HemosIL Heparin Calibrators recallZ-1020-2025 · Instrumentation Laboratory · initiated Dec 18, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range.
FDA root cause category: Under Investigation by firm
What the company is doing
On December 18, 2024 URGENT MEDICAL DEVICE CORRECTION letters were email to customers. Mandatory Customer Actions Based on the above, please take the following immediate actions: - Check inventory for lots of HemosIL Heparin Calibrators (Part No. 0020300600). See Appendix A for a list of in-date lots with corrected value assignments for the HEP CAL 2 and HEP CAL 3 levels. - Follow the instructions below to update the ACL TOP test parameters (same instructions for all models): - Select Setup from the menu bar and then select Test List from the dropdown menu. - From Test List , double click on the Test Code Anti-Xa to open Test Definition Anti-Xa . - From Test Definition Anti-Xa, click on Pre-diluted calibrators under the header Calibration . An editable table titled Calibration points will appear with HEP CAL 1, HEP CAL 2, and HEP CAL 3 and their predefined Target Values IU/mL of 0.00, 0.80 and 2.00, respectively. - In the Calibration points table, change the target values for HEP CAL 2 and HEP CAL 3 to the lot-specific calibrator values provided in Appendix A. - After entering the corrected HEP CAL 2 and HEP CAL 3 values, click on the save icon (picture of floppy disk) to save the changes. - Recalibrate using the HemosIL Liquid Anti-Xa assay lot in use after making the test parameter update. - Run quality controls per labeled instructions for use. If needed, determine control ranges for HemosIL UF Heparin Controls (Part No. 0020300300) and HemosIL Low Molecular Weight Heparin Controls (Part No. 0020300200) in accordance with your internal protocol until the new acceptance ranges are available. - Repeat the above test parameter update, recalibration, and quality control for all ACL TOP instruments. - Share this information with your laboratory staff. - Forward this notification to all affected locations within your facility. - Retain a copy of this notification for your records. - Complete and sign the enclosed Mandatory Response Tracking Form a
Recall details
- Recall number
- Z-1020-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 18, 2024
- Posted by the FDA
- Jan 24, 2025
- Recalling firm
- Instrumentation Laboratory
- Quantity
- 10,952 units (5,303 US, 5,649 OUS) (Additional 11,562 units updated 4/15/2026)
- Product code
- KFF
- FDA submissions
- K213464
- Distribution
- Worldwide - US Nationwide and the countries of Andorra, Algeria, Armenia, Australia, Austria, Belarus, Belgium, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Cyprus, Czechia, Denmark, Finland, France, Georgia, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Korea (the Republic of), Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Macao, Malaysia, Mexico, Netherlands (the), Norway, Philippines (the), Poland, Portugal, Qatar, Romania, Russian Federation (the), Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan (Province of China), Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Product description
HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic
Code information (models, lots, serial numbers)
Part Number: 0020300600; UDI-DI: 08426950472483; Lot Numbers: N0128933, N0220450, N0522479, N0724070, N1026509, N0239584, N0531891, N0733658, N0249207, N0442161, N0744673; ****Lot numbers added 4/15/2026**** N0531891, N0733658, N0249207, N0442161, N0744673, N1147734, N020463, N0250464, N0452277, N0553155, N0754506;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HemosIL Heparin Calibrators 00020300600Instrumentation Laboratory Company · CALIBRATOR, PRIMARY1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.