Nobel Biocare N1 recallZ-1101-2025 · Nobel Biocare · initiated Jan 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nobel Biocare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.
FDA root cause category: Process control
What the company is doing
On 01/10/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Nobel Biocare became aware through internal quality control one of the plastic components of the packaging that is used to hold the OsseoShaper in place may contain a manufacturing error. This results in a sharp edge that could puncture the blister packaging, thereby breaching the sterile barrier. Customers and Distributors are instructed to: 1. Inspect their stock and quarantine affected devices. The use of affected devices should be ceased immediately. 2. For Distributors, complete attached Distributor Acknowledgement Form with information collected from end user and return the firm to Nobel Biocare via email to [email protected] with in 5 days of receipt of this notice. 2. Complete attached Customer Acknowledgment Form, even if customers do not have any affected stock, and return it to Nobel Biocare, via email to [email protected], within 5 days of receipt of this notice. 3. Return all affected stock on hand to Nobel Biocare using the shipping label attached to this notice. 4. Ensure relevant staff members are informed of this recall. If products were supplied or transferred to another facility or organization, inform these facilities of the recall by providing a copy of this notice. For question or further support, contact local customer support representative at 800-322-5001 ext. 1527 or at [email protected]
Recall details
- Recall number
- Z-1101-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 10, 2025
- Posted by the FDA
- Feb 5, 2025
- Recalling firm
- Nobel Biocare
- Quantity
- 1,558 units (90 units - U.S.)
- Product code
- DZE
- FDA submissions
- K211109
- Distribution
- U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.
Product description
. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function
Code information (models, lots, serial numbers)
Nobel Biocare N1 TiUltra TCC RP 4.0x 9mm Catalog Number: 300861 UDI-DI code: 07332747161731 Lot Numbers: 12253672 12253673 Nobel Biocare N1 TiUltra TCC RP 4.0x11mm Catalog Number: 300862 UDI-DI code: 07332747161748 Lot Numbers: 12253448 12254209 Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Catalog Number: 300863 UDI-DI code: 07332747161755 Lot Numbers: 12253620 12254213
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Nobel Biocare N1™ 300861Nobel Biocare AB · IMPLANT, ENDOSSEOUS, ROOT-FORM1 recall1,393 reports
- Nobel Biocare N1™ 300862Nobel Biocare AB · IMPLANT, ENDOSSEOUS, ROOT-FORM1 recall1,280 reports
- Nobel Biocare N1™ 300863Nobel Biocare AB · IMPLANT, ENDOSSEOUS, ROOT-FORM1 recall641 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page