Sklar Instruments recallZ-0929-2025 · Sklar Instruments · initiated Dec 3, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sklar Instruments or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reports of various packaging issues that may result in a breach of the sterile barrier.
FDA root cause category: Under Investigation by firm
What the company is doing
On December 3, 2024 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Action required: 1. Check your inventory for any affected product lots (see Attachment A). If you have affected product in inventory, discontinue use immediately. 2. Notify Additional Users. If you have further distributed any of the items referenced in this notification, you must notify those downstream customers of this recall. Forward a copy of this letter, along with copies of Attachment A and the Recall Acknowledgement form, for such customers to complete and return to Sklar. 3. Notify Sklar. Please complete the enclosed Recall Acknowledgement form as soon as possible, indicating whether you do or do not have any of the affected products remaining in your inventory and whether you have distributed any of the affected products to downstream customers. Return the form to Sklar by e-mail to [email protected]. 4. Return the affected product. If you do have any of the affected products in inventory, Sklar customer service will contact you to arrange for the affected product to be returned. Sklar will credit its direct customers for returned affected product once it receives the affected product back. Sklar Instruments is committed to providing high quality, safe, and effective products. We sincerely apologize for any inconvenience this action may cause your operations. If you have experienced an adverse event associated with any of these products, please contact us and/or report the event to FDA s MedWatch Adverse Event Reporting program online.
Recall details
- Recall number
- Z-0929-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 3, 2024
- Posted by the FDA
- Jan 16, 2025
- Recalling firm
- Sklar Instruments
- Quantity
- 75 units
- Product code
- FFO
- Distribution
- Domestic: Nationwide Distribution
Product description
ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE WEITLANER RETR SH 2X3 4.5"CS25, Model Numbers: 96-2201A; 2) ECONO STERILE WEITLANER RETR BL 2X3 4.5"CS25, Model Numbers: 96-2203A; 3) ECONO STERILE WEITLANER RETR BL 3X4 5.5"CS25, Model Numbers: 96-2207A; 4) ECONO STERILE WEITLANER RETR SH 3X4 5.5"CS25, Model Numbers: 96-2208A; 5) ECONO STERILE WEITLANER RETR 3X4 SH 8" CS/10, Model Numbers: 96-2209A; 6) ECONO STERILE WEITLANER RETR 3X4 BL 8" CS/10 , Model Numbers: 96-2210A; 7) ECONO STERILE HOLZHEIMER RETR 4PRG SH 4"CS25, Model Numbers: 96-2241A; 8) ECONO STERILE HOLZHEIMER RETR 4PRG BL 4"CS25, Model Numbers: 96-2242A; surgical retractor
Code information (models, lots, serial numbers)
1) 96-2201A, UDI/DI 50649111493100, Lot Codes: SK-120; 2) 96-2203A, UDI/DI 50649111493094, Lot Codes: SK-120; 3) 96-2207A, UDI/DI 50649111493117, Lot Codes: SK-120; 4) 96-2208A, UDI/DI 50649111493124, Lot Codes: SK-120 5) 96-2209A, UDI/DI 50649111493155, Lot Codes: SK-120; 6) 96-2210A, UDI/DI 50649111493162, Lot Codes: SK-130; 7) 96-2241A, UDI/DI 50649111492820, Lot Codes: SK-115; 8) 96-2242A, UDI/DI 50649111492837, Lot Codes: SK-115;
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- Sklar Instruments 96-2241ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Sklar Instruments 96-2242ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Sklar Instruments 96-2203ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Sklar Instruments 96-2201ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Sklar Instruments 96-2207ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Sklar Instruments 96-2208ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Sklar Instruments 96-2209ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Sklar Instruments 96-2210ASklar Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95946).
- Z-0925-2025 · Open, ClassifiedECONO STERILE brand medical procedure kits labeled as: 1) ECONO STERILE" BUCK EAR CUR BL STR#0STRLCS/25, Model Number 96-2273; 2) ECONO STERILE" BILLEAU EAR LOOP MD STRL CS/25,…
- Z-0930-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE VIENNA NSL SPEC MD STRL CS/25, Model Number 96-2236; 2) ECONO STERILE VIENNA NSL SPEC MD STRL CS/25, Model Numb…
- Z-0931-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: ECONO STERILE" CURETTE EXCAVATOR STRL CS/25, Model Number 941-1320; general surgical
- Z-0932-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE D/A NAIL NIPPER 5.75" STER CS5, Model Number 96-2424A; 2) ECONO STERILE TISS NIPPER 14MM CVX 5"CS/25, Model Num…
- Z-0933-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Nu…
- Z-0934-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: ECONO STERILE CEMENT SPATULA #24 STRL CS/25, Model Number 96-2245M, mixing utensil, dental
- Z-0935-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE VEIN HOOK 1.5MM STERILE CS/25, Model Number 96-4020M; 2) ECONO STERILE VEIN HOOK 2MM STERILE CS/25, Model Numbe…
- Z-0939-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ROCH-PEAN FCP 8" CVD CS/25,Model Number 94-0286; 2) ECONO STERILE DRESSING FCP 5.5" CS/25LG,Model Number 94-02…
- Z-0940-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: ECONO STERILE BLAKE GALLSTONE FCP CVD 8"CS25, Model Number 96-4921A
- Z-0941-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE IRIS SCISS 4 1/2" CVD CASE/50,Model Number 941-0918; 2) ECONO STERILE METZENBAUM SCIS STR 9"STER C25,Model Numb…
- Z-0943-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE SIMS UTERINE CURETTE #3 STER25, Model Number 96-4113; 2) ECONO STERILE KEVORK ENDO CURET W/BAS STER25, Model Nu…
- Z-0944-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE GRAVES SPEC STRL SM BX/25, Model Number 96-2603; 2) ECONO STERILE GRAVES SPEC STRL MD BX/25, Model Number 96-…
- Z-0947-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE MINI TISCH 2.3X4.2MM STER CS10, Model Number 96-4134M; 2) ECONO STERILE OVAL BIOPSY FCP 3X7MM STER10, Model Num…
- Z-0950-2025 · Open, ClassifiedMedical procedure kits labeled as: 1) ECONO STERILE STRABIS SCISS STR 4.5"STER 50, Model Number 96-2540; 2) ECONO STERILE STRABIS SCISS CVD 4.5"STER 50, Model Number 96-2541; 3…
- Z-0951-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: ECONO STERILE IRIS HOOK QK REL 4MM 9" STER25, Model Number 96-2224A; eye surgery
- Z-0953-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE PROBE W/EYE CS/50, Model Number 94-5923; 2) ECONO STERIL" PROBE W/EYE 5.5" STERL CS/25, Model Number 96-2361; o…
- Z-0954-2025 · Open, ClassifiedECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE OLSEN-HEG NH SMO DEL 5.5STER25, Model Number 96-2352A; 2) ECONO STERILE HALSEY NH SERR 5" STRL CS/25, Model Nu…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page