See DocMan recallZ-0782-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Nov 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Recall details
- Recall number
- Z-0782-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 21, 2024
- Posted by the FDA
- Dec 30, 2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 48655 units
- Product code
- OGR
- Distribution
- Worldwide distribution - US Nationwide and the countries of UAE, and Canada.
Product description
See DocMan
Code information (models, lots, serial numbers)
Show the full code information (23k characters)
1) REF CDS981185B, UDI/DI 10884389647141 (EA) 40884389647142 (CS), Lot Numbers: 19CDB151, 19DDD096, 19FDA612, 19IDA540, 19LDB150, 20BDA647, 20DDB481, 20IDA733, 20LDA459; 2) REF CDS983087G, UDI/DI 10193489237351 (EA) 40193489237352 (CS), Lot Numbers: 21LBS976; 3) REF CDS983302G, UDI/DI 10888277753181 (EA) 40888277753182 (CS), Lot Numbers: 19JDB741; 4) REF CDS983302I, UDI/DI 10193489425659 (EA) 40193489425650 (CS), Lot Numbers: 23FDA273, 23IDA781; 5) REF CDS983483G, UDI/DI 10889942946068 (EA) 40889942946069 (CS), Lot Numbers: 19CDB914, 19DDB938, 19GDB412, 19GDC427, 19JDA647, 19KDB084, 19LDC222, 20BDC250, 20EDB879, 20GDC001, 20JDA007, 20JDB542, 20LDB199; 6) REF CDS983483I, UDI/DI 10193489919585 (EA) 40193489919586 (CS), Lot Numbers: 21FDB817, 21HDA571, 21KDB608, 22BDA109; 7) REF CDS984313G, UDI/DI 10193489540437 (EA) 40193489540438 (CS), Lot Numbers: 19GBL203, 20EBC321, 20KBB606; 8) REF DYNJ0101278F, UDI/DI 10193489352351 (EA) 40193489352352 (CS), Lot Numbers: 20FDA999, 20HDA734, 20KDB838, 21ADC024, 21CDA355, 21EDA527, 21GDC529, 21IDA040, 21IDA786, 21JDB214, 21KDC443, 22ADC117, 22CDB133; 9) REF DYNJ0826316K, UDI/DI 10889942092642 (EA) 40889942092643 (CS), Lot Numbers: 19FKB533, 19EKD537; 10) REF DYNJ0826316L, UDI/DI 10193489413496 (EA) 40193489413497 (CS), Lot Numbers: 21IME029, 21IMC729, 20IKB206; 11) REF DYNJ0867411F, UDI/DI 10889942620579 (EA) 40889942620570 (CS), Lot Numbers: 19EDA018; 12) REF DYNJ10130D, UDI/DI 10193489432176 (EA) 40193489432177 (CS), Lot Numbers: 21BBC490, 21GBS164, 22BBE642; 13) REF DYNJ36728B, UDI/DI 10889942389209 (EA) 40889942389200 (CS), Lot Numbers: 19BBG860, 19DBL861, 19HBF653, 19KBY296; 14) REF DYNJ37254F, UDI/DI 10889942450121 (EA) 40889942450122 (CS), Lot Numbers: 19BBG398, 19DBH566, 19GBJ891, 19IBU877, 19LBQ548, 20BBO050; 15) REF DYNJ37254G, UDI/DI 10193489317794 (EA) 40193489317795 (CS), Lot Numbers: 20DBV689, 20EBP226, 20LBT897, 20LBU922; 16) REF DYNJ37254I, UDI/DI 10193489849677 (EA) 40193489849678 (CS), Lot Numbers: 21EBB287, 21FBS039, 21KBB916, 21WBA934, 22BBP179, 22DBN656, 22FBE094, 22IBA792, 22IBC342, 23BBD691, 23BBP725, 23CBS713, 23IBR830, 23KBB006; 17) REF DYNJ37254J, UDI/DI 10195327595272 (EA) 40195327595273 (CS), Lot Numbers: 24CBN777, 24EBV639, 24GBP625, 24HBL327; 18) REF DYNJ37383C, UDI/DI 10889942465255 (EA) 40889942465256 (CS), Lot Numbers: 23HBK851; 19) REF DYNJ37551A, UDI/DI 10888277585720 (EA) 40888277585721 (CS), Lot Numbers: 23HBO656; 20) REF DYNJ40094B, UDI/DI 10888277208711 (EA) 40888277208712 (CS), Lot Numbers: 19AKD477, 19BKD188, 19DKA822, 19FKC285, 19HKB997, 19JMB401, 19LMC591, 20BMB740, 20BMC380, 20DMA323; 21) REF DYNJ40094C, UDI/DI 10193489371369 (EA) 40193489371360 (CS), Lot Numbers: 20HMB370, 20KMC960, 20LMD514; 22) REF DYNJ40094D, UDI/DI 10193489814491 (EA) 40193489814492 (CS), Lot Numbers: 21AMD504, 21FMA749, 21HMH004, 21JMA896, 21LMB149; 23) REF DYNJ40360D, UDI/DI 10889942697175 (EA) 40889942697176 (CS), Lot Numbers: 19ABK188, 19CBR141, 19FBR439, 19HBF132, 19IBP615, 19KBQ585, 19LBG346, 19LBX594, 19PBA454, 20BBF530, 20BBL599, 20CBL193, 20FBG957, 20JBZ214, 21DBA832, 21DBB800, 21FBG597, 21FBH899, 21KBC034, 22ABP202, 22DBL101, 22EBO742, 22GBK777, 22IBS429, 22IBU440, 22JBL587, 23ABG728, 23CBC426, 23DBD304, 23FBQ109, 23GBP104, 23IBM800, 23KBV657, 24ABX199; 24) REF DYNJ40360F, UDI/DI 10195327646943 (EA) 40195327646944 (CS), Lot Numbers: 24CBD964, 24DBK537, 24EBU384, 24HBD935, 24HBU664; 25) REF DYNJ40367C, UDI/DI 10889942530847 (EA) 40889942530848 (CS), Lot Numbers: 19ABD489, 19DBC716, 19DBS940, 19FBX750, 19KBR568, 19LBG353, 20ABD302, 20GBV337, 21DBA938, 21EBD979, 21KBD411, 22ABJ472, 22CBK103; 26) REF DYNJ40546B, UDI/DI 10889942576050 (EA) 40889942576051 (CS), Lot Numbers: 19AKA421, 19CKB561, 19DKD095, 19FKC891; 27) REF DYNJ40548B, UDI/DI 10889942576067 (EA) 40889942576068 (CS), Lot Numbers: 19BKB151; 28) REF DYNJ41923B, UDI/DI 10889942380091 (EA) 40889942380092 (CS), Lot Numbers: 19ABM921; 29) REF DYNJ41923C, UDI/DI 10193489677614 (EA) 40193489677615 (CS), Lot Numbers: 19FBQ574, 19HBH201, 19SBA225, 20BBN192, 20CBK650, 20DBG283, 20LBG893, 21ABA695, 21JBE214, 21KBZ942, 21KBZ951, 22ABD668, 22JBA053, 23EBN364, 23FBM705, 23IBA234, 23IBM790, 23KBO421, 23KBV637, 24FBG873, 24GBJ293, 24GBR489,; 30) REF DYNJ43582A, UDI/DI 10889942255573 (EA) 40889942255574 (CS), Lot Numbers: 19ADD164, 19BDC840, 19EDA377, 19GDC175, 19IDC514, 20ADA566, 20CDA176, 20FDA533, 20HDA715, 20IDA515, 21EDA529, 21HDB003, 21IDA824, 21JDC783; 31) REF DYNJ45576A, UDI/DI 10888277893665 (EA) 40888277893666 (CS), Lot Numbers: 21JBV694; 32) REF DYNJ46551B, UDI/DI 10193489715286 (EA) 40193489715287 (CS), Lot Numbers: 19JMB469; 33) REF DYNJ46776D, UDI/DI 10195327037178 (EA) 40195327037179 (CS), Lot Numbers: 21KBO473, 21LBO435; 34) REF DYNJ47085C, UDI/DI 10889942844418 (EA) 40889942844419 (CS), Lot Numbers: 19CDD240, 19GDC707, 19HMA803, 20BMB244, 20EMB373, 20HMC737, 20IMB980, 21DMD706, 21GMC329, 21HMF183, 21JMD528, 22BMF758; 35) REF DYNJ47239B, UDI/DI 10888277820029 (EA) 40888277820020 (CS), Lot Numbers: 19ABN793, 19CBO125, 19EBX648, 19GBX624, 19JBC114, 19WBA380, 20BBE712, 20EBI061, 20GBE240, 20LBV787, 21CBO951, 21CBS176, 21DBA820, 21DBB544, 21JBW651, 21KBC241, 21LBA985, 22BBT765, 22DBL195, 22EBF781, 22HBN419, 22JBP786, 22KBS728, 23BBO504, 23CBS055, 23DBL233, 23JBV371, 23KBW804, 24ABU614, 24BBN264, 24DBV437; 36) REF DYNJ47555C, UDI/DI 10889942376094 (EA) 40889942376095 (CS), Lot Numbers: 19ABA628, 19ABK936, 19DBB411, 19GBJ579, 19JBB782, 19LBA766, 19LBM575, 19LBY411, 19TBD241, 20EBD213, 20EBM194, 20LBQ927, 21ABC141; 37) REF DYNJ49926B, UDI/DI 10889942678136 (EA) 40889942678137 (CS), Lot Numbers: 19ABG084, 19BBJ125, 19CBN213, 19DBN436, 19EBX441, 19EBX695, 19JBC761, 19JBQ465, 20ABR294, 20EMB477, 20FMA631, 20HMA734; 38) REF DYNJ49926C, UDI/DI 10193489415360 (EA) 40193489415361 (CS), Lot Numbers: 20JMA659; 39) REF DYNJ49926D, UDI/DI 10193489471595 (EA) 40193489471596 (CS), Lot Numbers: 21AMA902, 21AMD139, 21CMD402, 21EME405, 21HME635, 21JMD409, 21KMG897; 40) REF DYNJ50687C, UDI/DI 10193489605952 (EA) 40193489605953 (CS), Lot Numbers: 19ADC422, 19BDB987, 19DDC853, 19IDC428, 19KDB096, 19LDC226, 20ADC318, 20FDA097, 20GDB406, 20HDB868, 20IDA319, 20LDB765, 21BDA108, 21DDA227; 41) REF DYNJ50687D, UDI/DI 10193489904383 (EA) 40193489904384 (CS), Lot Numbers: 21FDC030, 21HDB849, 21KDA466, 22ADA004; 42) REF DYNJ50687F, UDI/DI 10195327103446 (EA) 40195327103447 (CS), Lot Numbers: 22BDB471, 22EDA359, 22GDB132, 22LDA898, 23CDA622, 23EDA893, 23HDC392, 23JDB014, 23JDB096, 23LDC131, 24GDA465, 24IDA748, 24IDA814; 43) REF DYNJ50692D, UDI/DI 10193489904390 (EA) 40193489904391 (CS), Lot Numbers: 22AMB929; 44) REF DYNJ50864, UDI/DI 10889942029723 (EA) 40889942029724 (CS), Lot Numbers: 19ADB495; 45) REF DYNJ51726B, UDI/DI 10889942975921 (EA) 40889942975922 (CS), Lot Numbers: 19ABC323, 19BBO624, 19DBG311, 19EBO941, 19HBN410, 19IBA770, 19IBC937, 19JBQ566, 19KBY548, 19QBB234, 20ABE830, 20ABV915, 20BBQ948, 20CBM145, 20FBN532, 20FBV841, 20GBK875, 20HBN815, 20JBE568, 20KBC854, 20LBK395; 46) REF DYNJ52217B, UDI/DI 10889942706464 (EA) 40889942706465 (CS), Lot Numbers: 19CBU746, 19EBC205, 19GBK195, 19IBN942, 19KBE949, 19LBQ997, 19VBE458, 20ABD510, 20ABO061; 47) REF DYNJ52217C, UDI/DI 10193489439212 (EA) 40193489439213 (CS), Lot Numbers: 21BBO084, 21DBS699, 21EBB088; 48) REF DYNJ52217D, UDI/DI 10193489948318 (EA) 40193489948319 (CS), Lot Numbers: 22KBG459, 22OBG364; 49) REF DYNJ53548A, UDI/DI 10889942765942 (EA) 40889942765943 (CS), Lot Numbers: 19ABN458, 19EBA147; 50) REF DYNJ54720, UDI/DI 10889942504176 (EA) 40889942504177 (CS), Lot Numbers: 19ABT998, 19DBU644, 19DBU645, 19FBD509, 19GBE418, 19HBW788, 19IBN909, 19JBM472, 19LBB903, 20ABW499, 20BBO627, 20CBE901, 20DBA778, 20GBH109; 51) REF DYNJ55161, UDI/DI 10889942530649 (EA) 40889942530640 (CS), Lot Numbers: 19ABY393, 19DBC739, 19FBU339, 19HBE884, 19IBF379, 19KBR550, 19LBN878, 20BBC013, 20DBC977, 20EBG090, 20GBS648, 21ABG356, 21CBG361, 21FBL722, 21LBO396, 22CBK047, 22DBG379, 22EBI751, 22IBL566, 22JBQ711, 22KBS818, 22OBK860, 23ABN517, 23CBP855, 23DBG358, 23HBF223, 23JBC620; 52) REF DYNJ56362, UDI/DI 10889942637980 (EA) 40889942637981 (CS), Lot Numbers: 19ABC115, 19BBK164, 19CBS923, 19DBR713, 19EBT879, 19HBT840, 19IBN948, 19KBE484, 19LBY561, 20DBV346, 20GBP800, 20HBW339, 20JBC737, 21DBA837, 21DBP071, 21GBO709, 21IBN160, 21KBG434, 21KBY327, 21LBU350, 21VBB870; 53) REF DYNJ57770A, UDI/DI 10193489669916 (EA) 40193489669917 (CS), Lot Numbers: 19EBY202; 54) REF DYNJ58247C, UDI/DI 10193489266634 (EA) 40193489266635 (CS), Lot Numbers: 21KBQ502; 55) REF DYNJ58471, UDI/DI 10889942856251 (EA) 40889942856252 (CS), Lot Numbers: 19ABB664, 19DBB655, 19LBL298, 20EBG804, 20GBU730, 20WBC107; 56) REF DYNJ58648K, UDI/DI 10195327596538 (EA) 40195327596539 (CS), Lot Numbers: 24BMJ716, 24FME844, 24FMI959, 24HMC399, 24HME448; 57) REF DYNJ58856A, UDI/DI 10193489750263 (EA) 40193489750264 (CS), Lot Numbers: 19IBK249, 19KBB669, 20BBK962, 20FBC493, 20JBS044, 21CBT495, 21CBV908, 21FBB419, 21GBA467, 21HBL409, 21KBF933, 21LBQ393, 22ABH862; 58) REF DYNJ58857A, UDI/DI 10193489733631 (EA) 40193489733632 (CS), Lot Numbers:19GBL697, 19JBG476, 20CBA709, 20CBF266, 20KBW594, 21ABH961, 21BBF515, 21BBS123, 21FBH835, 21JBC414; 59) REF DYNJ59018A, UDI/DI 10193489233469 (EA) 40193489233460 (CS), Lot Numbers: 19LBV363, 20ABV672, 20CBF463, 20FBU157; 60) REF DYNJ59109A, UDI/DI 10888277799288 (EA) 40888277799289 (CS), Lot Numbers: 19ABC064; 61) REF DYNJ59170A, UDI/DI 10193489958973 (EA) 40193489958974 (CS), Lot Numbers: 21GLB014, 21HLA567, 21ILA983, 21JLA186, 21KLA534, 22ALA556, 22DLB034, 22FLB039, 22ILA418, 22JLA976, 22LLA419, 22LLA613, 23CLB115, 23DLA689, 23ELA631, 23FLA255, 23HLB006, 23ILA760, 23KLA492, 24ALA462, 24DLB099, 24FLA916, 24HLA319; 62) REF DYNJ59384B, UDI/DI 10193489564327 (EA) 40193489564328 (CS), Lot Numbers: 19ABC019; 63) REF DYNJ59384C, UDI/DI 10193489606034 (EA) 40193489606035 (CS), Lot Numbers: 19BBJ867, 19BBL515, 19EBW226, 19GBW144; 64) REF DYNJ59384D, UDI/DI 10193489764758 (EA) 40193489764759 (CS), Lot Numbers: 19LBD983, 20ABJ547; 65) REF DYNJ59613, UDI/DI 10888277763487 (EA) 40888277763488 (CS), Lot Numbers: 19ABD529, 19CBG723, 19QBB146, 19SBE142; 66) REF DYNJ60111B, UDI/DI 10195327527617 (EA) 40195327527618 (CS), Lot Numbers: 24BMI703; 67) REF DYNJ62068, UDI/DI 10193489622812 (EA) 40193489622813 (CS), Lot Numbers: 19CBN388, 19CBQ974, 19EBH826, 19EBL004, 19RBB843; 68) REF DYNJ62646, UDI/DI 10193489675498 (EA) 40193489675499 (CS), Lot Numbers: 19EKD615, 19IKB018, 20IKA058, 20IKA452, 20JMD352, 20LMA201, 20LMD695, 21IMF226, 21LMD801, 22CME540, 22HMD949; 69) REF DYNJ62646A, UDI/DI 10195327267414 (EA) 40195327267415 (CS), Lot Numbers: 23AMH061, 23BMI003, 23EMJ495, 23HMG270, 24AME185; 70) REF DYNJ62764, UDI/DI 10193489688061 (EA) 40193489688062 (CS), Lot Numbers: 19EBT170; 71) REF DYNJ62764A, UDI/DI 10193489755961 (EA) 40193489755962 (CS), Lot Numbers: 19HBC794, 19HBW785, 20IBB843, 20JBR893, 20KBV951, 20WBG812, 20XBG045, 21CBZ486, 21EBS505, 21FBR944, 21GBE120, 21JBO722, 21LBG676; 72) REF DYNJ62780A, UDI/DI 10193489230697 (EA) 40193489230698 (CS), Lot Numbers: 20ADC718, 20BDB082, 20EDB446; 73) REF DYNJ62780B, UDI/DI 10193489421828 (EA) 40193489421829 (CS), Lot Numbers: 20JDA431, 20KDB071, 21DDA022, 21EDA262, 21HDA312, 21JDB852, 21KDA414; 74) REF DYNJ62780C, UDI/DI 10195327067144 (EA) 40195327067145 (CS), Lot Numbers: 22ADA579l; 75) REF DYNJ63205, UDI/DI 10193489728699 (EA) 40193489728690 (CS), Lot Numbers: 19JDA642; 76) REF DYNJ63205A, UDI/DI 10193489303421 (EA) 40193489303422 (CS), Lot Numbers: 20EDC001, 20FDA134, 20FDC056, 20KDB738, 20LDA228, 21CDB555; 77) REF DYNJ63205B, UDI/DI 10193489962390 (EA) 40193489962391 (CS), Lot Numbers: 21LDC090, 22BDC017; 78) REF DYNJ63315A, UDI/DI 10193489290783 (EA) 40193489290784 (CS), Lot Numbers: 20BBC775, 20CBO634, 20EBA792, 20EBL285, 20HBW678, 20IBC968, 21CBT485, 21CBU820, 21DBF497; 79) REF DYNJ63315B, UDI/DI 10193489958102 (EA) 40193489958103 (CS), Lot Numbers: 21KBW892, 22BBV989, 22FBR138; 80) REF DYNJ63335, UDI/DI 10193489736212 (EA) 40193489736213 (CS), Lot Numbers: 19HBB301, 19LBL786, 20ABT141, 20BBI774; 81) REF DYNJ63335A, UDI/DI 10193489322361 (EA) 40193489322362 (CS), Lot Numbers: 20DBC076, 20EBM372, 20GBA396, 20HBA333, 20JBE350, 20XBH866; 82) REF DYNJ63335B, UDI/DI 10193489836578 (EA) 40193489836579 (CS), Lot Numbers: 21CBT574, 21EBA752, 21GBS999, 21KBE861; 83) REF DYNJ63622, UDI/DI 10193489770407 (EA) 40193489770408 (CS), Lot Numbers: 19HBP048, 19LBA378, 19LBY005, 20ABX203, 20BBL566, 20CBJ379; 84) REF DYNJ63622A, UDI/DI 10193489340235 (EA) 40193489340236 (CS), Lot Numbers: 20DBP547, 20EBU631, 20FBN537, 20FBO604, 20GBA484, 20HBH050, 20JBK458, 20KBD037, 20WBF187, 20XBC232, 21BBO769, 21DBG288; 85) REF DYNJ63622B, UDI/DI 10193489899344 (EA) 40193489899345 (CS), Lot Numbers: 21FBB733, 21GBP791, 21IBF749, 21JBF288, 21KBK620, 21LBG779, 22BBV814, 22CBF337; 86) REF DYNJ64210, UDI/DI 10193489794595 (EA) 40193489794596 (CS), Lot Numbers: 23HBN332; 87) REF DYNJ65792A, UDI/DI 10193489895322 (EA) 40193489895323 (CS), Lot Numbers: 21KBQ541; 88) REF DYNJ66448, UDI/DI 10193489356106 (EA) 40193489356107 (CS), Lot Numbers: 20EBV507, 21HBP922; 89) REF DYNJ66749, UDI/DI 10193489375947 (EA) 40193489375948 (CS), Lot Numbers: 20HBM389, 20JBS998, 21ABO570, 21ABT938; 90) REF DYNJ67384, UDI/DI 10193489892161 (EA) 40193489892162 (CS), Lot Numbers: 21FMA932, 21JMA446, 21KMF313, 22AMA930, 22AMF818, 22CMG635, 22FMB146; 91) REF DYNJ67384A, UDI/DI 10195327209087 (EA) 40195327209088 (CS), Lot Numbers: 22HMG870; 92) REF DYNJ67384B, UDI/DI 10195327284985 (EA) 40195327284986 (CS), Lot Numbers: 23BLA676, 23ELA051, 23FLA625, 23GLA180, 23GLB044, 23HLA430, 23HLB295, 23JLB234, 24BLA786, 24FLA125; 93) REF DYNJ67423, UDI/DI 10193489892499 (EA) 40193489892490 (CS), Lot Numbers: 21FLA896, 21JLA004, 22ALA838, 22BLB045; 94) REF DYNJ67423A, UDI/DI 10195327163402 (EA) 40195327163403 (CS), Lot Numbers: 22GLA325, 22GLA884, 22HLA029, 22KLA652, 22LLA582, 23BLA617, 23DLA185, 23DLA463, 23FLA068, 23HLB011, 23HLB296, 23ILA503, 23ILB043, 23KLA058, 24BLA235, 24DLA092; 95) REF DYNJ67423B, UDI/DI 10195327686871 (EA) 40195327686872 (CS), Lot Numbers: 24DLA759, 24ELA763, 24FLA258, 24GLB018; 96) REF DYNJ67706, UDI/DI 10193489425154 (EA) 40193489425155 (CS), Lot Numbers: 20JBA421, 21DBH563, 21FBP348, 21IBS916; 97) REF DYNJ68171, UDI/DI 10193489492323 (EA) 40193489492324 (CS), Lot Numbers: 20LBV298, 21BBK016, 21BBL776, 21CBD825, 21CBU780, 21CBZ345, 21FBF710, 21FBO048; 98) REF DYNJ68171A, UDI/DI 10193489991925 (EA) 40193489991926 (CS), Lot Numbers: 21JBE229, 21LBJ092, 21VBC813, 22CBS835, 22EBG817, 22EBU962, 22GBK323, 22KBD430, 23DBB591, 23DBO819, 23EBX020, 23GBJ979, 23KBR717; 99) REF DYNJ68985, UDI/DI 10193489871654 (EA) 40193489871655 (CS), Lot Numbers: 21DBM210, 21GBU579; 100) REF DYNJ68985A, UDI/DI 10193489970494 (EA) 40193489970495 (CS), Lot Numbers: 21IBQ118, 21IBQ119, 21JBO506, 21KBA738, 21LBF016, 21LBQ309, 22CBA233, 22EBT803, 22HBP282, 22LBT057, 23GBQ067, 23IBJ504, 23KBR815, 24BBL934, 24CBN950; 101) REF DYNJ68985B, UDI/DI 10198459019036 (EA) 40198459019037 (CS), Lot Numbers: 24GBP722; 102) REF DYNJ69260, UDI/DI 10193489920871 (EA) 40193489920872 (CS), Lot Numbers: 21GBF659, 21IBM258, 21KBD037, 21KBS889, 21KBY234, 22ABX295, 22BBP691; 103) REF DYNJ69719, UDI/DI 10193489973051 (EA) 40193489973052 (CS), Lot Numbers: 21IBM594; 104) REF DYNJ69829, UDI/DI 10193489982671 (EA) 40193489982672 (CS), Lot Numbers: 21LBH546; 105) REF DYNJ80225, UDI/DI 10195327016661 (EA) 40195327016662 (CS), Lot Numbers: 22ABA788; 106) REF DYNJ80526A, UDI/DI 10195327063566 (EA) 40195327063567 (CS), Lot Numbers: 22BBB176; 107) REF DYNJ80764, UDI/DI 10195327076726 (EA) 40195327076727 (CS), Lot Numbers: 22BBC884, 22LBS552, 24DBE999, 24FBD723; 108) REF DYNJ81950B, UDI/DI 10195327401931 (EA) 40195327401932 (CS), Lot Numbers: 23HBN340; 109) REF DYNJ83679, UDI/DI 10195327296698 (EA) 40195327296699 (CS), Lot Numbers: 23BBN423; 110) REF DYNJ84244, UDI/DI 10195327345860 (EA) 40195327345861 (CS), Lot Numbers: 23DBF981, 23IBO352; 111) REF DYNJ84975, UDI/DI 10195327443467 (EA) 40195327443468 (CS), Lot Numbers: 23FBN786, 23GBO320, 23JBG699, 24GBA863; 112) REF DYNJ85422, UDI/DI 10195327499716 (EA) 40195327499717 (CS), Lot Numbers: 24ADB100, 24FDA925; 113) REF DYNJ900727C, UDI/DI 10193489275674 (EA) 40193489275675 (CS), Lot Numbers: 20CDA183, 20FDB427, 20JDD618, 21CDA708, 21CDB318, 21EDB862, 22BDA961, 22CDA005, 22EDA639, 22GDB814, 22LDB298, 23DDA533, 23EDB580, 23FDB726, 23HDA714; 114) REF DYNJ900727D, UDI/DI 10198459000263 (EA) 40198459000264 (CS), Lot Numbers: 24FDA037; 115) REF DYNJ900777B, UDI/DI 10889942819768 (EA) 40889942819769 (CS), Lot Numbers: 19BBC646, 19BBI519, 19DBW920, 19IBF903, 19QBD239, 19SBA173, 19TBA702, 20ABE354, 20DBP989, 21ABM626, 21GBL292, 21JBV958, 21KBA451, 21KBY795, 22ABG006, 22FBX280, 22GBR293, 22HBL267, 23ABC477, 23BBM671, 23EBE690, 23EBT053; 116) REF DYNJ901073G, UDI/DI 10193489944600 (EA) 40193489944601 (CS), Lot Numbers: 21KMH462; 117) REF DYNJ902512C, UDI/DI 10889942781218 (EA) 40889942781219 (CS), Lot Numbers: 19FBF050, 19JBV446, 19LBS994, 20BBD323, 20BBP150, 20CBA326, 20DBP970, 20EBF588, 20IBE007, 20IBL285, 20KBH533, 20KBL976, 21CBV550, 21FBK159, 21FBS455, 21JBW039, 22ABT817, 22OBC132, 22OBC143; 118) REF DYNJ903036C, UDI/DI 10889942757954 (EA) 40889942757955 (CS), Lot Numbers: 19ABW972; 119) REF DYNJ903444B, UDI/DI 10889942898008 (EA) 40889942898009 (CS), Lot Numbers: 19CBS101, 19DBQ708, 19FBZ889, 19HBD273, 19KBZ023, 20BBW965, 20LBY882, 21CBY280, 21IBQ946, 21JBJ772, 21KBH122, 21KBK371, 22CBB889, 22OBM621; 120) REF DYNJ903565A, UDI/DI 10889942975266 (EA) 40889942975267 (CS), Lot Numbers: 19BBR817; 121) REF DYNJ903565B, UDI/DI 10193489681307 (EA) 40193489681308 (CS), Lot Numbers: 19FBR670, 19GBZ347, 19IBS468, 19KBV841, 19LBJ444, 19LBX494, 20ABQ639, 20DBH136, 20GBQ975; 122) REF DYNJ903565C, UDI/DI 10193489459395 (EA) 40193489459396 (CS), Lot Numbers: 21LBW280, 21LBY303, 22BBA386, 22CBM614, 22DBR276; 123) REF DYNJ903834A, UDI/DI 10889942677108 (EA) 40889942677109 (CS), Lot Numbers: 19BBR974, 19CBK529, 19EBD263, 19IBC620, 19SBE004; 124) REF DYNJ903834B, UDI/DI 10193489204322 (EA) 40193489204323 (CS), Lot Numbers: 19KBH824, 20ABP109, 20BBW872, 20IBE124, 20JBI672, 20JBM256, 20LBS257, 21ABB044, 21GBP417, 21LBX238, 22GBK200; 125) REF DYNJ904555B, UDI/DI 10889942965656 (EA) 40889942965657 (CS), Lot Numbers: 19CBT669, 19EBG357, 19FBJ790, 19GBS312; 126) REF DYNJ904555C, UDI/DI 10193489784305 (EA) 40193489784306 (CS), Lot Numbers: 19JBF520, 19KBL046, 20BBE148; 127) REF DYNJ904555D, UDI/DI 10193489308952 (EA) 40193489308953 (CS), Lot Numbers: 20CBW502, 20DBL265, 20EBI849, 21ABW297; 128) REF DYNJ904555F, UDI/DI 10193489832839 (EA) 40193489832830 (CS), Lot Numbers: 23IBL165; 129) REF DYNJ905249A, UDI/DI 10193489504392 (EA) 40193489504393 (CS), Lot Numbers: 19IBY297, 19KBH885, 19VBF326, 20ABT938, 20BBM086, 20DBH137, 20JBJ922, 20KBD413, 21ABR613, 21FBS428, 21FBS429, 21HBF347; 130) REF DYNJ905340A, UDI/DI 10193489502138 (EA) 40193489502139 (CS), Lot Numbers: 19CBI094, 19DBC223, 19EBC012, 19EBO259, 19EBY277, 19GBO678, 19KBM221, 19SBC082, 19UBA530, 19VBC017, 20ABW774, 20BBW624, 20CBP625, 20DBP908, 20EBJ878, 20EBO371, 20EBS238, 20JBX262, 20WBG752, 21CBU561, 21CBV937, 21DBO239, 21EBA135; 131) REF DYNJ905649, UDI/DI 10888277780774 (EA) 40888277780775 (CS), Lot Numbers: 19CBF660; 132) REF DYNJ905685C, UDI/DI 10195327536619 (EA) 40195327536610 (CS), Lot Numbers: 24EDA215; 133) REF DYNJ905866, UDI/DI 10193489547184 (EA) 40193489547185 (CS), Lot Numbers: 19CBG162, 19GBF254, 19JBT864, 20ABU002, 20CBJ194, 20DBL272, 20HBG841, 20IBL795, 20JBX950, 21ABB887, 21GBI107, 21GBW360, 21JBK091, 21KBA508, 21LBX118, 22ABZ555; 134) REF DYNJ906334, UDI/DI 10193489740066 (EA) 40193489740067 (CS), Lot Numbers: 20BBA182; 135) REF DYNJ906334A, UDI/DI 10193489480726 (EA) 40193489480727 (CS), Lot Numbers: 21ABW672, 21CBZ278; 136) REF DYNJ906838, UDI/DI 10193489261660 (EA) 40193489261661 (CS), Lot Numbers: 20BBE353, 20DBB620, 20HBQ735, 20IBL479, 21ABX036, 21CBM474, 21DBK177, 21DBP124, 21EBK758, 21FBL081, 21IBR113, 21LBT107, 22EBB407, 22IBB422, 22IBD323, 22IBE309, 22JMG245, 22KMG721, 22OBA564, 23DMF269, 23EMH758, 23FMA525, 23GMF521, 23HMB863, 23KMF381; 137) REF DYNJ906838A, UDI/DI 10195327592745 (EA) 40195327592746 (CS), Lot Numbers: 24EMD602, 24EMF972, 24FMF235, 24GME809; 138) REF DYNJ906839A, UDI/DI 10195327092528 (EA) 40195327092529 (CS), Lot Numbers: 22CBB869, 22FBY308, 22JMB394, 23AMB381, 23DMB259, 23EMF065, 23FMH068; 139) REF DYNJ906922, UDI/DI 10193489288070 (EA) 40193489288071 (CS), Lot Numbers: 20CBF639, 20CBX011, 20EBO876, 20GBO466; 140) REF DYNJ906922A, UDI/DI 10193489391107 (EA) 40193489391108 (CS), Lot Numbers: 20IBL327, 20KBY241, 21ABA913; 141) REF DYNJ906922B, UDI/DI 10193489493887 (EA) 40193489493888 (CS), Lot Numbers: 21FBP621, 21GBL188, 21HBU984, 21JBM757, 22CBP206; 142) REF DYNJ907268, UDI/DI 10193489385748 (EA) 40193489385749 (CS), Lot Numbers: 20HBH761; 143) REF DYNJ908032, UDI/DI 10193489963809 (EA) 40193489963800 (CS), Lot Numbers: 21JDC057, 22FDB542; 144) REF DYNJ908237, UDI/DI 10195327013820 (EA) 40195327013821 (CS), Lot Numbers: 21KDA699, 21KDB723; 145) REF DYNJ908294, UDI/DI 10195327031800 (EA) 40195327031801 (CS), Lot Numbers: 21LBI523, 22BBD332, 22GBP001; 146) REF DYNJ909872, UDI/DI 10195327495367 (EA) 40195327495368 (CS), Lot Numbers: 23KBH139; 147) REF DYNJ909872A, UDI/DI 10195327607890 (EA) 40195327607891 (CS), Lot Numbers: 24CBB812, 24CBF038, 24CBU588; 148) REF DYNJG001006B, UDI/DI 10195327678968 (EA) 40195327678969 (CS), Lot Numbers: 24ELA384, 24GLB039; 149) REF DYNJG001008, UDI/DI 10195327080020 (EA) 40195327080021 (CS), Lot Numbers: 22CLB277, 23BLA957, 23HLA656; 150) REF DYNJG001021, UDI/DI 10195327130800 (EA) 40195327130801 (CS), Lot Numbers: 22GLA576, 22JLB233, 23ALA650, 23CLA663, 23CLA733; 151) REF DYNJG001021A, UDI/DI 10195327563226 (EA) 40195327563227 (CS), Lot Numbers: 23KLA960, 24DLB111, 24ELA368, 24ELA781, 24FLA975; 152) REF DYNJVB1295, UDI/DI 10193489440973 (EA) 40193489440974 (CS), Lot Numbers: 21CLA332, 21ELA779, 21GLA166, 21ILA793, 21ILA930, 21JLA670, 21JLA925, 21KLA312, 22BLA834, 22BLA837, 22CLA611, 22CLB061, 22DLA832, 22FLA966, 22GLA936; 153) REF DYNJVB1295A, UDI/DI 10195327227265 (EA) 40195327227266 (CS), Lot Numbers: 22HLA826, 22ILA601, 22KLA904, 23BLA429, 23FLA974, 23GLA884, 23GLA981, 23HLA532, 23KLA358, 23KLA845, 23KLB023, 23LLA548; 154) REF PHS408470B, UDI/DI 10889942818303 (EA) 40889942818304 (CS), Lot Numbers: 19ABD941, 19ABK626, 19CBU486; 155) REF SYNJ10083C, UDI/DI 10889942848492 (EA) 40889942848493 (CS), Lot Numbers: 19BBQ934, 19HBU085, 19JBF267, 19JBX380, 20DBJ553, 20FBH753, 20KBW738, 21FBB565, 21FBP956, 21FBP960.
Devices named in this recall
153 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- Medline Industries, Inc. DYNJ0101278FMedline Industries, Inc. · Ear, nose, and throat surgical tray4 recalls0 reports
- Medline Industries, Inc. DYNJ905866Medline Industries, Inc. · Ear, nose, and throat surgical tray4 recalls0 reports
- Medline Industries, Inc. DYNJ58857AMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ58856AMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ62764AMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ901073GMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ63315BMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ68171AMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ68555AMedline Industries, Inc. · Cardiovascular procedure kit3 recalls0 reports
- Medline Industries, Inc. DYNJ906839AMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67423AMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ60111BMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67423BMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ37551AMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ37383CMedline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ55161Medline Industries, Inc. · Ear, nose, and throat surgical tray3 recalls0 reports
- Medline Industries, Inc. DYNJ903834BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. CDS983087GMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ906838Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ58247CMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63205AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ66448Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ66749Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0826316LMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. CDS983302IMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ10130DMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903565CMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ906922BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ905340AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ62068Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ64210Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904555FMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63335BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ65792AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63622BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. CDS983483IMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69260Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ59170AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63205BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69719Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69829Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ80225Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908294Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ46776DMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ80526AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ62780CMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ50687FMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ67384AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ81950BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ84975Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ85422Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ906838AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ37254JMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ58648KMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ909872AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ40360FMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJG001006BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ900727DMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. CDS981185BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ47239BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ45576AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43582AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ40367CMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ56362Medline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ40360DMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ53548AMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. PHS408470BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ47085CMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903444BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ51726BMedline Industries, Inc. · Ear, nose, and throat surgical tray2 recalls0 reports
- Medline Industries, Inc. DYNJ62780AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ59018AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ900727CMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ906922Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ63315AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ904555DMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ37254GMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ63335AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ63622AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ40094CMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ907268Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ906922AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ52217CMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJVB1295Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ49926DMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ906334AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ68171Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ905249AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. CDS984313GMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ59384BMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ50687CMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ59384CMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ57770AMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ62646Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ41923CMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ903565BMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ62764Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ46551BMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ63205Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
- Medline Industries, Inc. DYNJ63335Medline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95852).
- Z-0777-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) CATH LAB PACK-LF, REF DYNJ0382559M; 2) CRANI, REF DYNJ56001B; 3) GENDER AFFIRMING SURGERY, REF DYNJ66252; 4) GENDER AFFIRM…
- Z-0778-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) OPEN HEART CDS-LF, REFCDS840451M; 2) NEURO PACK, REFDYNJ41516M; 3) RFD- AAA PACK, REFDYNJ54242B; 4) OPEN HEART A&B, REFDYN…
- Z-0779-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) VP SHUNT CDS, REF CDS983470D; 2) 1/2X1/2 NEURO SPONGE PACK-LF, REF DYNJ0713177; 3) BURR HOLE SHUNT PACK, REF DYNJ56819D; 4…
- Z-0780-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) UMC DR. BAKER SUPPLEMENT, REF DYNJ69476; 2) DIEP FLAP, REF DYNJ910120
- Z-0781-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) NASAL PACK, REF DYNJ67789A.
- Z-0783-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) FREE FLAP BASIC SET-UP PK, REF DYNJ0991320J; 2) MVFF/FIBU #54-RF, REF DYNJ21927R; 3) FUNCTIONAL SINUS ENDOSCOPY, REF DYNJ40…
- Z-0784-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) ENDOSCOPIC SKULL BASE, REF DYNJ58648B; 2) ENDOSCOPIC SKULL BASE, REF DYNJ58648C; 3) ENT ENDO SINUS WILMINGTON PACK, REF DYN…
- Z-0785-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) PANENDOSCOPY-LF, REF DYNJ36728C; 2) ENDOSCOPIC SINUS PACK, REF DYNJ40644A; 3) ENDO SINUS PACK, REF DYNJ66449.
- Z-0786-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) MEM MAXILLOFACIAL PACK-LF, REF DYNJ0660040R; 2) FREE FLAP PACK-LF, REF DYNJ42833A; 3) FREE FLAP PACK-LF, REF DYNJ42833B; 4) T…
- Z-0787-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) NEURO BASIN PACK-LF, REF DYNJ39461J; 2) NEURO BASIN KIT, REF DYNJ80200B.
- Z-0788-2025 · Open, ClassifiedSee DocMan
- Z-0789-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) MICRODISECTOMY PACK, REF DYNJ80585A; 2) MICRODISECTOMY PACK, REF DYNJ80585B.
- Z-0790-2025 · Open, ClassifiedSee DocMan
- Z-0791-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: DEEP BRAIN STIMULATION PACK-LF
- Z-0792-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) ENSEMBLE RACHIS-LF, REF DYNJ47568B; 2) UROLOGY ROBOTIC, REF DYNJ55555D; 3) KIT,SPINE LH, REF DYNJ906448A.
- Z-0793-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) CERVICAL PACK-LF, REF DYNJ0867145G; 2) CERVICAL SPINE PACK HUM-LF, REF DYNJ35184F; 3) ANTERIOR CERVICAL-CORPECTOMY, REF DYN…
- Z-0794-2025 · Open, ClassifiedMedline custom medical procedure kits labeled as: 1) CRANI, REF CDS780089V; 2) CRANIOFACIAL CDS, REF CDS780102K; 3) CRANIOFACIAL CDS, REF CDS780102L; 4) CRANIOFACIAL CDS, REF …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page