Hillrom Temperature Management Controller, MP, Model #… recallZ-0841-2025 · Augustine Temperature Management, LLC · initiated Mar 14, 2022
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Augustine Temperature Management, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
FDA root cause category: Device Design
What the company is doing
Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.
Recall details
- Recall number
- Z-0841-2025
- Status
- Completed
- Initiated by the company
- Mar 14, 2022
- Posted by the FDA
- Jan 10, 2025
- Recalling firm
- Augustine Temperature Management, LLC
- Quantity
- 77 units
- Product code
- DWJ
- Distribution
- Worldwide
Product description
Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI-DI 00850011479582, Serial Numbers: 23010166, 23010167, 23010168, 23010169, 23010170, 23010171, 23010172, 23010173, 23010174, 23010175, 23010176, 23010177, 23010178, 23010179, 23010180, 23010181, 23010182, 23010183, 23010184, 23010185, 24020001, 24020002, 24020003, 24020004, 24020005, 24020006, 24030001, 24030002, 24030003, 24030004, 24030005, 24030021, 24030022, 24030023, 24030024, 24030025, 24030026, 24030027, 24030028, 24030029, 24030030, 24030016, 24030017, 24030018, 24030019, 24030020, 22120033, 24030011, 24030012, 24030013, 24030014, 24030015, 24030006, 24030007, 24030008, 24030009, 24030010, 24060001, 24060002, 24060003, 24060004, 24060005, 24060006, 24060007, 24060008, 24060009, 24060010, 24060024, 24060025, 24060026, 24060027, 24060028, 24060011, 24060012, 24060013, 24060014, 24060015.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Hillrom powered by Augustine Surgical HD 2083516Augustine Temperature Management LLC · System, Thermal Regulating2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95913).
- Z-0839-2025 · CompletedAugustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # WC71, 2) Model # WC71R (refurbished); used in conjunction wit…
- Z-0840-2025 · CompletedAugustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # WC77, 2) Model # WC77D (Demo), 3) Model # WC77R (Refurbished…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.