Neptune recallZ-0807-2025 · Stryker Corporation · initiated Dec 3, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
On December 3, 2024, the firm notified affected consignees via "Urgent Medical Device Field Action" letters dated December 2, 2024. Customers were instructed to locate Neptune S Waste Management Systems and to quarantine affected products at their facility; to contact their Stryker Sales Rep to schedule the replacement of the affected device; to complete Business Reply Form after scheduling replacement and email completed form to [email protected] and maintain information internally until all required action have been completed. Please contact Recall Coordinator at 269-290-0524 or [email protected] with questions or concerns. An URGENT Medical Device Field Action dated 10/13/25 was sent to customers. The purpose of this notification is to advise you that Stryker Instruments has expanded the scope of impacted lots for the Neptune S Waste Management System recall. Due to a distribution error, your facility may have received a recalled device. Next Steps: 1. Locate Neptune S Waste Management Systems and quarantine affected products at your facility Reference the catalog and affected serial numbers in the table above to know if your product is impacted. 2. Contact your Stryker Sales Rep. Your rep will help you schedule the repair of your Neptune device. 3. Complete Business Reply Form (BRF) Once you have scheduled your repair, fill out the attached BRF on page 2. Please email completed form to [email protected]. 4. Maintain this notification internally until all required actions have been completed. Please contact Brooke Thompson at 269-290-0524 or [email protected] with questions or concerns.
Recall details
- Recall number
- Z-0807-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 3, 2024
- Posted by the FDA
- Jan 6, 2025
- Recalling firm
- Stryker Corporation
- Quantity
- 14 (Expanded to 21 units on 10/9/25)
- Product code
- GEI
- FDA submissions
- K143145
- Distribution
- US (Domestic) distribution to states of: WA, GA, and IL. OUS (International) distribution to Japan and Netherlands.
Product description
100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)
Code information (models, lots, serial numbers)
0703 002 000ENG GTIN 07613327382082, Lot Numbers; 2427418723 2427423863 2427418453 2427423873 2427418743 2427419773; 0703 002 000ES GTIN 07613327382068, Lot Numbers: 2427418453; 0703 003 000 GTIN 07613327381948, Lot Numbers: 2427418093 2427418053 2427418073 2427418233 2427418013 2427418723; 0711 001 000 GTIN 07613327576887, Lot Numbers: 2429720623 2429721033 2429720903 2429721203 2429720973 2429721223 2429721023 2303501403 2323610223 2403600253;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Neptune 0711001000Stryker Corporation · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page