Karl Storz - Endoskope - REF 723400 - Optical Biopsy and… recallZ-0837-2025 · Karl Storz Endoscopy · initiated Dec 19, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Karl Storz Endoscopy or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to failure of manual cleaning validation, biopsy and grasping forceps like sterility assurance.

FDA root cause category: Reprocessing Controls

What the company is doing

On December 19, 2024, Karl Storz issued a "Urgent: Medical Device Recall" Notification to affected consignees. In addition to informing consignees about the recall, Karl Storz ask consignees to take the following actions: 1. Immediately quarantine and discontinue use of associated part numbers listed. 2. Pass on this Urgent Medical Device Recall Notice to all users of the recalled product and all other persons who need to be aware within your organization. 3. If you have further distributed the recall product, please identify and promptly notify those recipients by providing a copy of the recall notification and response form, or provide KARL STORZ a list of customers who received/may have received the product listed. 4. Return the completed Customer Reply Form by Fax or E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Contact your KARL STORZ representative or call KARL STORZ Customer Support to return affected product. " Call Customer Support at 800-421-0837 Option 1 to set up an RMA. " Or call your KARL STORZ representative to assist with setting up an RMA. 6. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Recall details

Recall number
Z-0837-2025
Status
Open, Classified
Initiated by the company
Dec 19, 2024
Posted by the FDA
Jan 10, 2025
Recalling firm
Karl Storz Endoscopy
Quantity
285 units
Product code
LRC
Distribution
US: AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OR, PA, SC, TX, VA OUS: None

Product description

Karl Storz - Endoskope - REF 723400 - Optical Biopsy and Grasping Forceps, CE

Code information (models, lots, serial numbers)

Lot Numbers: All lots of the affected material number/Primary DI Number: 04048551194416

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.