Skip to content

1) Pyxis MedStation ES, Catalog No. 323 recallZ-1119-2025 · CareFusion 303, Inc. · initiated Jan 8, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issues could potentially result in: 1) delays in accessing dispensing software application, 2) wrong therapy (dose, item, timing, etc.), or 3) unauthorized access and/or compliance related issues

FDA root cause category: Software design

What the company is doing

On January 8, 2025, the firm notified customers via email and mailed letter titled "Urgent Medical Device Correction". Firm states customers can continue to use Pyxis products in accordance with their labeling and any previously provided field action notifications, including the following guidance: 1. Facilities should establish policies and procedures for when an automated dispensing system is unavailable. 2. Device Keys can manually access unavailable drawers, but must be readily available and retrievable to adequately mitigate the risk. 3. Emergency supplies should be physically available in care areas for critical or urgent medications. 4. If using critical override mode to access contents of the device, users must use extra precautions and verify correct medications are selected. The firm anticipates releasing software update ES 1.11 beginning June 2025, which will address these software issues.

Recall details

Recall number
Z-1119-2025
Status
Open, Classified
Initiated by the company
Jan 8, 2025
Posted by the FDA
Feb 7, 2025
Recalling firm
CareFusion 303, Inc.
Quantity
137,741 total
Product code
BRY
Distribution
Domestic distribution nationwide. International distribution to North America, Latin America, EMEA, and Greater Asia.

Product description

(1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis MedStation ES Tower, Catalog No. 352; (4) BD Pyxis CII Safe ES, Catalog No. 1116-00

Code information (models, lots, serial numbers)

(1) Pyxis MedStation ES, Catalog No. 323, UDI-DI 10885403512667; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327, UDI-DI 10885403477836; (3) Pyxis MedStation ES Tower, Catalog No. 352, UDI-DI 10885403512674; (4) BD Pyxis CII Safe ES, Catalog No. 1116-00, UDI-DI 10885403512605; All Serial Numbers; Software Versions v1.10 and prior Affected

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates