VITROS recallZ-0902-2025 · Quidel Ortho · initiated Dec 5, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Quidel Ortho or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.
FDA root cause category: Device Design
What the company is doing
On December 5, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to consignees. REQUIRED ACTIONS - For customers who have VITROS 5,1 FS/4600/5600/XT 3400/XT 7600 Systems: Load ADD DRV 6308 upon receipt. - For customers who have VITROS 250/350 Systems: Please be aware of the lowering of the Interferent Concentration limit from 300 mg/dL to 100 mg/dL at a digoxin concentration of 2.0 ng/mL (2.6 nmol/L). - Complete and return the enclosed Confirmation of Receipt form no later than December 13, 2024. - Save this notification with your User Documentation or post this notification by each VITROS 250/350/5,1 FS/4600/5600/XT 3400/XT 7600 System in your laboratory until the issue has been resolved. - Please forward this notification if the affected product was distributed outside of your facility. - If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Resolution Our investigation to revise the limit at which hemoglobin interference may reach an unacceptable level is ongoing. In the interim we will be issuing a Technical Bulletin to implement the temporary changes to the Sample Indices Threshold default hemolysis limit on the ADD and the hemoglobin Interferent Concentration limit for digoxin listed in the IFU. We will communicate again after our investigation has completed. QuidelOrtho is targeting to have an update by Q1 2025. Contact Information We apologize for the inconvenience this may cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-421-3311. To report adverse events, contact your local Global Services Organization. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088. On December 5, 2024 URGENT PRODUCT CORRECTION NOTIFICATION lette
Recall details
- Recall number
- Z-0902-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 5, 2024
- Posted by the FDA
- Jan 10, 2025
- Recalling firm
- Quidel Ortho
- Quantity
- 1,541 units (1,125 US, 416 OUS)
- Product code
- LFM
- FDA submissions
- K942610
- Distribution
- Domestic: Nationwide Distribution International: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom,
Product description
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhythmias. Digoxin measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. One sales unit contains 5 cartridges. Each cartridge holds 18 slides (equivalent to one test), for 90 slides per sales unit. Component: No, not a component
Code information (models, lots, serial numbers)
Model No 8343386; UDI/GTIN 10758750004782; Coating 0272, expiration date; 01-April-2025 The scope of this recall is limited to consignees who received VITROS Chemistry Products DGXN slides in the last 18 months. This will include in-date lots manufactured from coatings (production formulation) 0271 and/or 0272. Of the in date product, the latest expiration date is listed for the coating.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 8343386Ortho-Clinical Diagnostics, Inc. · ENZYME IMMUNOASSAY, DIGITOXIN2 recalls46 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95904).
- Z-0901-2025 · Open, ClassifiedBrand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VI…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗