Brand Name: HALYARD recallZ-0983-2025 · O&M Halyard, Inc. · initiated Nov 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from O&M Halyard, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Elevated residues of anti-block/slip additive used in manufacturing found within sterile packaging.
FDA root cause category: Process control
What the company is doing
On November 26, 2024 URGENT VOLUNTARY RECALL letters were emailed to consignees. ACTIONS TO TAKE: " Immediately discontinue use of the Laparotomy Pack identified in APPENDIX 1. Please review your inventory, and segregate and quarantine all affected Packs. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. " RETURN the completed Field Action Response Form to [email protected] as soon as possible even if you have no current product inventory. ADDITIONAL INFORMATION: Report any quality problems or adverse events associated with the products listed in this notification to [email protected]. Your assistance is appreciated and necessary to prevent any risk to patient health as a result of not following the field correction instructions. If you have any questions, please email [email protected] using the Event # FA-2024-031 in the header of the response. This will ensure prompt response by the O&M Halyard, Inc, team.
Recall details
- Recall number
- Z-0983-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 26, 2024
- Posted by the FDA
- Jan 17, 2025
- Recalling firm
- O&M Halyard, Inc.
- Quantity
- 1440 units
- Product code
- KKX
- FDA submissions
- K083234
- Distribution
- US Nationwide distribution in the states of CO, FL, GA, HI, IA, IL, IN, KS, LA, MI, MN, MO, NC, NE, NY, TN, TX, VA, WI.
Product description
Brand Name: HALYARD Product Name: Laparotomy Pack Model/Catalog Number: 88231 Software Version: N/A Product Description: Laparotomy Pack Component: N/A
Code information (models, lots, serial numbers)
Model No 88231; UDI-DI: 30680651882319; Lot Number: AC2335502D; Exp Date 21-12-28
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Halyard 88231O&M Halyard, Inc. · Drape, surgical1 recall0 reports
Other devices under the same FDA submission
8 devices cleared under K083234 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Halyard 77165O&M Halyard, Inc. · Drape, surgical1 recall0 reports
- Halyard 88041O&M Halyard, Inc. · Drape, surgical1 recall0 reports
- Halyard 42306O&M Halyard, Inc. · Drape, surgical1 report
- Halyard 77769O&M Halyard, Inc. · Drape, surgical1 report
- Halyard 77823O&M Halyard, Inc. · Drape, surgical1 report
- Halyard 88171O&M Halyard, Inc. · Drape, surgical1 report
- Halyard 88719O&M Halyard, Inc. · Drape, surgical1 report
- Halyard 88841O&M Halyard, Inc. · Drape, surgical1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page