Idylla INSTRUMENT, IVD, REF P0010 recallZ-0836-2025 · Biocartis Nv · initiated Dec 10, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biocartis Nv or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

False-positive MSI-H results generated by the Idylla MSI Test.

FDA root cause category: Environmental control

What the company is doing

Biocartis issued an URGENT - Medical Device Correction notice to its consignees on 12/10/2024 via email. The notice provided the description of the problem, its impact, and requested the following actions be taken: 1. Until your Instrument is serviced, please move forward with the option that applies to your situation: o In case you have another Idylla Instrument available that is not affected, use that Instrument to run the Idylla MSI Test and discontinue running the Idylla MSI Test on the listed Instrument(s). o No new Idylla MSI testing should be performed using the affected Idylla Instruments. o For results already obtained using an affected Idylla Instrument, the following apply: o MSS results are VALID and may be reported. o MSI-H (Microsatellite Instability High) results with six or seven positive identified biomarkers (as depicted in the result report) are VALID and may be reported. o MSI-H results other than those with six or seven positive identified biomarkers are INVALID and should NOT be reported or used to manage patient health. Patient samples with these results should be re-tested on an Idylla Instrument that is NOT affected. 2. If you are not the Idylla user, please forward this notice to the Idylla end user or Idylla supervisor and provide Biocartis with the correct contact details for our record keeping. 3. Completion of the Acknowledgement of Receipt (Appendix 2) is required for compliance. Please complete and sign the attached Acknowledgement of Receipt form by December 17, 2024, and send it to [email protected].

Recall details

Recall number
Z-0836-2025
Status
Open, Classified
Initiated by the company
Dec 10, 2024
Posted by the FDA
Jan 10, 2025
Recalling firm
Biocartis Nv
Quantity
8 units
Product code
OOI
Distribution
US: AL, NJ, TX, OH, CA

Product description

Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.

Code information (models, lots, serial numbers)

UDI/DI 05415219000119, Serial Numbers: 00002706, 00002820, 00003103, 00003262, 00003419, 00003434, 00003444, and 00003445.