Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM… recallZ-0920-2025 · Sin Sistema De Implante Nacional S.A · initiated Sep 1, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sin Sistema De Implante Nacional S.A or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
FDA root cause category: Process control
What the company is doing
On 9/1/2024, "URGENT: MEDICAL DEVICE RECALL" notices were emailed to customers and distributors who were told the following: The product should not be used. It does not pose a handling risk and should be separated and discarded. Complete and return the Medical Device Recall Return Response Acknowledgement and Receipt Form via email to: [email protected]. Customer with questions may contact the company via telephone at 0800-770-8290 between the hours of 9 and 19 (Brasilia time) or email the firm at [email protected].
Recall details
- Recall number
- Z-0920-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 1, 2024
- Posted by the FDA
- Jan 15, 2025
- Recalling firm
- Sin Sistema De Implante Nacional S.A
- Quantity
- 123 (US 98, OUS 25)
- Product code
- DZE
- FDA submissions
- K222231
- Distribution
- US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
Product description
Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N
Code information (models, lots, serial numbers)
UDI-DI: 07899995273437, Lot: X010469757, Expiration: 2028/01
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IMP. EPIKUT PLUS MT16° Ø4,5X18,0MM ILM 4518NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form2 recalls0 reports
Other devices under the same FDA submission
32 devices cleared under K222231 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- IMP. EPIKUT PLUS MT16° Ø4,0X20,0MM ILM 4020NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø4,0X18,0MM ILM 4018NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø3,8X20,0MM ILM 3820NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø5,0X15,0MM ILM 5015NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø5,0X13,0MM ILM 5013NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø5,0X11,5MM ILM 5011NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø5,0X10,0MM ILM 5010NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø5,0X8,5MM ILM 5085NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø4,5X15,0MM ILM 4515NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø4,5X13,0MM ILM 4513NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø4,5X8,5MM ILM 4585NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
- IMP. EPIKUT PLUS MT16° Ø4,5X10,0MM ILM 4510NSin Sistema de Implante Nacional S/A · Implant, Endosseous, Root-Form1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page