NAMIC recallZ-0764-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Dec 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
FDA root cause category: Under Investigation by firm
What the company is doing
An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 12/16/24 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-24-270-FG Recall Code: 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. Update: A follow-up letter titled IMMEDIATE ACTION REQUIRED dated 2/11/25 was sent to customers. Medline would like to update our action to instruct you to please destroy all impacted convenience kits and Medline will provide replacement kit(s) with the correct Fluid Delivery Set. REQUIRED ACTION: 1. Immediately check your stock for the affected kit number(s) and lot number(s) listed on the attachment. Please destroy all affected kits. 2. Please complete and return the enclosed destruction form listing the quantity of affected kits destroyed. Even if you do not have any affected product, please complete and return the form. 3. Upon receipt of your completed destruction form, Medline will issue replacem
Recall details
- Recall number
- Z-0764-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 16, 2024
- Posted by the FDA
- Jan 15, 2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 13,600 units
- Product code
- FPA
- FDA submissions
- K852140
- Distribution
- US Nationwide Distribution
Product description
Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- NAMIC 99000069Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE 949000223Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE 949000222Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE 949000221Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE 949000224Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K852140 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NAMIC 64000935Navilyst Medical, Inc. · Set, administration, intravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 95988).
- Z-0765-2025 · Open, ClassifiedMedline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page