Skip to content

Baxter SIGMA Spectrum Infusion Pump with Master Drug… recallZ-0775-2025 · Baxter Healthcare Corporation · initiated Nov 19, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices were identified as released after repair without full testing being performed, which includes flow testing.

FDA root cause category: Process control

What the company is doing

Customers were contacted via telephone beginning 11/19/24. Baxter Healthcare is calling to notify you of a potential issue with the Spectrum infusion pump. One of your devices has been identified as released after repair without full testing being performed, which includes flow testing. Baxter is requesting the return of the impacted infusion pump to perform all tests required and confirm that the infusion pump performs as expected. As part of this process, we need to make you aware of the potential hazard associated with flow related issues. Flow issues may cause excessive therapy, free-flow condition, insufficient therapy or interruption of therapy. These may lead to serious adverse health consequences, including permanent injury and death. Send signed form to Global CORP Service Quality Field Actions [email protected]

Recall details

Recall number
Z-0775-2025
Status
Open, Classified
Initiated by the company
Nov 19, 2024
Posted by the FDA
Dec 27, 2024
Recalling firm
Baxter Healthcare Corporation
Quantity
1 units
Product code
FRN
FDA submissions
K133801
Distribution
US Nationwide distribution in the states of AL, IN.

Product description

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), Product Code 35700BAX

Code information (models, lots, serial numbers)

GTIN 00085412091570; Serial Number: 779515

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95848).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page