Optima recallZ-0813-2025 · GE Medical Systems, LLC · initiated Nov 25, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
FDA root cause category: Component design/selection
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/25/2024 via letter issued using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by the user: "1. Perform a check of your device prior to each patient exam to look for the following behaviors. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. . Column vertical movement requires increased effort. . Column has change in resistance (i.e. movement is not smooth) during vertical movement. . Column makes abnormal noises during vertical movement. 2. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. 3. Please retain this document for your records. 4. Please complete and return the attached acknowledgement form to [email protected]" For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. On 02/11/2025, the firm send an updated notice with a corrected GTIN for Optima XR200amx.
Recall details
- Recall number
- Z-0813-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 25, 2024
- Posted by the FDA
- Jan 7, 2025
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 120 units
- Product code
- IZL
- FDA submissions
- K162990, K173602, K182234
- Distribution
- Worldwide
Product description
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1) 5555000-27: UDI/DI 840682122351, Serial Numbers: 228867XR240, 252847PP241, 281274OPT1, 281338CL240, 281VHVXR240, 305682DIG1, 305682DIG3, 305KEN240B, 401OPTI2, 405949OPT1, 408559XRDR7, 413447OPT1, 419251OPT, 419690DP1, 504842XR242, 602344240A, 602344MARY240, 713440DP2, 713440DP3, 720854DF1, 732222DP3, 763520XR10, 763520XRA6, 806725OPT, 865305DP1, 941276XRA, 970495OPTI1, 973926OPT1, BOLDWPACPORT02, GLEDWPACPORT02, HMHWT240P5, LIHOPTIMA240B, OCHSNERWB240, SAMC240A; 2) 5555000-31: UDI/DI 840682122351, Serial Numbers: 00158MOB17, DF2401800221WK, U416249MOB240; 3) 5555000-33: UDI/DI 840682122351, Serial Numbers: 082407100689, 228867XR240C; 4) 5555000-41: UDI/DI 840682122351, Serial Numbers: 00130MOB16, 5973XR0225, DEF11034, RS4025RX05, SA1058MM50; 5) 5555000-47: UDI/DI 840682122351, Serial Numbers: 206223OPTIMA1, 206248OPT2, 206341OPT2402, 212838OPTIMA1, 219326AMX2, 256HHPORT5, 256HHPORT6, 260824XR240, 270793OPT, 270793OPT1, 304256OPTIMA240, 307324XR240, 318626OPX240FP, 336716WFP2, 336716WFP3, 404616OPT2, 405657OPT, 406345OPT1, 406653XR240, 406653XRA1, 414266XR243, 423837X240, 425228OPT240, 435713OPX240, 435789XR240, 502897M241, 504842OPTFP1, 504894XR244, 513952OPT1, 530888OPT2, 541789M3, 551996DIG1, 706527OPT2, 712662OPT240, 713440DP4, 718918XR240B, 785623OPT2, 786308XR240, 801465XR240, 805652M2, 816943XR240A, 828369OPT1, 850864FWDH240, 865977DP2, 901595XRA1, 914787XR2401, 916781OPT4, 951788XR240P2, 954723XR240B, 972596OPT3, 973877OPT1, AMX240GEN2SYSTEM17, AMX240GEN2SYSTEM9, HMHXR240P7, KAFB228OPTIMA2, LNMCDP1, LVCCER240B, MBMCPORTD; 6) A0700DL: UDI/DI 840682122351, Serial Numbers: 082407250223; 7) A1700DA: UDI/DI 840682122351, Serial Numbers: 210METM2, 256543OPT, 337261OPT, 360744OPT2402, 425899OPT, 425SNORIDGE, 509946AMX1, 573814DP8, 574231OPT2, 702852OPT, 850864FWDH2401, HMHXR240P1, OMCTOPT1; 8) To be provided: UDI/DI To be provided, Serial Numbers: 256543OPT1, 405657FP1417, 513952OPT2, RU1018RX04.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Optima XR240amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K162990, K173602, K182234 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Optima XR240amx HD UpgradeGE Medical Systems, L.L.C. · System, x-ray, mobile1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95958).
- Z-0811-2025 · Open, ClassifiedGE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB
- Z-0812-2025 · Open, ClassifiedGE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page