systems, Inc. · Interventional fluoroscopic x-ray system3 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95846).
- Z-0703-2025 · Open, ClassifiedOEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physi…
- Z-0704-2025 · Open, ClassifiedOEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, …
- Z-0706-2025 · Open, ClassifiedOEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page