Temporary Titanium Abutments recallZ-1111-2025 · DDS Lab · initiated Dec 6, 2024

Class II Open, Classified All recalls posted in February 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DDS Lab or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain definitive dental implant restoration cases may have included a temporary titanium abutment (cylinder), rather than a definitive titanium abutment, which may lead the titanium abutment fracture, which may risk loosening of the affixed restoration; fragments could also injure oral mucosa, and could be aspirated.

FDA root cause category: Component design/selection

What the company is doing

On 12/6/2024, correction notices were mailed to customers asking them to do the following: Patients that received a dental implant restoration processed with a temporary titanium abutment (cylinder) rather than a definitive titanium abutment should be contacted to inform them about this Correction. Dental providers should continue to monitor a restoration subject to this Correction for potential problems, consistent with their clinical standard of care for all dental implant restorations. Firm is actively collaborating with dental support organizations and dental offices to ensure patients are notified of this event. If a dental provider identifies a potential problem with a restoration subject to this Correction, please contact the firm. To use case ID to access additional information including patient names, use the following steps in the My DDS Lab Portal: 1) Access the MyDDSLab Portal (https://my.ddslab.com/Login.aspx) 2) Can't remember your office's log-in or password? Call 877-966-9931 to request a reset. 3) Once logged into the portal, navigate to the "Case Management" tab. 4) Search by Case ID to find the patient's name and additional details about the fabrication. 5) Once in the case, you can also choose to download the invoice that will have more specific information about the fabrication. Please direct any inquiries to the vendor supporting the market correction effort via phone at 855-544-4821 Monday thru Friday 8am to 7pm ET or by email at [email protected].

Recall details

Recall number
Z-1111-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Dec 6, 2024
Posted by the FDA
Feb 5, 2025
Recalling firm
DDS Lab
Quantity
11,259
Product code
NHA
FDA submissions
K103691, K071638, K093321, K223697
Distribution
US Nationwide distribution including in the states of FL, TN, TX, MA, NH, CT, ME, WA, IN, CA, PA, NV, MO, IL, AK, MN, VT, OH, MI, VA, WI, AL, AZ, KY, MT, OK, CO, LA, GA, NY, WV, NC, NJ, MD, OR, DE, SC, ID, KS, IA, AR, NM, HI, DC, MS, NE, RI, UT.

Product description

Temporary Titanium Abutments

Code information (models, lots, serial numbers)

Show the full code information (98k characters)

Dental implant restoration cases processed between August 2017 and November 2022: Case ID: 60772314 60822385 60812506 60246730 60342214 60325436 60239491 60000163 60000228 60000253 60000263 60000271 60000295 60000314 60000335 60000341 60000357 60000362 60000365 60000366 60000388 60000398 60000503 60000525 60000529 60000551 60000581 60000582 60000583 60000610 60000616 60000631 60000633 60000646 60000670 60000697 60000700 60000707 60000717 60000718 60000723 60000745 60000749 60000760 60000787 60000798 60000802 60000807 60000808 60000811 60000813 60000817 60000823 60000836 60000848 60000850 60000855 60000877 60000890 60000903 60000926 60000927 60000945 60000968 60000982 60000997 60001014 60001015 60001016 60001037 60001040 60001059 60001062 60001079 60001081 60001096 60001106 60001117 60001127 60001130 60001137 60001151 60001160 60001171 60001174 60001182 60001187 60001194 60001200 60001209 60001222 60006583 60001245 60001267 60001277 60001287 60001289 60001291 60001298 60001305 60001313 60001332 60001344 60001347 60001366 60001368 60001371 60001372 60001387 60001391 60001410 60001416 60001427 60001428 60001452 60001466 60001480 60001482 60001501 60001547 60001550 60001555 60001557 60001576 60001580 60001592 60001594 60001598 60001599 60001612 60001614 60001621 60001628 60001634 60001674 60001700 60003149 60001706 60001720 60001725 60001726 60001742 60001762 60001815 60001865 60001872 60001882 60001898 60003195 60001905 60001917 60001920 60001931 60001933 60001955 60001964 60001967 60001999 60002010 60002011 60002012 60002017 60002018 60002034 60002038 60002048 60002067 60002074 60002085 60002122 60002131 60002133 60002136 60002138 60002160 60002176 60002189 60002191 60002223 60002224 60002241 60002270 60002274 60002276 60002282 60002291 60002295 60002296 60002301 60002320 60002321 60002365 60002368 60002369 60003188 60002389 60002422 60002423 60002452 60002460 60002481 60002499 60002526 60002543 60002595 60002607 60002610 60002623 60002633 60002656 60002706 60002758 60002761 60002787 60002985 60002987 60003013 60003060 60003113 60003163 60003164 60003175 60003184 60003223 60003229 60003258 60003274 60003293 60003332 60003337 60003338 60003347 60003366 60003376 60003456 60003466 60003532 60003534 60003535 60003573 60003795 60003829 60003830 60003835 60003838 60003910 60004130 60004184 60004192 60004347 60004348 60004420 60004441 60004560 60004790 60004809 60004873 60004877 60004878 60005105 60005127 60005148 60005176 60005212 60005244 60005249 60005394 60005406 60005469 60005539 60018810 60005724 60005751 60005824 60006092 60006122 60006131 60006402 60006684 60006685 60006736 60007022 60007240 60007290 60007293 60007301 60007305 60007569 60007610 60007832 60008069 60008071 60008119 60019181 60008121 60008307 60008315 60008326 60008370 60008404 60008498 60008631 60008835 60008947 60009098 60009342 60009421 60009450 60009665 60009870 60009933 60009934 60009954 60009957 60010592 60011187 60011202 60011255 60011256 60011278 60011292 60011688 60012058 60012231 60012284 60012364 60012671 60012982 60013206 60013864 60013888 60014174 60014197 60014213 60014305 60014543 60014580 60014582 60014635 60015299 60015365 60015427 60015912 60015990 60016096 60016548 60016759 60016818 60016970 60017199 60017338 60037585 60017796 60018272 60018601 60019120 60019591 60019655 60020261 60020337 60020750 60020850 60020852 60021485 60021503 60023337 60023844 60023918 60024241 60025298 60026536 60027408 60027476 60028168 60028996 60029014 60030647 60031478 60031505 60032629 60032643 60033211 60034036 60034037 60034053 60034691 60035074 60035677 60037118 60037164 60037221 60038380 60039127 60039139 60039194 60040128 60042869 60043521 60043559 60043926 60044573 60044638 60044740 60045671 60045815 60047577 60047629 60047806 60048267 60048745 60048860 60048924 60049142 60049650 60052011 60052577 60053624 60054418 60054503 60054509 60054941 60056748 60056983 60057047 60057817 60057883 60454880 60057895 60057990 60059220 60059840 60059845 60061057 60062524 60062814 60063159 60063299 60065305 60065547 60065812 60066187 60066293 60066524 60066585 60066824 60067632 60068501 60068738 60068864 60069665 60069827 60070858 60071563 60071736 60071970 60072123 60073916 60073924 60074063 60076701 60077280 60077390 60078656 60079542 60080667 60081614 60081775 60082705 60082970 60084062 60084979 60085159 60085178 60087865 60088524 60088799 60089539 60089553 60089610 60089637 60091183 60091227 60093808 60093960 60094235 60137762 60095171 60095182 60095244 60095380 60095493 60096818 60096827 60096837 60096852 60097523 60097632 60097651 60099978 60100659 60100667 60101981 60103433 60105218 60105632 60105652 60107722 60107971 60108150 60108540 60109706 60109790 60113465 60114893 60115476 60116475 60116572 60116590 60209947 60117025 60117829 60118187 60118192 60118249 60118312 60119005 60119626 60119660 60119702 60121452 60122814 60122866 60123031 60125622 60125809 60127048 60127090 60127139 60128746 60128846 60129750 60129761 60131498 60131506 60131543 60132968 60134276 60134332 60134558 60134911 60134949 60134993 60136050 60136156 60137550 60138665 60138968 60139010 60139363 60139559 60139586 60139870 60140438 60140772 60176076 60140862 60141930 60142184 60142204 60142214 60142502 60142905 60143666 60143702 60143708 60143766 60144058 60144587 60144944 60145062 60145070 60145393 60145402 60146118 60146156 60146222 60146246 60146325 60146369 60146407 60146500 60146509 60147217 60147743 60147747 60147752 60147904 60147929 60149013 60149043 60149300 60149651 60149681 60150939 60150958 60151104 60151112 60151115 60151119 60151912 60151954 60152211 60152279 60153641 60153997 60154092 60154982 60155093 60155120 60155360 60155441 60156792 60157103 60157121 60158101 60158107 60158525 60158640 60159820 60160013 60225171 60160047 60160932 60161033 60161569 60161624 60161658 60161673 60163098 60164061 60164079 60164088 60164448 60164525 60165851 60165881 60165963 60166293 60167411 60167671 60167780 60167784 60168878 60169099 60169627 60170905 60170981 60170983 60171000 60171029 60171391 60171682 60172758 60172797 60172803 60173682 60174507 60174586 60174641 60174760 60175565 60175602 60175615 60175924 60175980 60176456 60176475 60177043 60177081 60177134 60234882 60177552 60177560 60177929 60177959 60178537 60744250 60178681 60178988 60180188 60180204 60181023 60181107 60181953 60182332 60182345 60182394 60182647 60182769 60184205 60184691 60184698 60185447 60185649 60185934 60187361 60187377 60187381 60187561 60187847 60189157 60189179 60189186 60189238 60189770 60189853 60189881 60190667 60191546 60191614 60193079 60194676 60194741 60194766 60194803 60196045 60196878 60196885 60298005 60196952 60197954 60198698 60199606 60199629 60200065 60200109 60200612 60200777 60200831 60200862 60202237 60202268 60202330 60203815 60203851 60204005 60204102 60204131 60204178 60205353 60205623 60205865 60206222 60206238 60206247 60206291 60207595 60208069 60208099 60209288 60209512 60210095 60211556 60211837 60211921 60211946 60211958 60213733 60214076 60214098 60214133 60215283 60215441 60215577 60215959 60215968 60215985 60216032 60216039 60217539 60218089 60218109 60219561 60219644 60219665 60220450 60221529 60222309 60222457 60223131 60224225 60224420 60224447 60224454 60225730 60225762 60226124 60226271 60226775 60228066 60228517 60228695 60228754 60230113 60230271 60230282 60230583 60230909 60231117 60231176 60231201 60232423 60232478 60283165 60232716 60232921 60233441 60233595 60234661 60234683 60234728 60234770 60235741 60237407 60237531 60238013 60597515 60238070 60238281 60239422 60305398 60240175 60240387 60241474 60241590 60241603 60242387 60243932 60246087 60246366 60249305 60249709 60250698 60250885 60251702 60251725 60253107 60253201 60253246 60253758 60253776 60253970 60254034 60255687 60255898 60256363 60256843 60257501 60261281 60261411 60261459 60263365 60263731 60263732 60265089 60265103 60265462 60268135 60269165 60269460 60269491 60269520 60269539 60270232 60270528 60270579 60270730 60270825 60270957 60271315 60275755 60276013 60276569 60276618 60278311 60278549 60278628 60278756 60278802 60278994 60363580 60279496 60280085 60280638 60281131 60281151 60281230 60281243 60281310 60281658 60282243 60282616 60282971 60282973 60283011 60283447 60283498 60284794 60285130 60285804 60285945 60286318 60287354 60290036 60290049 60290942 60290962 60290972 60291124 60292980 60293195 60293503 60295120 60541467 60298809 60298891 60299124 60299501 60299581 60299940 60300011 60300257 60301713 60301958 60302071 60347105 60302094 60302159 60302858 60304164 60306260 60307337 60307434 60308401 60308637 60308790 60484102 60309697 60310025 60310188 60310692 60310793 60311543 60311699 60311757 60392087 60501636 60670623 60313582 60314250 60315215 60315233 60316437 60317643 60318350 60318774 60318921 60319270 60319984 60322505 60323848 60324092 60326218 60327313 60327610 60327666 60327789 60327909 60328383 60328511 60329502 60329585 60329657 60329681 60329702 60329867 60331277 60331617 60331671 60331712 60331756 60333338 60335628 60335954 60336165 60336891 60337447 60338145 60338545 60338844 60338945 60339459 60339515 60339873 60340845 60341115 60484851 60342452 60647242 60727696 60839973 60918276 61168390 61235532 60343214 60343266 60343309 60343887 60344225 60344655 60345256 60345388 60345486 60347219 60397500 60347311 60347656 60347853 60350333 60350997 60351017 60352086 60352104 60352210 60352761 60353015 60354856 60355648 60356051 60356854 60357004 60358034 60359545 60361562 60364068 60364191 60365674 60406976 60365704 60366680 60366880 60367531 60367625 60367927 60367954 60368936 60369379 60369536 60369683 60369714 60370317 60370396 60372157 60372552 60372645 60372729 60372891 60373242 60373900 60374019 60374797 60374948 60375050 60375739 60376999 60377518 60377601 60379413 60380068 60380123 60382053 60382310 60382393 60382423 60382523 60382573 60383919 60383955 60384739 60384898 60447845 60745174 60385040 60385095 60385774 60386524 60389357 60389393 60389817 60389909 60390321 60390839 60390979 60391270 60393869 60394196 60395482 60395754 60396520 60396558 60396603 60398854 60399267 60399293 60399502 60399824 60399934 60400535 60404350 60404971 60405089 60405115 60405178 60407530 60407566 60408401 60409421 60409944 60410505 60410621 60410643 60410659 60411565 60412037 60413438 60413576 60413597 60413663 60414255 60766422 60415902 60418081 60419016 60421497 60422747 60423395 60423791 60424624 60424728 60424856 60426007 60426022 60426066 60427388 60429040 60430024 60431288 60432472 60436187 60436866 60438315 60438446 60438464 60439137 60439377 60439474 60439500 60440630 60441128 60441637 60441821 60442376 60442532 60443745 60444309 60444366 60444377 60444399 60444401 60444407 60444418 60445018 60446932 60448846 60448880 60449196 60450054 60450241 60450528 60453079 60453427 60453522 60453613 60453720 60454144 60454197 60454228 60456299 60456405 60457513 60459184 60460729 60462417 60463818 60465291 60466060 60466705 60467333 60467354 60467406 60467683 60467784 60469401 60470921 60473576 60473588 60473769 60473865 60474126 60477146 60477289 60477440 60477458 60477582 60477659 60477693 60478381 60479373 60479739 60479762 61380879 60480301 60481519 60482401 60483017 60483659 60483699 60483999 60484239 60484384 60488038 60488052 60490717 60491114 60492540 60493245 60494259 60495817 60496159 60496216 60651154 60496995 60497131 60498587 60498642 60498956 60499275 60501840 60501866 60502025 60502657 60502679 60503181 60503464 60503488 60504505 60504702 60625416 61173640 60507794 60507927 60507983 60508005 60509584 60510595 60510696 60510765 60510822 60510948 60511217 60856942 60511596 60511607 60511775 60511816 60511857 60511882 60512167 60512320 60512498 60513508 60513611 60513698 60515002 60515041 60515430 60515448 60515483 60517281 60518039 60518088 60518552 60518824 60518832 60520254 60520808 60521535 60619531 60577098 60521997 60522097 60523017 60523171 60523716 60524487 60526064 60526259 60586342 60527703 60528203 60529967 60531187 60531218 60533233 60533870 60533944 60534721 60536441 60536561 60538833 60538929 60538942 60539959 60540133 60540147 60541657 60541747 60541776 60541789 60542228 60542491 60543025 60543110 60543114 60543394 60543788 60544380 60544438 60546149 60547299 60547394 60547426 60547645 60547775 60548145 60548831 60548840 60548859 60548866 60549745 60549898 60550545 60551092 60551902 60551911 60551915 60551963 60552845 60553374 60619470 60554328 60554370 60554442 60554816 60554829 60554851 60554857 60615343 60556151 60556739 60556756 60557714 60557849 60558752 60558832 60558918 60559769 60560201 60560355 60560373 60561235 60562730 60562842 60562860 60563259 60563438 60563689 60563708 60563783 60563793 60563965 60564148 60564163 60564272 60564345 60564588 60564598 60564712 60565002 60565341 60601713 60565501 60565787 60566401 60566454 60566462 60566552 60566741 60566759 60567157 60568087 60568680 60568688 60570534 60571236 60572098 60572238 60572895 60628838 60572918 60574591 60574643 60574653 60575131 60575277 60577005 60852065 60577140 60578018 61403850 60579319 60579380 60580477 60580793 60583319 60583604 60583816 60584870 60584932 60586015 60586671 60586685 60586958 60587101 60587202 60587233 60587317 60587332 60587630 60587758 60588138 60588280 60665789 60590196 60590475 60590508 60590602 60590707 60591243 60591259 60591532 60591554 60591620 60592102 60593555 60594388 60595219 60595237 60595422 60595956 60596077 60596469 60596545 60596683 60596895 60596974 60597229 60597297 60597506 60597716 60597918 60598014 60599572 60599589 60599874 60599904 60599942 60600352 60600549 60600583 60600768 60602115 60602352 60603310 60603484 60604160 60604281 60604546 60605469 60605569 60605582 60606302 60606635 60607380 60607802 60607930 60608320 60610290 60610305 60611066 60611603 60611746 60611793 60613152 60613272 60613548 60614006 60614092 60614755 60615917 60616809 60617179 60619249 60620965 60622808 60622864 60623048 60623081 60623149 60623205 60623217 60623267 60623500 60623519 60623534 60623672 60623677 60703136 60625137 60695503 60625275 60625373 61020563 60625924 60625952 60625982 60626072 60626218 60626508 60626964 60627256 60627372 60627392 60628821 60628909 60628995 60629070 60629660 60630247 60630387 60630451 60631036 60631109 60719271 60631760 60632747 60633441 60634307 60749945 60634791 60634938 60634994 60635590 60635616 60635705 60636696 60637994 60638034 60638048 60638175 60639167 60639520 60641008 60641030 60731629 60641186 60727030 60641421 60642464 60642587 60642801 60643064 60643169 60643246 60643421 60643594 60643859 60643879 60644092 60644252 60644832 60645550 60646357 60646659 60646674 60646914 60646976 60647587 60647704 60648241 60648332 60648386 60648776 60648883 60648951 60650593 60651128 60651322 60651349 60651565 60651916 60652082 60652479 60652579 60653009 60653757 60654320 60778767 60654414 60655777 60655803 60655857 60656534 61705436 60656774 60657212 60658002 60658135 60658262 60721041 60658379 60658522 60659052 60659073 60659548 60659579 60659773 60746812 60660264 60660322 60660796 60661236 60661845 60663029 60663117 60663553 60664857 60665007 60665200 60666406 60666915 60667291 60667680 60668652 60668724 60669223 60669287 60669612 60669639 60669844 60670611 60670972 60670993 60671123 60671242 60671326 60671422 60671781 60671846 60672365 60672610 60672621 60673335 60927143 61330853 60673760 60806449 60674083 60674209 60675132 60675817 60675850 60676133 60676455 60676602 60676825 60676871 60676934 60677326 60678346 60678364 60678374 60678710 60743158 60679543 60680493 60681261 60681537 60681630 60681663 60682033 60682116 60682920 60682956 60682990 60683448 60683495 60684355 60685131 60686598 60762378 60686754 60687863 60689106 60690294 60690462 60690792 60690907 60690927 60692876 60693065 60693754 60694081 60694710 60696053 60696281 60696738 60696761 60697275 60697319 60698373 60698506 60698519 61395964 60699257 60699335 60699490 60700255 60700293 60700314 60700318 60700639 60700682 60760310 60700954 60701263 60702112 60703183 60703439 60704050 60705235 60705268 60705510 60707648 60813472 60882887 60708445 60708453 60709618 60710722 60711401 60711437 60712030 60712650 60714301 60714447 60714859 60715159 60715531 60715557 60715596 60716233 60716554 60716604 60716820 60717426 60717463 60717500 60717730 60719578 60720053 60720338 60720633 60721122 60722303 60722411 60722761 60722777 60725418 60725424 60739546 60726280 60726977 60728664 60729046 60729226 60729239 60729630 60729637 60730301 60730303 60770673 60730676 60731098 60731107 60731114 60731142 60731395 60731403 60731451 60731503 60731747 60731831 60731922 60731933 60731941 60732107 60732111 60732341 60732719 60733111 60733130 60733270 60733417 60733623 60733631 60733640 60733839 60734129 60734166 60734185 60735310 60735490 60735494 60735496 60735506 60736003 60736134 60798095 60737327 60737422 60923838 60737674 60738237 60739104 60739211 60739231 60739273 60739349 60739366 60739422 60740021 60813915 60740042 60740060 60740161 60740386 60740498 60740893 60741124 60741161 60825717 60741188 60741221 60741716 60741750 60741759 60741773 60742209 60742301 60742307 60742619 60743553 60743627 60743747 60743754 60743779 60744064 60744103 60744207 60744262 60744267 60744551 60745253 60745728 60800756 61121653 61455866 61605366 60745772 60745774 60745777 60745792 60745834 60746297 60746707 60746792 60746797 60747277 60747460 60747493 60747494 60747520 60747547 60747730 60748057 60748141 60748231 60748926 60749023 60749169 60749203 60749210 60749232 60882004 60966204 60749302 60749640 60749709 60750046 60750488 60750523 60750735 60750757 60750784 60750962 60751236 60751244 60751661 60751678 60751712 60751716 60751752 60751770 60751939 60752254 60752300 60752337 60753157 60753201 60753206 60753210 60753499 60753561 60753606 60753624 60753793 60753796 60753810 60753984 60754041 60754362 60754412 60754431 60754454 60754676 60754867 60755081 60755771 60755799 60756221 60756262 60756504 60756946 60799292 60757734 60758101 60758293 60758364 60758579 60758584 60758664 60758675 60759022 60759209 60759399 60759413 60759420 60759513 60759534 60759535 60759845 60760061 60760109 60760355 60760372 60761063 60761105 60761820 60761827 60761974 60762380 60762398 60762746 60762751 60762942 60763420 60800426 60763590 60763653 60763912 60764092 60764101 60764266 60764538 60764630 60764661 60764665 60764828 60765567 60765678 60765684 60765904 60765911 60765924 60765945 60765969 60766043 60766241 60766303 60766332 60766440 61460124 60766444 60766630 60766718 60766785 60767249 60767410 60767451 60829883 60767473 60768881 60769222 60769330 60769662 60875982 60769667 60769832 60770892 60771014 60771077 60771494 60771531 60771789 60772358 60772490 60773297 60773449 60774120 61202698 60774352 60774482 60774546 60774601 60774637 60774971 60775012 60819532 60775244 61959575 60775419 60775420 60775448 60775719 60776278 60776426 60776450 60776734 60776950 60777273 60777675 60777887 60777986 60778147 60778149 60778452 60778611 60778674 60779084 60779180 60779812 60780738 60781111 60781180 60781463 60782065 60782118 60782134 60782214 60782633 60782993 60783040 60783364 60783411 60783432 60980668 60783871 60783905 60784612 60863030 61010777 60785045 60785580 60785593 60785612 60785616 60785626 60785977 60785991 60786083 60786265 60786379 60786458 60786483 60786571 60786838 60786955 60786986 60787179 60787225 60787267 60787382 60787419 60787474 60787820 60788215 60788361 60788451 60788560 60788728 60836546 60788914 60789033 60855507 60789108 60789123 60789197 60789592 60789843 60790542 60790562 60790783 60790821 60790853 60790855 60791067 60791090 60791121 60791135 60791241 60791607 60791644 60792346 60792411 60792858 60792954 60793224 60793308 60793411 60793468 60873619 60793991 60794060 60794066 60794591 60836875 60795515 60795568 60795855 60795876 60796069 60796260 60796269 60797186 60797405 60797464 60797515 60798026 60798170 60798301 60798754 60798812 60799153 60799156 60799160 60799280 60852806 61107772 61230928 60800630 60800876 60800929 60801369 60802004 60802064 60802108 60802516 60802523 60802699 60804377 60804432 60804552 60804714 60804717 60804730 60805085 60806430 60806457 60806464 60945516 60806575 60807083 60808046 60808289 60808864 60808922 60809043 60809335 60809625 60809641 61050372 60809763 60809843 60810137 60810385 60810463 60810808 60811160 60811511 60812233 60812297 60812590 60813035 60813437 60813765 60813789 60813860 60814313 60814693 60814753 60814941 60815086 60815096 60815111 60815983 60816032 60816266 60816538 60816744 60816772 60816793 60816804 60818576 60818595 60819704 60819714 60819725 60819756 60820022 60820113 60820128 60820144 60820169 60820373 60820826 60821105 60821295 60822182 61892142 60822579 60823566 60823796 60823909 60823936 60823947 60823962 60823970 60823983 60823984 60824170 60824236 60824921 60825391 60825577 60825588 60825756 60827322 60827326 60827721 60827757 60828794 60828823 60828925 60828956 60828998 60829346 60829414 60829446 60829483 60829535 60918162 60829598 60829876 60829973 60830046 60830081 60830091 60831042 60831570 60831638 60831655 60831907 60832519 60832554 60832820 60833035 60833117 60833604 60833821 60833964 60834021 60834035 60834052 60834233 60834244 60834274 60834322 60835176 60835225 60835394 60835796 60891198 60835832 60835861 60836579 60836865 60837152 60837382 60837678 60837696 60838230 60838263 60838551 60838830 60838928 60839497 60839522 60955288 60839775 60839823 60840665 60841935 60842114 60842130 60842652 60843361 60843386 60843443 60843552 60843569 60843589 60843602 60843838 60844102 60844129 60844162 60844404 60844573 60845151 60845183 61574033 60845256 60934064 60845290 60845318 60845365 60846931 60846957 60847120 60847471 60847685 60848209 60848563 60849183 60849273 60849297 60849321 60849363 60849861 60850133 60850976 60851095 60851314 60851328 60851534 61109308 60851963 60853006 60853047 60853213 60853420 60853815 60853841 60853909 60854015 60952613 60854407 60855456 60855551 60855626 60855646 60855771 60855848 60855875 60856174 60856431 60856971 60857961 60857962 60858182 60858218 60858344 60858799 60858826 60859085 60859584 60860367 60860374 60860390 60860408 60860472 60860589 60928468 60860649 60860941 60862425 60862841 60862922 60862975 60863718 60863932 60968694 60864226 60865305 60865352 60865888 60866340 60866433 60866484 60866798 60867119 60867141 60867649 60867900 60868069 60868584 60869633 60870028 60870504 60870860 60870884 60870893 60871172 60872374 60872406 60872527 60872976 60873301 60874100 60874407 60874483 60875130 60875613 60875627 60875647 60875679 60875806 60875858 60875876 60875955 60875969 60875978 60876212 60876288 60876295 60876309 60876328 61045515 60876608 60877330 60878114 60933972 60878321 60878358 60880505 60880855 60880988 60882251 60883269 60883358 60884945 60885415 60885602 60885798 60885964 60886042 60886234 60886255 60886681 60886765 60886981 60886997 60934031 60887034 60887429 60888707 60888727 60889154 60889196 60889326 60890454 60891040 60891706 60892488 61285208 60893277 60893436 60893595 60893933 60895045 60895442 60895531 60895592 60895690 60897527 60897652 60897996 60898152 60898196 60898964 60900695 60900924 60901143 60901458 60902269 60902580 60902951 60903072 60903146 60903893 60903927 60904048 60905698 60905778 60906061 60906881 60906892 60906940 60907464 60907517 60907522 60907539 60907719 60908293 60908640 60908654 60908902 60909221 60910577 60910765 60910852 60910998 60911188 61048231 61603745 61683534 60911697 60912777 60914303 60914315 60914326 60914347 60914392 60914795 60915095 60915101 60915231 60915716 60915795 60915818 61019818 60916245 60916259 60916565 60916596 60918099 60918369 60918402 60918593 60919819 60976751 60919993 60921453 60921715 60922867 60922886 60922996 60923317 60923636 60923672 60923912 60923936 60924362 60925127 60925229 60925249 60925852 60926311 60926582 60927554 60927599 60927885 60928348 60928353 60928440 60928513 60929295 60929651 60929733 60929814 60929826 60929841 60930183 60930422 60930482 60931059 60931682 60931696 60931699 60932173 60933563 60933694 60934282 60934488 60934838 60935302 60935618 60935757 60935990 60936046 60936076 60936124 60937335 60938682 60938718 60938960 60939971 60940187 60940357 60994979 60940732 60940802 60941636 60941761 60942017 60942116 60942303 60942444 60942514 60943272 60943853 60944477 60944966 60945579 60945712 60946011 60946075 60946384 60946397 60946475 60946554 60946961 60947206 60947572 60947596 60947649 60948111 60948404 60949495 60949905 60950246 60950281 60950520 60951261 60951469 60952545 60953077 60953317 60953414 60953478 60953731 60953907 60955191 60955194 60955710 60955780 61266054 60956035 60956831 60958099 60958202 60958462 60958552 60958574 60959381 60959475 60960087 60960335 60960375 60960426 60960858 60961027 61078612 60961537 60963876 60964191 60964238 60964437 60964970 60964990 60965133 60965686 60965700 60967021 60967086 60967478 60967699 60967933 60970848 60971352 60973530 60973720 60974190 60974407 60974423 60974456 60974526 60974652 60974794 60975010 60975590 60976072 60976947 60978908 60979587 60980082 60980127 60980135 60980786 60981168 60981642 60985406 61033274 60985636 60986018 60986102 60987168 60987534 60987817 60987833 60987842 60987855 60987910 60989916 60990473 60990585 60990760 60991260 60991498 60991607 60991661 60992182 60992225 60992723 60992745 60992779 60993683 61050856 61108056 60994810 60995666 60997046 60997061 60997318 60997549 60998406 61001738 61002469 61003033 61003055 61004227 61005811 61005839 61006175 61006709 61007776 61008186 61008217 61008418 61009299 61009316 61009729 61463793 61011394 61081913 61011441 61011458 61012686 61013015 61013069 61013629 61016019 61016509 61018186 61018717 61019194 61020241 61114084 61021867 61022947 61023291 61023526 61025017 61025565 61025693 61025834 61025848 61027949 61028016 61028111 61028153 61028181 61028321 61028657 61030527 61030883 61033029 61033357 61033480 61033848 61033982 62141707 61034263 61034491 61036059 61036093 61036111 61036839 61037089 61037479 61037551 61382301 61501365 61038442 61038543 61038624 61038943 61038961 61039007 61039730 61040248 61147051 61331200 61042436 61043915 61043989 61044838 61044923 61045036 61045305 61045332 61045351 61045457 61046274 61046441 61046544 61046672 61047334 61047937 61047944 61049634 61049881 61049921 61049948 61050039 61052397 61052410 61052422 61052774 61052999 61053226 61053281 61053304 61053373 61053580 61054398 61054860 61055065 61055580 61055701 61056019 61056040 61056403 61057697 61057783 61058452 61059125 61061774 61061983 61062023 61062206 61062517 61064151 61065246 61066444 61066930 61067286 61067376 61067392 61067713 61067870 61067897 61067939 61067962 61067965 61068637 61069026 61069215 61069292 61069312 61069703 61069820 61069848 61070544 61070584 61071044 61095408 61072344 61072540 61072549 61072664 61072698 61075116 61075136 61075290 61075313 61075335 61075359 61075403 61075483 61075563 61076758 61077546 61077557 61077665 61078402 62123493 61078508 61078720 61078763 61079037 61152688 61079206 61079546 61079592 61079802 61080114 61080626 61080722 61080757 61080839 61081453 61084061 61084486 61084574 61084675 61085033 61085303 61085457 61085745 61085805 61085919 61086645 61086776 61086920 61087250 61087331 61087368 61087389 61088252 61089239 61090796 61090886 61091576 61091591 61091633 61092584 61093382 61093551 61093640 61094207 61094445 61095095 61095240 61095416 61096513 61097205 61097422 61097710 61161749 61097793 61097814 61097909 61099250 61099989 61100033 61100147 61100240 61100265 61100326 61101186 61101474 61102019 61102046 61102199 61102303 61102922 61102991 61103073 61103444 61105807 61105883 61105910 61106633 61106653 61106900 61106926 61107117 61107214 61107382 61107466 61107483 61107764 61109286 61109299 61109327 61109423 61109444 61109464 61109622 61110238 61110275 61110545 61110958 61111426 61111662 61111762 61160380 61112787 61112836 61113016 61113110 61115113 61115925 61116006 61116699 61117135 61117205 61117648 61117661 61118783 61118794 61119453 61119577 61120988 61121027 61121774 61123300 61123416 61123667 61202618 61124485 61434864 61125201 61125255 61125292 61126467 61127261 61127518 61127729 61127748 61128337 61128372 61128573 61128759 61128787 61265953 61128839 61129811 61129896 61130064 61131088 61131124 61131140 61131431 61132659 61212240 61132782 61133167 61133412 61133863 61134548 61134788 61134809 61134943 61136184 61136515 61137203 61137248 61137740 61137952 61137972 61138004 61138378 61139046 61142074 61142124 61143750 61143958 61144313 61145312 61145575 61145631 61145767 61146814 61146877 61147126 61148035 61148099 61148188 61148409 61148718 61149865 61150860 61150889 61150915 61151508 61151537 61151834 61152266 61152510 61152538 61153642 61153720 61153835 61155028 61244897 61155270 61155687 61155694 61155808 61156163 61156462 61156538 61157031 61158023 61158145 61589267 61158411 61158695 61159443 61159529 61159580 61300622 61160092 61160142 61160727 61160911 61161219 61161312 61162252 61162265 61162338 61163374 61164048 61165165 61165271 61165337 61250898 61165796 61165812 61165878 61165927 61166663 61166781 61166801 61166859 61167104 61167174 61167751 61167893 61167988 61169357 61169511 61169600 61170063 61267700 61170258 61170548 61171010 61171067 61485942 61171209 61172233 61172494 61172610 61172684 61172856 61173256 61173746 61174510 61174980 61175404 61175618 61175897 61176419 61176933 61177269 61177332 61178164 61178377 61178841 61178865 61179573 61179596 61179613 61179803 61180554 61181666 61182583 61183184 61183548 61184077 61184208 61184312 61184375 61184383 61184420 61184470 61184500 61184503 61184581 61184832 61185019 61185764 61185785 61185845 61185857 61185877 61186033 61186428 61187026 61188875 61189475 61189495 61189521 61189536 61190071 61191615 61191730 61253855 61357693 61424166 61485768 61191873 61191961 61192315 61192419 61192430 61192601 61192608 61192722 61253485 61192934 61193148 61194364 61194443 61194472 61194840 61195422 61195709 61195812 61196240 61196629 61196895 61197241 61198075 61198570 61199151 61199180 61199424 61199434 61199490 61199592 61199672 61199812 61200557 61201239 61201279 61201301 61201505 61201542 61201554 61202081 61202320 61202602 61202760 61202841 61202867 61203204 61203335 61203574 61203579 61203597 61203917 61204125 61204788 61205157 61205243 61205420 61205473 61207574 61207589 61207716 61207780 61207862 61207988 61208879 61208915 61209046 61209511 61210936 61210945 61211051 61211134 61212314 61213517 61213714 61213753 61213830 61214017 61214077 61214497 61214568 61214596 61214663 61214871 61215424 61215575 61215629 61215884 61216109 61217803 61218214 61218609 61218717 61218762 61218790 61218838 61219429 61220446 61220618 61220837 61220943 61220955 61221244 61221557 61221627 61221905 61221917 61222484 61222991 61223015 61223404 61223408 61223438 61223483 61223492 61223741 61223938 61223990 61224017 61224030 61224729 61224864 61225422 61226030 61226385 61226706 61226962 61227353 61227645 61227770 61228582 61228963 61229026 61229064 61229186 61229586 61229636 61230007 61230153 61230650 61231358 61231996 61232340 61232525 61232554 61233180 61234052 61234718 61656974 61235209 61235234 61235291 61235308 61235332 61235355 61235502 61236016 61236038 61236225 61237657 61237695 61237712 61237751 61237835 61238553 61238577 61238730 61355096 61238761 61239553 61240180 61240649 61241111 61241137 61241381 61241393 61241489 61241780 61241895 61241991 61242183 61242241 61242770 61242811 61243693 61243856 61243960 61244002 61244360 61245061 61246524 61246532 61246633 61246847 61246938 61247000 61247044 61248375 61248391 61248897 61249208 61249234 61249476 61249690 61249801 61249839 61249856 61250432 61250536 61250550 61250693 61251219 61251680 61252199 61252326 61252552 61252576 61252831 61252880 61253265 61253453 61253910 61254795 61254835 61255644 61256021 61256589 61256599 61256896 61256928 61258781 61258912 61258975 61260090 61260249 61260380 61260603 61260760 61261150 61261609 61262266 61262692 61262872 61262941 61263046 61263424 61263576 61264020 61264047 61264292 61264344 61264854 61265013 61265155 61265172 61265348 61338897 61265372 61265397 61265408 61265420 61265426 61265437 61265447 61266177 61266186 61266195 61266285 61266718 61266838 61268417 61268683 61268742 61268784 61268817 61268870 61269469 61269646 61269812 61270199 61270225 61270922 61271737 61271825 61272019 61272033 61272298 61272801 61273019 61273065 61273444 61276711 61276795 61277216 61277275 61277334 61277547 61277731 61277951 61278093 61278173 61279613 61279737 61279751 61279764 61279769 61279782 61279787 61279793 61279803 61279805 61279829 61279865 61279933 61279957 61280016 61280096 61280216 61345974 61281843 61281860 61282576 61282607 61282664 61282682 61283030 61283069 61283549 61283736 61283758 61283816 61283848 61283892 61284158 61284348 61284607 61284693 61284720 61284916 61285002 61389084 61285243 61285493 61328725 61285862 61285961 61286125 61286141 61372761 61450100 61286377 61371793 61286766 61287095 61287675 61287725 61287756 61287780 61287814 61287823 61288244 61288250 61288271 61288290 61288308 61288359 61288380 61288400 61288460 61288862 61290353 61290411 61293222 61293231 61293528 61293855 61294167 61294350 61294456 61294710 61294922 61294947 61295278 61295307 61295826 61296642 61296658 61296761 61297361 61297642 61297719 61298006 61298587 61298616 61298674 61298703 61298769 61299132 61299262 61299624 61300632 61300663 61300747 61378933 61301768 61302458 61303689 61303837 61303902 61304046 61304058 61304179 61304328 61305176 61305657 61306935 61307328 61366148 61307463 61307890 61383252 61307977 61308217 61308241 61308674 61308883 61309059 61309420 61309585 61310172 61310262 61310431 61310492 61310526 61310567 61310659 61310722 61440196 61311423 61312058 61312082 61312130 61312447 61313183 61463146 61313225 61313263 61313422 61313433 61313818 61313927 61313975 61314139 61421247 61315389 61315567 61315658 61315733 61315750 61316770 61316881 61317094 61317318 61317866 61318658 61318710 61318799 61318839 61318862 61318866 61318874 61318886 61320454 61320546 61320564 61320720 61446152 61320731 61320816 61320838 61320854 61320910 61321276 61321297 61321412 61321777 61321960 61322395 61322810 61323551 61324485 61324865 61324942 61325093 61325294 61325296 61326179 61326625 61326681 61326850 61326877 61327241 61449408 61512847 61328023 61328121 61328307 61329347 61329597 61329922 61330590 61330787 61330807 61331035 61331122 61331142 61331213 61331553 61331681 61332203 61332603 61333098 61333217 61333382 61334665 61334841 61334885 61335123 61335142 61335360 61335521 61335617 61335736 61335911 61336854 61336900 61338435 61338591 61338690 61338956 61339234 61339923 61339989 61340030 61340568 61340798 61340891 61341066 61341962 61341982 61342113 61342178 61342270 61342894 61343368 61343640 61343702 61343814 61344381 61345137 61345656 61345857 61346231 61346270 61346286 61346575 61347044 61347344 61347364 61347548 61347570 61348249 61348275 61348557 61348562 61348727 61348987 61349067 61349125 61924426 61349157 61349254 61349785 61349948 61350048 61350065 61350510 61351581 61352165 61352208 61352221 61352303 61352415 61352525 61352693 61353012 61353344 61436894 61643923 62025745 61353768 61353977 61354824 61355111 61355858 61356002 61356176 61356358 61356659 61357774 61357910 61358043 61358239 61358393 61358447 61358484 61358896 61359005 61359023 61464367 61359710 61359726 61359750 61359757 61361024 61361472 61361574 61361753 61362610 61362849 61363967 61363972 61364566 61486454 61364610 61365084 61365511 61365587 61365704 61366063 61366107 61366278 61366304 61366411 61366598 61366599 61366834 61487448 61367532 61368653 61369085 61369232 61369372 61369445 61369517 61369630 61370046 61370235 61370261 61370515 61370584 61371258 61371577 61372105 61523728 61372596 61373067 61373140 61373141 61373373 61373437 61373468 61373580 61373629 61373638 61373700 61373775 61373801 61374049 61374325 61374405 61374546 61374566 61374664 61374703 61375769 61376223 61574497 61376281 61376332 61377065 61377583 61377609 61377631 61378170 61378419 61378653 61379207 61379405 61379545 61634167 61379875 61380159 61380241 61905342 61380350 61380434 61380468 61380724 61381102 61381184 61835412 61381244 61381294 61381328 61381593 61381909 61381920 61382351 61382409 61382481 61382794 61465311 61384051 61529815 61384371 61384430 61479373 61384865 61385251 61385574 61385629 61385795 61387022 61387105 61387138 61387757 61387893 61387943 61387992 61388018 61388041 61388092 61388174 61388671 61388719 61388861 61388890 61389438 61389863 61390060 61390503 61390828 61391055 61391593 61391614 61391782 61391951 61391967 61392059 61392230 61392387 61392590 61392924 61393424 61393665 61393729 61394385 61394651 61394722 61394796 61394807 61394998 61395388 61395582 61395943 61396021 61396128 62046147 61396849 61396913 61397445 61397754 61397986 61398088 61398130 61398160 61398444 61398753 61399334 61480700 61603591 61399617 61400111 61400177 61605367 61400824 61401190 61401220 61401297 61401317 61401358 61401556 61401699 61401824 61402123 61402279 61402292 61402648 61402794 61402808 61402815 61402823 61403393 61403460 61403681 61403701 61404974 61405456 61405530 61405762 61580229 62034283 62377065 61406112 61406240 61406435 61406490 61406725 61407149 61407172 61407217 61407253 61407300 61407610 61408277 61408383 61409349 61409725 61409875 61410140 61410161 61410763 61410801 61410928 61411224 61411229 61411858 61412429 61412442 61412655 61413273 61494807 61413615 61413744 61413788 61413953 61413995 61414045 61414538 61414777 61416382 61417462 61417478 61418146 61418156 61418506 61418636 61418976 61419024 61419091 61419412 61419473 61419643 61420476 61420498 61420579 61420595 61420636 61420739 61420863 61421104 61421119 61421267 61421524 61421613 61421918 61422041 61422071 61422192 61422205 61422391 61580208 61422405 61423153 61423442 61423616 61423704 61423765 61423819 61423929 61424008 61424125 61424158 61424239 61424240 61424336 61424350 61424588 61424653 61424719 61424797 61424926 61425017 61425047 61425335 61426719 61427743 61427906 61427948 61428081 61428117 61428158 61428184 61428344 61428424 61428445 61428466 61428651 61429724 61429966 61430098 61517299 61430129 61430322 61430585 61430601 61430629 61430934 61431318 61431335 61431369 61431436 61431544 61431573 61431840 61432444 61432557 61432768 61433018 61433497 61433691 61433754 61433852 61434185 61434193 61434309 61434344 61434357 61434387 61434517 61434749 61435014 61435116 61435283 61435790 61436014 61436086 61436307 61436450 61437182 61577775 61437332 61437343 61437391 61437492 61437631 61438184 61439407 61439451 61439511 61439544 61439553 61439866 61440300 61440858 61440865 61440904 61441023 61441396 61720995 61441868 61442074 61442170 61442198 61442422 61442715 61442739 61442979 61444306 61444360 61445096 61445132 61708185 61445327 61446167 61446266 61446336 61446879 61446895 61447472 61447551 61447637 61447717 61726514 61448071 61448271 61448278 61449011 61449096 61449207 61449473 61449566 61449847 61450110 61450148 61450163 61450204 61450232 61450251 61450273 61450297 61450363 61450387 61450655 61451202 61451234 61452552 61452560 61605664 61452686 61452768 61452872 61453074 61453124 61453286 61453680 61453736 61453761 61453778 61454892 61454931 61455069 61455189 61455205 61455333 61455523 61455537 61456345 61456491 61456655 61456695 61456929 61457426 61457569 61457607 61619586 61459074 61459111 61459415 61460730 61461049 61461104 61461167 61461199 61461289 61461403 61461559 61461681 61462696 61463242 61463320 61463350 61463437 61463466 61463526 61463620 61463744 61464809 61464881 61464947 61464979 61465147 61579785 61465196 61465740 61466519 61466547 61466629 61467101 61467168 61467192 61468389 61469148 61469290 61469398 61469814 61469849 61469868 61469957 61470003 61470063 61470075 61470113 61470135 61470164 61470334 61470637 61470757 61471694 61472014 61472064 61472092 61474970 61474975 61497257 61475010 61475020 61475316 61475432 61475482 61475672 61475692 61476238 61477083 61477288 61477431 61477560 61590204 61477595 61478054 61479419 61479525 61479734 61480720 61480727 61513010 61481388 61481394 61481865 61482045 61482342 61482959 61483114 61483457 61483689 61483719 61483812 61484193 61484399 61484647 61485483 61485600 61485670 61485676 61486246 61487327 61487357 61487534 61488331 61488419 61488560 61489195 61489391 61489417 61489929 61490011 61490059 61490568 61491226 61491289 61492444 61492740 61492926 61492958 61493013 61493146 61493291 61493293 61494323 61495286 61495444 61496135 61496558 61496709 61496758 61496769 61496936 61496944 61497391 61497408 61497570 61498039 61498403 61498464 61498847 61498955 61499139 61499143 61499325 61499807 61500167 61500454 61500729 61500893 61501231 61501458 61501698 61501914 61502276 61502308 61502425 61502521 61665839 61502792 61503546 61503831 61503911 61503930 61503937 61504220 61504289 61504313 61504671 61504700 61504804 61505233 61505273 61505456 61505479 61505856 61505950 61506094 61506115 61506191 61506566 61506797 61507080 61507510 61508235 61508282 61508418 61508463 61508489 61508566 61508692 61508704 61508855 61508928 61508959 61509163 61509596 61509672 61510072 61510478 61510580 61511545 61512360 61512443 61512707 61512724 61512810 61512954 61513299 61513565 61513605 61513687 61513915 61514074 61514192 61514662 61514937 61515291 61515736 61515781 61515798 61516122 61856927 61516738 61517849 61518019 61518031 61518043 61518058 61518298 61518351 61518505 61518958 61519589 61519758 61519769 61519875 61520031 61520332 61520352 61520493 61702870 61520605 61520616 61521235 61521522 61521897 61522023 61522867 61523101 61523196 61703185 61523348 61523472 61523587 61523629 61598567 61523630 61523700 61524076 61524102 61524167 61524271 61524364 61524516 61612099 61850206 62145231 62186426 61525158 61525442 61525453 61526533 61526567 61526709 61526742 61526867 61527060 61527077 61527445 61527452 61527860 61528533 62143821 61528545 61528709 61528918 61528949 61528956 61529128 61529947 61529972 61530469 61530508 61530845 61530931 61532591 61532993 61533064 61533071 61533152 61533153 61533274 61533302 61533636 61533640 61894126 61534145 61534302 61534597 61534784 61536313 61536588 61536658 61536729 61536803 61537210 61537348 61537699 61537726 61537887 61537905 61538053 61538109 61538347 61538770 61539085 61539876 61540319 61540330 61540468 61540486 61540839 61540999 61541315 61541360 61541389 61589319 61541411 61541537 61542246 61542270 61542366 61542554 61542613 61542634 61542654 61542809 61543288 61543497 61543523 61544050 61544062 61544063 61544355 61544502 61816922 61545467 61545680 61545805 61545824 61546665 61546859 61547140 61610833 61547271 61547431 61547640 61548286 61548396 61548504 61548525 61548583 61548769 61549325 61549536 61549558 61550015 61550093 61550133 61550144 61550253 61550433 61550508 61550511 61550678 61550865 61550955 61550987 61551463 61551532 61551968 61552014 61552150 61552326 61552451 61552490 61552491 61553497 61553535 61553858 61553886 61553968 61554229 61554463 61554634 61554783 61554958 61554972 61555558 61555572 61555611 61555865 61555899 61555990 61556903 61557607 61558029 61559202 61559525 61559673 61643674 61559699 61560025 61560029 61560114 61560126 61560268 61560843 61561287 61561290 61561470 61561567 61561644 61562058 61562282 61562509 61562510 61562551 62060026 61562914 61563120 61563254 61563564 61563984 61564071 61564326 61564550 61564648 61564802 61565020 61565138 61565275 61565411 61565521 61566256 61566500 61566511 61566745 61566915 61567675 61568903 61569050 61569515 61570050 61651490 61570098 61570178 61571402 61571411 61571442 61571493 61571539 61571549 61571581 61571938 61571954 61571997 61572295 61572409 61572899 61573013 61573343 61573873 61574109 61574713 61574822 61574834 61575884 61575917 61576154 61576216 61576450 61576508 61576528 61576530 61576651 61576775 61577618 61577687 61577795 61577801 61577828 61577884 61577955 61578002 61578176 61578195 61578248 61578286 61578432 61578754 61578755 61578962 61579472 61579701 61579943 61580049 61580125 61580185 61580345 61580473 61640674 61580717 61580721 61580738 61581676 61581736 61581768 61581824 61581871 61581936 61582097 61582222 61582306 61582539 61583464 61653588 61583554 61583663 61583710 61583726 61583901 61583989 61584292 61584433 61584966 61585161 61585248 61585321 61585350 61585540 61585848 61586246 61587610 61587694 61711005 61850434 61588179 61588208 61588229 61588284 61588312 61588332 61588372 61588419 61588765 61589061 61589236 61589890 61590411 61590488 61590717 61591104 61591217 61591352 61591463 61591545 61591650 61591711 61591760 61592523 61592694 61592703 61592819 61593483 61593576 61593629 61594014 61594854 61595604 61595697 61595797 61595854 61596185 61596312 61596834 61597367 61597633 61597677 61597846 61597875 61597936 61598041 61598114 61598165 61598168 61598687 61599135 61599183 61599279 61707741 61599305 61599326 61599523 61599529 61599684 61600349 61600531 61600663 61601059 61601183 61601402 61601577 61601815 61602070 61602128 61602175 61915972 61602318 61602631 61602643 61602653 61602867 61603232 61603378 61603668 61603776 61603945 61603977 61604238 61790848 61604324 61604444 61604672 61604837 61605499 61605568 61605699 61605905 61606099 61606849 61606872 61606997 61606999 61607052 61607811 61607907 61608179 61608284 61608676 61608908 61608966 61609198 61609389 61609409 61609481 61609823 61609870 61610218 61610337 61610340 61610716 61610725 61610744 61610892 61611034 61611194 61611427 61611468 61611513 61611603 61611944 61611959 61612333 61612370 61612498 61613185 61613203 61613236 61613292 61613497 61613534 61613670 61613941 61614130 61614255 61614435 61614516 61614889 61614938 61695956 61752987 61615174 61615789 61615845 61616611 61616964 61617425 61617551 61617569 61617850 61617884 61618166 61618190 61618559 61618587 61618663 61619385 61619509 61619999 61620384 61621070 61621088 61621213 61621303 61621312 61621384 61621655 61621705 61621805 61621912 61621997 61622527 61622583 61622795 61622853 61623528 61623664 61623691 61623937 61623991 61624027 61624352 61700858 61624879 61624960 61625047 61625299 61625800 61625839 61626009 61626436 61626587 61626643 61626647 61626713 61626825 61627010 61627018 61627299 61627334 61627428 61970687 61627601 61627885 61628004 61628049 61628060 61628192 61628489 61628586 61628818 61629287 61629900 61630323 61630342 61630362 61630475 61630880 61631408 61631827 61632693 61632751 61632910 61633132 61633265 61633275 61633622 61633957 61633987 61634014 61634061 61634106 61634560 61635049 61635520 61635620 61635745 61635897 61636471 61636511 61636532 61636745 61638076 61638314 61638340 61638476 61638674 61639332 61639598 61687187 61640059 61640592 61640652 61640681 61640828 61641177 61641486 61641614 61641682 61641784 61642214 61642941 61642978 61643152 61643155 61643156 61643567 61644068 61645346 61645872 61646119 61646613 61698817 61646809 61646888 61647646 61647855 61648472 61649018 61649020 61649118 61649712 61650273 61650295 61650326 61650548 61650811 61650909 61651116 61651452 61651656 61651692 61652067 61652408 61652414 61654810 61655062 61655095 61655320 61655339 61655514 61656392 61656486 61841599 61656586 61656742 61656746 61656913 61887852 61657013 61657234 61657593 61658021 61658673 61658966 61659483 61659737 61660177 61030064 61660309 61660384 61660872 61661007 61661047 61661166 61661445 61662008 61662388 61662561 61663674 61663993 61715270 61664021 61664080 61664857 61664953 61665340 61665359 61665590 61665974 61666131 61666522 61666660 61666714 61666973 61666997 61667211 61667387 61667620 61667782 61668190 61668569 61668664 61668979 61669013 61669587 61669620 61669860 61670120 61670870 61671157 61672066 61672234 61672692 61673068 61673139 61673435 61673483 61673632 61673638 61673844 61674213 61674361 61674435 61675239 61675282 61675422 61675733 61675944 61676094 61676730 61676999 61677502 61677698 61677763 61677832 61678022 61678150 61678497 61679098 61679105 61679218 61679231 61679272 61679340 61679416 61679531 61679817 61680215 61680308 61680563 61680702 61680721 61680797 61680916 61680996 61681250 61681474 61681540 61681827 61681988 62142948 61682642 61682809 61682835 61682888 61683179 61683212 61683250 61683301 61683740 61683834 61684542 61685216 61685321 61754938 61685478 61685895 61685905 61686118 61686603 61687109 61687153 61687368 61687516 61687662 61688384 61688473 61688815 61688824 61688947 61838278 61689449 61689499 61689777 61690725 61690768 61691419 61691795 61692416 61692607 61692919 61692956 61692973 61693608 61693739 61694127 61694721 61786420 61973653 61694762 61694854 61900871 62034332 62106743 61695024 61695210 61695803 61695816 61695893 61696107 61696251 61696565 61697055 61697076 61697082 61697101 61697140 61697224 61697619 61697724 61697917 61698373 61698481 61698801 61698840 61698984 61699175 61699266 61699363 61699904 61699960 61700736 61700765 61700937 61701005 61701331 61701936 61702225 61702233 61702388 61702554 61702723 61703265 61774948 61703321 61703330 61703352 61703423 61703564 61704038 61704215 61704230 61704335 61704362 61704449 61704595 61704739 61704744 61705106 61705215 61705434 61705525 61705727 61705901 61705974 61706127 61706364 61706375 61706516 61706606 61707137 61707146 61707581 61707733 61707921 61708362 61708427 61708486 61708921 61709970 61710032 61710044 61710152 61710357 61710401 61711065 61711980 61712324 61712738 61713489 61713506 61713516 61713545 61713809 61713878 61714043 61714504 61714623 61714901 61714976 61716768 61716816 61827773 61717098 61717149 61717257 61717280 61717471 61717910 61718105 61718118 61718413 61718670 61718856 61719125 61719188 61719207 61719308 61719974 61719982 61720410 61720555 61720594 61720669 61720874 61772643 61721205 61722118 61722136 61722270 61722699 61723323 61723514 61723523 61723706 61723753 61724406 61724640 61725360 61726346 61726535 61726736 61726827 61726839 61726967 61726985 61727017 61728277 61728820 61729065 61729204 61729783 61730241 61730426 61730606 61730874 61730894 61731090 61731110 61731672 61908159 61732129 61732197 61732386 61732428 61732628 61732648 61732852 61732879 61733150 61733263 61733419 61733740 61734004 61734064 61734348 61734431 61734532 61734765 61734898 61734931 61734953 61734997 61735363 61735385 61735744 61735763 61736274 61736476 61736480 61737164 61737434 61737806 61737906 61738351 61739008 61739057 61739254 61739332 61739534 61739557 61740172 61740280 61740328 61741200 61741209 61741519 61741532 61741699 61742058 61742060 61742492 61742524 61742579 61742968 61743062 61743212 61743652 61744151 61923127 61744205 61744212 61744221 61744346 61842564 61744503 61744589 61744648 61745271 61745273 61745399 61745549 61745875 61745890 61746107 61746348 61746379 61746958 61817127 61747440 61747534 61748379 61748393 61748704 61748813 61748862 61748929 61749585 61749740 61749934 61750096 61750203 61751282 61751637 61751931 61751979 61752003 61752049 61752077 61752131 61752175 61752185 61752402 61752422 61752980 61753523 61753665 61753678 61754311 61754365 61754458 61754666 61754672 61754692 61754864 61754902 61754917 61755015 61755114 61755913 61755942 61756264 61756467 61756602 61756674 61757215 61757402 61757714 61758193 61758279 61758405 61758945 61801215 61902837 61974246 62035825 62158208 61758993 61759245 61759940 61760598 61760785 61760871 61761037 61761075 61761135 61761208 61761235 61761427 61761458 61761609 61761667 61761794 61863514 61761834 61761874 61761930 61762265 61762457 61762770 61762775 61763190 61763219 61763563 61764467 61764487 61764588 61764647 61764719 61764937 61765174 61765198 61765511 61765525 61765893 61765965 61765977 61766040 61766129 61766345 61766659 61766738 61767778 61768015 61768040 61768479 61768684 61768996 61769730 61769758 61770094 61770168 61770682 61770805 61988952 62130032 62293399 61770883 61771185 61869145 61772399 61772487 61772768 61772867 61773409 61773439 61774416 61775244 61775260 61775274 61775588 61776395 61777453 61778071 61778081 61778533 61780072 61780133 61780412 61781033 61781493 61782316 61782379 61782437 61782585 61782652 61782746 61782793 61783141 61783220 61783323 61783501 61783513 61783562 61783588 61783712 61784622 61784907 61784964 61785495 61785648 61786238 61786429 61786756 61787045 61787256 61787267 61787651 61787691 61787702 61787932 61788170 61788902 61789163 61789257 61789944 61790336 61790371 61790516 61790763 61790887 61790890 61791139 61791331 61791439 61791457 61791553 61792118 61792129 61792156 61792423 61792427 61792703 61793041 61793129 61793842 61793861 61793864 61794002 61794572 61795696 61796847 61796859 61796901 61797157 61797342 61797438 61797962 61799679 61799748 61799782 61799817 61799888 61799957 61800021 61800428 61800578 61800909 61801010 61801184 61913220 61801469 61801516 61801569 61801722 61801724 61801748 61801754 61801769 61801889 61801913 61802184 61802212 61802273 61802362 61802438 61802485 61802971 61803440 61803758 61804001 61804140 61804281 61804316 61804401 62024757 61804939 61804949 61805029 61805221 61805234 61805277 61805386 61805503 61805634 61805714 61805755 61806214 61806578 61807685 61807706 61807836 61807979 61808075 61808219 61808308 61808367 61808467 61808872 61808890 61809891 61809909 61810167 61810212 61810407 61810583 61810642 61810880 61811011 61811710 61811743 61811776 61812368 61812388 61812580 61814286 61814506 61855746 61815219 61815877 61816320 61816332 61816487 61816751 61817589 61817800 61818007 61818058 61818140 61818484 61818506 61818613 61818648 61818799 61819559 61819756 61819990 61820013 61820605 61820715 61820824 61820841 61820874 61821160 61822120 61822134 61822210 61822222 61822309 61822632 61823229 61823271 61823404 61823456 61823816 61823927 61823951 61824073 61824454 61919817 61824749 61825351 61825696 61825728 61825741 61825773 61825992 61826021 61826155 61826554 61826555 61826633 61826868 61826890 61827237 61828445 61828482 61828503 61829303 61829477 61830366 61830377 61830851 61831041 61831059 61831440 61831474 61831611 61831817 61831850 61831906 61832011 61832023 61832121 61938375 61832173 61832183 61832737 61833052 61833135 61833194 61833299 61833310 61833727 61833975 61834057 61834079 61834277 61834659 61834823 61835034 61835752 61835769 61835792 61835802 61836575 61836644 61836994 61837004 61837123 61837323 61837372 61837639 61838252 61838371 61838485 61838676 61838908 61840226 61840396 61840468 61840532 61840596 61840636 61840660 61840675 61841093 61841982 61842824 61843297 61843643 61843855 61844455 61844774 61845087 61845732 61846010 61846052 61846151 61846249 61846561 61847354 61847434 61847446 61847573 61847637 61848220 61848387 61941901 61849147 61849486 61849582 61849850 61850267 61850814 61851068 61851235 61851781 61852105 61852199 61852593 61852743 61853178 61854896 61854960 61855051 61855284 61855916 61856353 61856398 61856613 61856867 61856879 61857115 61857979 61858766 61858824 61859615 61859877 61859891 61860336 61860707 61861234 61861264 61861541 61862042 61862092 61862510 61862574 61862601 61862620 61863040 61863477 61863764 61863848 61863865 61864254 61864268 61864328 61864708 61865773 61865983 61866343 61866448 61866622 61866677 61867801 61867838 61867912 61869135 61986247 61870369 61870402 61870511 61870523 61871407 61871430 61872082 61872100 61872124 61872691 61873173 61873454 61874241 61875333 61876209 61994871 61876527 61876716 61876828 61876872 61877456 61877536 61877648 61877920 61877951 61878020 61878902 61879153 61879568 61879587 61879762 61880588 61881112 61881780 61881879 62053731 61881972 61882013 61882032 61882093 61882154 61882229 61882258 61882474 61883238 61883294 61883703 61883706 61884701 61884753 61885505 61885568 61885606 61885632 61885704 61885734 61886853 61887328 61887416 61888830 61889209 61889946 61890082 61890411 61890934 61890989 61891010 61891145 61891233 61892836 61892913 61893033 61893172 61893307 61893635 61895046 61895877 61953711 61895989 61896081 61896082 61896097 61896113 61896156 61897437 61897499 61897645 61897659 61897993 61898722 61898999 62091277 61899083 61899964 61900559 61900770 61901110 61901197 61901211 61901226 61901505 61901576 61901957 61902347 61902869 61902890 61902937 61904647 61905099 61905244 61905263 61905655 61905820 61905901 61906913 61907327 61907668 61908281 61908674 61908698 62010993 61909086 61910369 61910538 61910850 61910929 61910957 61911460 62030001 61911576 61912028 61912063 61912099 61912500 61912583 61912590 61912970 61913022 61913259 61913711 61913842 61913945 61914550 61915018 61915060 61915223 61915289 61915403 61916060 61916945 61917328 61917365 61917393 61917414 61918709 61918968 61920766 61920815 61920876 61921867 61922019 61922503 61923378 61923405 61923454 61923488 61923516 61923728 61923997 61924652 61925883 61927170 61927664 61928182 61928229 61928266 61928492 61928972 61929293 61929309 61929371 61929934 61930056 61931155 61931232 61931236 61931251 61932160 61932851 61933185 61933281 61933632 61934690 61934849 61934955 61935211 61935225 61935257 61935587 61937704 61937865 61937922 61937924 61938265 61938556 61938907 61940076 61940163 62003408 61940168 61940624 61940679 61940723 61940741 61941387 61941532 61942327 61942343 61942504 61942764 61942770 61942779 61942865 61943988 61944687 61945216 62009152 61945730 62048985 62153771 61945922 61946293 61946505 61946813 61946892 61948400 61948471 61949936 61950408 61950564 61951590 61951633 61952220 61953363 61953748 61953794 61954045 61954254 61954343 61954750 61955497 61955653 61957399 61957903 61958130 61959835 61959885 61960311 61960817 61960826 61961402 61961956 61962712 61962942 61963034 62045250 61963471 61963531 61963862 61963975 61964194 61964290 61964868 61965166 61965687 61966316 61966480 61966648 61967199 61967353 61968106 61968761 61969536 61969737 61969944 61970083 61970830 61973269 61973300 61973447 61973913 61974877 61974944 61974957 61975579 61975638 61975690 61976222 61976344 61976372 61976621 61977205 61977513 61977683 61978531 61978894 61979030 61979088 61980804 61980880 61980914 61980931 61981091 61981780 61982116 61982240 61982323 61982405 61982434 62162115 62274527 62463530 61983782 61983906 61984054 61984083 61984106 61985178 61985295 61985439 61985510 61985607 61985661 61986068 61986361 61986417 61986815 61986820 61986920 61987034 61988118 61988376 61988410 61989008 61989062 61989579 61989598 61989616 61990178 61990209 61990266 61990349 61990427 61990431 61990645 61991617 61992128 61992399 61992534 61992881 61992891 61993370 61993409 61993468 61993502 61993538 61993849 61993950 61995244 61996199 61996262 61996682 61996875 61996913 61997061 61997070 61997116 62060881 61997177 61997227 61997927 61998726 61999033 62133925 61999107 61999148 61999453 61999726 61999843 62000067 62000111 62000273 62000282 62000337 62000367 62000423 62000448 62000469 62000603 62000757 62001421 62001663 62001893 62001920 62002045 62002299 62002614 62003096 62003232 62003502 62003635 62004332 62005421 62005849 62005877 62005924 62005936 62006026 62006178 62006586 62006622 62006748 62007621 62008080 62009012 62009678 62009704 62009969 62010362 62010389 62010676 62011923 62012194 62013147 62013339 62013528 62014950 62015147 62015253 62015288 62015333 62015434 62015867 62015952 62016381 62016391 62017332 62018120 62018167 62018233 62018924 62019344 62019451 62019757 62019897 62020736 62020929 62020953 62021024 62021036 62021080 62021249 62021498 62021499 62021563 62021601 62138676 62021764 62021789 62021827 62022009 62022040 62022565 62022575 62022629 62136758 62022938 62023675 62023730 62024214 62024954 62025606 62025623 62025628 62025865 62026504 62026518 62026726 62027247 62027668 62027731 62027911 62214952 62027930 62028020 62028337 62028943 62029143 62029432 62029497 62029645 62030473 62030535 62030663 62118773 62030854 62030856 62030857 62031149 62031204 62031886 62031973 62032350 62032852 62032873 62032906 62033780 62033825 62033969 62213889 62034103 62034255 62034552 62034657 62034762 62034918 62035577 62035587 62035828 62035864 62035907 62036010 62036293 62037263 62037880 62038869 62039667 62039788 62039809 62039994 62040017 62124828 62040109 62040167 62040250 62040455 62040472 62041822 62041879 62041885 62041896 62041920 62042053 62042267 62042417 62042433 62042471 62042509 62042784 62043075 62043932 62044012 62044390 62198031 62044533 62044566 62044641 62044941 62045378 62045412 62045871 62045962 62046010 62046041 62046317 62046336 62046349 62046498 62046517 62046533 62046957 62047538 62048152 62220402 62048529 62049678 62049686 62049729 62049766 62049789 62049818 62049844 62049876 62049891 62049901 62050779 62050952 62051470 62051859 62184206 62232946 62052365 62052441 62052464 62052675 62053027 62053376 62053475 62053670 62053909 62054106 62177275 62054466 62054481 62054616 62054721 62054757 62055735 62055761 62056516 62172597 62056565 62056582 62056945 62057591 62057730 62057871 62057880 62057972 62057981 62058004 62058254 62058260 62058294 62058302 62058416 62059214 62059266 62060076 62060811 62061051 62061898 62061927 62062213 62062812 62062936 62063004 62063009 62063153 62063235 62064050 62064143 62064283 62064337 62064443 62064547 62064615 62064724 62065035 62065066 62065282 62065353 62065456 62065467 62065956 62066423 62066562 62067518 62068194 62227443 62068766 62068898 62068935 62069039 62069097 62069118 62069128 62260074 62069161 62069351 62072061 62072318 62072356 62072369 62072390 62072637 62073011 62073030 62073242 62073378 62073926 62073932 62074044 62074153 62074389 62074474 62074744 62074904 62074979 62075148 62075701 62075854 62076117 62076285 62076651 62077250 62078090 62078103 62078949 62079724 62079738 62198777 62079755 62080068 62080550 62081619 62082949 62083169 62083278 62083448 62083603 62083627 62083983 62084134 62084220 62084904 62086168 62086231 62086339 62087260 62087630 62088254 62173796 62090518 62090745 62091601 62091792 62092157 62092628 62092977 62093124 62093207 62093694 62162551 62093698 62093752 62093865 62094132 62094272 62094569 62095055 62095113 62095246 62095309 62096020 62243683 62097087 62097322 62097327 62097659 62097749 62097848 62098037 62098120 62098142 62098352 62098368 62098408 62098425 62388626 62098476 62098838 62098874 62099375 62099462 62099587 62099624 62228233 62099970 62099973 62100133 62100560 62100752 62101346 62101390 62101851 62102651 62102768 62102788 62102832 62102971 62103530 62103627 62238575 62103851 62104253 62104353 62104618 62220482 62104928 62105195 62105297 62105348 62105376 62105507 62105549 62106230 62106354 62106366 62106538 62106701 62106713 62107511 62107681 62107692 62107730 62107847 62108192 62108683 62108781 62108783 62108917 62108967 62109018 62109036 62109052 62109334 62109405 62109568 62110112 62110123 62110232 62180063 62110532 62110623 62110843 62111074 62111122 62111426 62111484 62111568 62111659 62112226 62112233 62112504 62112829 62112990 62113023 62113154 62113186 62113206 62113323 62113379 62113827 62113834 62114158 62114826 62114984 62115040 62115055 62115069 62204368 62115213 62115651 62115743 62115980 62116032 62116289 62116578 62116807 62117216 62117431 62117436 62117461 62117579 62117591 62117675 62117729 62117795 62117964 62118115 62118121 62118391 62118545 62119059 62119123 62120081 62120210 62120254 62120660 62120763 62120814 62120842 62120909 62120954 62121817 62121848 62121862 62121929 62122311 62122577 62123817 62123887 62123924 62123949 62124068 62124283 62124321 62124387 62124548 62124623 62125190 62126212 62127712 62128023 62128102 62128428 62128577 62128593 62128698 62128783 62128882 62129117 62129219 62129586 62130563 62130703 62130799 62131113 62131178 62131609 62131643 62131741 62131987 62132160 62132521 62133219 62133500 62133618 62134838 62134919 62135030 62135110 62135231 62135237 62135598 62136115 62136240 62136354 62137121 62311406 62137217 62137288 62137595 62137973 62137989 62138062 62138260 62138461 62138483 62138653 62138857 62138891 62139156 62139941 62140014 62140242 62267526 62140326 62140436 62140678 62140755 62140793 62140863 62141995 62142033 62142568 62142704 62143362 62143478 62143911 62144325 62144800 62144853 62144881 62144951 62145045 62146203 62146382 62146834 62146899 62147234 62147313 62147761 62147762 62147775 62147853 62148805 62149044 62149277 62149416 62149454 62149588 62149704 62149733 62149759 62149923 62149955 62149977 62150143 62150554 62150972 62151556 62151792 62151947 62152138 62152225 62152658 62152842 62153136 62153149 62153331 62153351 62153363 62388350 62153408 62154384 62245803 62154773 62154781 62154785 62155216 62155242 62155257 62155581 62155692 62156612 62156968 62157010 62157151 62157488 62157621 62157646 62157805 62158012 62158150 62158512 62158548 62158704 62262569 62158714 62159319 62159706 62159709 62159990 62160005 62161120 62161887 62162174 62162282 62162399 62163140 62285147 62163630 62163971 62164110 62164129 62164138 62164217 62164349 62164977 62165053 62165206 62165283 62165318 62166448 62166473 62167491 62167759 62530173 62168335 62168362 62168584 62327700 62168634 62168654 62168741 62168833 62168939 62170750 62170819 62170830 62170928 62170946 62171722 62171823 62173312 62173431 62265989 62173444 62174434 62174740 62175617 62175860 62175984 62176172 62176234 62176277 62176608 62176687 62176749 62176768 62176862 62176883 62177390 62178831 62610906 62178937 62179357 62179507 62179603 62179867 62182305 62182340 62182528 62182608 62182819 62183270 62183662 62183946 62183982 62184013 62184060 62184075 62184272 62184429 62184687 62185979 62186035 62186570 62186584 62186771 62187073 62187409 62187508 62187724 62188633 62188846 62189225 62189958 62190053 62190579 62191374 62191566 62192309 62192504 62192514 62193171 62194037 62194506 62195132 62195404 62195484 62195505 62195866 62195898 62252933 62195946 62196937 62197539 62198321 62198781 62198974 62201830 62202109 62202900 62202992 62203323 62203813 62204403 62205426 62206511 62207414 62207473 62207497 62207912 62208455 62209636 62210038 62210108 62210237 62210498 62210567 62210616 62210631 62210662 62212645 62212665 62212899 62213511 62213567 62213918 62214279 62215065 62215563 62215697 62215966 62215977 62216301 62217474 62218489 62218748 62218815 62218929 62219033 62219167 62219771 62220171 62220209 62220983 62220990 62221036 62221061 62221089 62221132 62221148 62221171 62221194 62221384 62221644 62222516 62222697 62222965 62223109 62223514 62223810 62224235 62294579 62224285 62224415 62224430 62224451 62225599 62225655 62225803 62226401 62226457 62227248 62228504 62228510 62228637 62229924 62230026 62230790 62231010 62231487 62231626 62232050 62232317 62232869 62233354 62233389 62233415 62233801 62234148 62235027 62235073 62235146 62235178 62235350 62235431 62235802 62235892 62236017 62236438 62237369 62239142 62239154 62239268 62240471 62241549 62241614 62241641 62241655 62241696 62241713 62242339 62242354 62242368 62242469 62242920 62243799 62243829 62244589 62244726 62245317 62245511 62245523 62245734 62245850 62247856 62247884 62247895 62247919 62247937 62248184 62248214 62248303 62248318 62248486 62248744 62249085 62249307 62249658 62249834 62249900 62250001 62250009 62250048 62250830 62251352 62251447 62251455 62251754 62252580 62252727 62252772 62252926 62253035 62253038 62253424 62253594 62253688 62253702 62253801 62253827 62253852 62255928 62255944 62256109 62257020 62257988 62777107 62258190 62258619 62259580 62260767 62260894 62261162 62261680 62261988 62262527 62263092 62263118 62263346 62263849 62264195 62264917 62264934 62264947 62265068 62265080 62265153 62265194 62265271 62265310 62267149 62267502 62267696 62268090 62268104 62268319 62269064 62269589 62269604 62269926 62270015 62270455 62270563 62270660 62270798 62270809 62271169 62271319 62272983 62273064 62273070 62274091 62274141 62274172 62274626 62274664 62276191 62276402 62277584 62277726 62278731 62280256 62280286 62280429 62280704 62280783 62280865 62280876 62280910 62281411 62282169 62282403 62283553 62283824 62284383 62285524 62286186 62286427 62287791 62288863 62288981 62290125 62292301 62292966 62302062 62302503 62309139 62311146 62311239 62315072 62315263 62317780 62318035 62318152 62319795 62410185 62418376 62500244 62517148 62528618 62615420 61727955 50706089 50720844 50721239 50721355 50722552 50723265 50724289 50724715 50724999 50725328 50725832 50727298 50728232 50728788 50728969 50729146 60001641 50731031 50731745 50732215 50732333 50732469 50732961 50733261 50733355 50733860 50734194 50734383 50734533 50735698 50736656 50737034 50737168 50737314 60012050 50738337 50741763 60002355 50742031 50742425 50744002 50744362 50745309 50745516 50746294 50746347 50746604 50746817 50746915 50747034 50748182 50749884 50750094 50750288 50750388 50750547 60006805 60019198 50751517 50751926 50751973 50752048 50752054 50752423 50752495 50752650 50753090 50753494 50753854 50755271 50756034 50756966 50757300 50761113 50761708 50761924 50762973 50762979 50762995 50766914 50768237 50769399 50770066 50770845 50773426 50777874 50778281 50779324 50779708 50779990 50780051 50780572 60008186 50780683 50780948 50781299 50781332 50782729 50782746 50782901 50783522 50783706 50784708 50785072 50787121 50787415 50789370 50789647 50789769 60989971 50790927 50791547 50791729 50792885 50794281 50796708 50798948 50803909 50804001 50806077 50808486 50809074 50809103 50809901 50811073 50813291 50814864 50816235 60895625 50818997 50820562 50821181 50823153 50823873 50828165 50832719 50835170 50835970 50837752 50838477 50839959 50840576 50840665 50841593 50843004 50845396 50845689 50845775 50846540 50852981 50854190 50860904 50861243 50862329 50862649 50865460 61326608 50865469 50873445 50874439 50879104 50880665 50881002 50881253 50882522 50884711 50889625 50892184 50898507 50898531 50905160 50908223 50908712 50909633 50910471 50910791 50911538 50914854 50919575 50921454 50924022 50924584 50926205 50926510 50927210 50929702 50930093 50931106 50931134 50932245 50936206 50939072 50940114 50942104 50942233 50943428 50945919 50947152 50947175 50947560 50948002 50948032 50948421 61545160 50949325 50951119 50951451 50952669 50952683 50952934 50953653 50953655 50953869 50955252 50955393 50955734 50956099 50956200 50957063 50957694 50957751 50958865 50959037 50959227 50959889 50960513 50961537 50961917 50961986 50962147 50962152 50962230 50962291 50962473 50962728 50963652 50964545 50964923 50965543 50966479 50967087 50967098 50967416 50967767 50968422 50968464 50969185 50969369 50970073 50970297 50971755 50971808 50972550 50972966 50972986 50973881 50973940 50975419 50976166 50976243 60479815 50977972 50977990 50978789 50979363 60229747 50979702 50980083 50980173 50980610 50983905 50984499 50984500 50984908 50985395 50985788 50986913 50987304 50988392 50988804 50988811 50988965 50989377 50990346 50990376 50990859 50990952 50991028 50991507 50991537 50992232 50992666 50994125 50994551 60480506 50994989 50995576 50995842 50997086 50997382 50997986 50998097 60520150 50999011 50999282 50999769 51000114 51000558 51002660 51003120 51003739 51003873 51003945 51005646 51006084 51006098 51006466 51006506 51008719 51010336 51010512 51010703 51011298 60342528 51011691 51011850 51012890 51013491 51014021 51014513 51014644 51014993 51015981 51016153 51016364 51016730 51017732 51018217 51018312 51018856 51019210 51019260 51020031 51021576 51021710 51022337 51022600 51023173 51023321 51023776 51023997 51024537 51024639 51024805 51025189 51025273 51025291 51025323 51026317 51026417 51027172 51027440 60305308 51027552 51027559 51028088 51029611 51030579 51030882 51031648 51032617 51032636 51032669 51032700 51033578 51033643 51034182 51034304 51034975 51035149 51036928 51037090 51037299 51037364 51038549 51038616 51038619 51038797 51038871 51038908 51039379 51039928 51040209 51040229 60366703 51040501 51041869 51042105 51043496 51044010 51044328 51045328 51046595 51046615 51046653 51046940 51047303 51047495 51047559 51047949 51047986 51049406 51049954 51049984 60393411 60485974 62252090 51050063 51050277 51050302 51050404 51050698 51051254 51051708 51051816 51051890 51052555 51053139 51053157 51053309 51053440 51053445 51053770 51054453 51054612 51054795 51054847 51056230 51057400 51057607 51057898 51057959 51058259 51058752 51058838 51059144 51060246 51060584 51060966 51061603 51061752 51062264 51062583 51063226 51063331 51063625 51063647 51065411 51065468 51065552 51065602 51065727 51065980 51066216 51067281 51067447 51068004 51068603 51069173 51069244 51069261 51069296 51069383 51069574 51069743 51070086 51071045 51071457 51071465 51071976 51072386 51073632 51074252 51074697 51075121 51075716 51076243 51076911 51077653 51078294 61383675 51079384 51079821 51081165 51081194 51081210 51081267 51081714 51082679 51082932 51083245 51083701 51083876 51084235 51084246 51084269 51084463 51084891 51084977 51084996 51085832 51087998 51089059 51089192 51089681 51092138 51092337 51093312 51094940 51094964 51095712 51096402 51097036 51097767 51098544 51098561 51098677 51098713 51098873 51100922 51100999 51101174 51101416 51101808 51102135 51102163 51103610 51104001 51106869 51106938 51106973 51107635 51107961 51108936 51110424 51110998 51111437 51111901 51112580 51113950 51113964 51114964 51115377 51115978 51116368 51116402 51116789 51116985 51117107 51117268 51117287 51117880 51120920 51121678 51121679 51122109 51122724 51123451 51123675 51124181 51124661 51124706 51124998 51126320 51126973 51127840 51127900 51128043 51128292 51128423 51129230 51129284 51129762 51130471 51131099 51132344 51133309 51133496 51133532 51134575 51136083 51136204 51137013 51137403 51139626 51140774 51141299 51142059 60601526 51145635 51146305 51146606 51146666 51146804 51147184 51147502 51148644 51148650 60663353 51149709 51150752 51151039 51152052 51152620 51152669 51154450 51156562 51157133 60647508 51158056 51158258 51158263 51158550 51158644 51161110 51161268 51161409 51161469 51161743 51162429 60660446 51163403 51163765 51164205 51164374 51164384 51165405 51165596 51165860 51166191 51166397 51166405 60748458 51166555 51166561 51166747 51167176 51167956 51168642 51169238 51169481 51169599 51169630 51169841 51170156 51170555 51170919 51170925 51171693 51172308 51172765 51174100 51174105 51174916 51174931 51175119 51175306 51175314 51175464 51175630 51175854 51176467 51176477 51176572 51177508 51177689 51178003 51178323 51178471 51178634 51179351 51180340 51180404 51180990 51181214 51181231 51181419 51181886 51182594 51183012 51183287 51183838 51184321 51188860 51189308 51189593 51189743 51189753 51190790 60702715 51191309 51191653 51191700 51192927 51192948 51193189 51194185 51195054 51195500 51195753 51196628 51196946 51197143 51197913 51198597 51199102 51199157 51199394 51200272 51200537 51200653 51200665 51201078 51201209 51202215 51202536 51203738 51204099 51205110 51205612 51205772 51206023 51206437 51206911 51207747 51208372 51209781 51210156 51210158 51210164 51211476 51211703 51211730 51211753 51212299 51213093 51214015 51214205 51214213 51214398 51214450 51214555 51215066 51216048 51220526 51223094 51223385 51226996 51228057 51228831 51229951 51229982 51230307 51230657 51230814 51231391 51232713 51233265 51233591 51233720 51233787 51233890 51234192 51234302 51235260 51235877 51235935 51236482 51236612 51236959 51237064 51237433 51237553 51237791 51238037 51238563 51238595 51238645 51238649 51238919 51239136 51239150 51239265 51239384 51239405 51239977 51240034 51240179 51240294 51240552 51240572 51240685 51240763 51241041 51241125 51241381 51241503 51241589 51241644 51241760 51242105 51242943 51242961 51243167 51243350 51243856 51244054 51244227 51244570 51244572 51245016 51245183 51245277 51245313 51245705 51245966 51246511 51246686 51246795 51246885 51246982 51247703 51247797 51248155 51248768 51248966 51249027 51249104 51249175 60934087 51249736 51249810 51250136 51250308 51250658 51250968 51251116 51251283 51252043 51252106 51252116 51252133 51252556 51252772 51252973 51252994 51253449 51253553 51253606 51254704 51254789 51254915 51254980 51256018 51256580 51257311 51257387 51257540 51257949 51258417 51259303 51259321 51259412 51259681 51259933 51260198 51260797 51261575 51262228 51262378 51262706 51263448 51263580 51263742 51263844 51264202 51264333 51264712 51265720 51265917 51266058 51266514 51267672 51268557 51268880 51269126 51269763 51270098 51270105 51270380 51271099 51271360 51272341 51272363 51273037 51273612 51274315 51274379 51274473 51275143 51275144 51275444 51275481 51275567 51276261 51276488 51276507 51276612 51276630 51277628 51278564 51278758 51278812 51279926 51280701 51280718 51280731 51280935 51280959 51280974 51281032 51281516 51281656 51281773 51282397 51282574 51282765 51283158 51284279 51284314 51284449 51285025 51285505 51285798 51285947 51286063 51286751 51287161 51287373 51287903 51288688 51288876 51289237 51289797 51289810 51290353 51291011 51291306 51292356 51292785 51293085 51293436 51293472 51293665 51293670 51293689 51293716 51294153 51294294 61860448 51294693 51295063 51295317 51295762 51295946 51296049 51297785 51298839 51299622 51299955 51300024 51300277 51300606 51301510 51301513 51301757 51301979 51302499 51303476 51303600 51303629 51304898 51305264 51305325 51305493 51305626 51305901 51306355 51308335 51308443 51308478 51308915 51309937 51310348 51310393 51310806 51310943 51311161 61068215 51311342 51311392 51311442 51312231 61084159 51312476 51312571 51313054 51313197 51313903 51314012 51314297 51315115 51315431 51316248 51317149 51317749 51317834 51317839 51317855 51317970 51318297 51318396 51319661 51319680 51319934 51320312 51320336 51320432 51320488 51323316 51323452 51323969 51324215 51324566 51324718 51324795 51324949 51324993 51325354 51325575 51325587 51325696 51325763 51325774 51325868 51326567 51327057 51327162 51328079 51328185 51328453 51328648 51328668 51328945 51330445 51331061 51331596 51332872 51332972 51333263 51333831 51334150 51334435 51334667 51334674 51335534 51335854 51336855 61148200 51337561 51339211 51339516 51343181 51343822 61208560 51343980 51344043 51344131 51344901 51345196 51345754 51345821 51346023 51347052 51347359 51348039 51348196 51348255 51348661 51348735 51348771 51350550 51350676 51351630 51351815 51352781 51353279 51354256 61220081 51354306 51354527 51354840 51354912 51355487 51355855 51355955 51356105 51356151 51356339 51356722 51356735 51356993 51357124 51358025 51358524 51358710 51359846 51359901 51360118 51360586 51361053 51361662 51361690 51361807 51361825 51362175 51363027 51363188 51363380 51363388 51363484 51363529 51364367 61250299 51364655 51364708 51364845 51365866 51366715 51366916 51367282 51368301 51368526 51368686 51369199 51369385 51369600 51369630 51369637 51369704 51369947 51370147 51370360 51370695 51370899 51370933 51370948 51371716 51371727 51371779 51371830 51371956 51372234 51372939 51373094 51373461 51373550 51373734 51373755 51374007 51374060 51374339 51374520 61273480 51374600 51374944 51375681 51375786 51375893 51376292 51376463 51376593 51376730 51377278 51377805 51378233 51378279 51378289 51378433 51378624 51378898 51378905 51379239 51379810 62207827 62507710 51380015 51380263 51380350 51380519 51381250 51381536 51381602 51381638 51381712 51381869 51383006 51383334 51383503 51383734 51384038 51384282 51384599 51384639 51384727 51386048 51387080 51387404 51387490 51387581 51387845 51387931 51388108 51388315 51388354 51388480 51388568 51388736 51389092 51389404 51389563 51389851 51390296 61442211 51390463 51390934 51390995 51391659 51391691 51391809 51392031 51392251 51392310 51393369 51393618 51393739 51394659 51394941 51395452 61349420 51395509 51395644 51397556 51398098 51398160 51398231 51398948 51399463 51399502 51399726 51399884 51400523 51400865 51400933 51401009 51401079 51401764 51402097 51402100 51402256 51402312 51403157 51403228 51404024 51404181 51404355 51404364 51404488 51405170 51405886 51406592 51407085 51407099 51407357 51407463 61390480 51407641 51408883 51409134 51409323 51409608 51409951 51410228 51410683 51410698 51410772 51411367 51411434 51411450 51411782 51412402 51412934 51413048 61480512 51414090 51414200 51414331 51415801 51416274 51417728 51418451 51419041 51419372 51419558 51420108 51420778 51420915 51421473 51421643 51422377 51422443 51422576 51422607 51422899 51422991 51423417 51423424 51423601 51423686 51423690 51423859 51424466 51424642 51424864 51424867 51425326 51426008 51426226 51426490 51427284 51427288 51427509 51427872 51428239 51428615 51428626 51429380 51430982 51431251 51431293 51431464 61463049 51431482 51431651 51431716 51431826 51431839 51432637 51432832 51433087 51433345 51433613 51434029 51434089 51434554 51434619 51434816 51435107 51435463 51435799 51436940 51437336 51437398 51437699 51438418 51438424 51438453 51438804 51439232 51439303 51439633 51439766 51440031 51440067 51440839 51440937 51441165 51441241 51441597 51441703 51442210 51443172 51443672 51443719 51444532 51444808 51445247 51445699 51446798 51446955 51446975 51447057 51447138 51447372 51448055 51448272 51448467 51448469 51448608 51448717 51448743 51449210 51449426 51449755 51450077 51450226 61569131 51450421 51450446 51450591 51451427 51452153 51452393 51452501 51452560 51452592 51452978 51453168 51453357 51453722 51454014 51454247 51454997 51455002 51455005 51455275 51457103 51457323 51457412 51457862 51458486 51458539 51458557 51458719 51459179 51459479 51459650 51460199 51460715 51461666 51461680 51461697 51461859 51462020 51462481 51462614 51462902 51463210 61764325 51463355 51463544 51463722 51463930 51463935 51463941 51463977 51464094 51464397 51464502 61657835 51465720 51465853 51465945 51467522 51467820 51467824 51467864 51467872 51467889 51468487 51468620 51468687 51468798 51468851 51469046 51469273 51469583 51470229 51470502 51470529 51470916 51470933 51471946 51471979 51472575 51472716 51472896 51472970 51473439 51473446 51473490 51473683 51474025 51474145 51474250 51474337 51474606 51474947 51475092 51475441 51475610 51475724 51476188 51476364 51476667 51476806 51477407 51477625 51477758 51477799 51477934 51479032 51479059 51479070 51479214 51479307 51479816 51480049 51480058 51480125 51480710 51480794 51480933 51481127 51481481 51481518 51481544 51481683 51482480 51482483 51482801 51483455 51483607 51483720 61677404 61923449 51484396 51484766 51485579 51485790 51485958 51486998 51487379 51487442 51487472 51487570 51488361 51488756 51488883 51489169 51489690 51489945 51490181 51490975 51491359 51491864 51492556 51493289 51493735 51493837 61640712 51494334 51495329 51495358 51495920 51496553 51496679 51496805 51497305 51497560 51497843 51498329 51498367 51498741 51499402 51499767 51499954 51499982 51500976 51501057 51501771 51501861 51501879 51501904 51502246 51503577 51504227 51504243 51504295 51504868 51505232 51506205 51506430 51507523 51507820 51508111 51508652 51508989 51509182 51509314 51509427 51509702 51509981 51510655 51511092 51511099 51511528 51511539 51511666 51512035 51512047 51512287 51512993 51513153 51513715 51513979 51514413 51514819 51515091 51515482 51516400 51516407 51516663 51516759 51517863 51517865 51518761 51519136 51520203 51520362 51520394 61752972 51521371 51521541 51521662 51521798 51522745 51522770 51523496 51523695 51524180 51524357 51525855 51526321 51526565 51526665 51526690 51526954 51526958 51527143 51527599 51529155 51529520 51530088 51530605 51530683 51531147 51531611 51531678 51531958 51532021 51532291 51532344 51532801 51533745 51534290 51535463 51535482 51535504 51535632 51537326 51537680 51537937 51538128 51538303 51539797 51541734 51541785 51542161 51542283 51542552 51542864 61854305 51543138 51543145 51543273 51543681 51544719 51545082 51545192 51545485 51545865 51546136 51546789 51546824 51547324 51547461 51547637 51547845 51547946 51548018 51548289 51548672 51548730 51548825 51549068 51549708 51550152 51550477 51550532 51551376 51551623 51552168 51553317 51553666 51553783 51554649 51554949 51555445 51555587 51557770 51558762 51559224 51559550 51559622 51559879 51560162 51561190 51561776 51562135 51562193 51562410 51563040 51563252 51563265 51563547 51563657 51564025 51564251 51564303 51564452 51565561 51565562 51565635 51565909 51566438 51566471 51566577 51567339 51567378 51568145 51568210 51568281 51568307 51569005 51569290 51569368 51569454 51569997 61969612 62060180 51570129 51570407 51571031 62028352 62120519 51571813 51572159 51572201 51572202 51572212 51572755 51572793 51572812 51573317 51573789 51574248 51574510 62073295 51574894 51575518 51576710 51577396 51577406 51577662 51577731 51578039 51578123 51578142 51578542 51579878 51580444 51580574 51581035 51581066 51581378 51581621 51582248 51582251 51582269 51582554 51582555 51582615 51583173 51583666 51583821 51583841 51584378 51584521 51584674 51584739 51584835 51585067 51585209 51585689 51585851 51586248 51586342 51586454 51586529 51586535 51586595 51587036 51587326 51587997 51588713 51588729 51588961 51590449 51590451 51590594 51590622 51591677 51592216 51592437 51592499 51592605 51593584 51593679 51593752 51595591 51596119 51597049 51597568 51598289 51598456 51598635 51598718 51598776 51598810 51598970 51599151 51599201 51599218 51599285 51599936 51600287 51600321 51601004 51601854 51601989 51602555 51602841 51603178 51603845 51604451 51605432 51605704 51606238 51606563 51606802 51606831 51607182 51607254 51607594 51607614 51607860 51607924 51608109 51608221 51608297 51608490 51608760 51608763 51608917 51608960 51609004 51609178 51609250 51609330 51610247 51610440 51610518 51610613 51610700 51610820 51611125 51611160 51611196 51611278 51611307 51611450 51611566 51611794 51611805 51611904 51611979 51612199 51612484 51612515 51612562 51612729 51612735 51612796 51612841 51612907 51612975 51613151 51613155 51613267 51613760 51613845 51613945 51613957 62230387 62235223 51614010 51614011 51614462 51614676 51615259 51615520 51615542 51615543 51616231 51616305 51616365 51616553 51616554 51616564 51616670 51616939 51617436 51617464 51617541 51617764 51617806 51618544 51618841 51618887 51619721 51620103 51620197 51620441 51620548 51620675 51621322 51621502 51621523 51621535 62227164 51621593 51621888 51621952 51622212 51622467 51622565 51622762 51622861 51623167 51623436 51623469 51623864 51624059 51624442 51624757 51625354 51625648 51625681 51625853 51625882 51625910 51626750 51626821 51627555 51627907 51628082 51628277 51628397 51628519 51629290 51629379 51629382 51629688 51629815 51629853 51630453 51630519 51631158 51631581 51631946 51632002 51632264 51632292 51632452 51632590 51632803 51633335 51633382 51633383 51633756 51634006 51634010 62187876 51634211 51634261 51634521 51634664 51634678 51634693 51635653 51635658 51635744 51635792 51636097 51636529 51636847 51637006 51637511 51637899 51638255 51638331 51638505 51639560 51639588 51640065 51640474 51640728 51641000 51641062 51641127 51644112 51644115 51645214 51645557 51646318 51646986 51647145 51647826 51650928 51651655 51652824 51653448 62270449 51653452 51653915 51653982 51654024 51654323 62311805 51654434 51654486 51654736 51654797 51654957 51655088 51655565 51655887 51656041 51656209 51656230 51657569 51657571 51657729 51657994 51658076 51658286 51658527 51658814 51659201 51659313 51659331 51659424 51660045 51660170 62318792 51660374 51660637 51661525 51661844 51661876 51662020 51662236 51662296 51663339 51663816 51663824 51664876 51664971 51665201 51665221 51665684 51665908 51665964 51665984 51666410 51666553 51667050 51667321 51667324 51667758 51668055 51668061 51668393 51668403 51668421 51668622 51668957 51669089 51672045 51673155 51674170 51676432 51677457 51677526 51678065 51678355 51679280 51699653 62489695 50525503 50526954 50528163 50528201 50530385 50531478 50531616 50534511 50539115 50539946 50540012 50540261 50542362 50543182 114746606 114747511 114747553 114749702 114754401 114754657 114756675 114758378 114758535 114759040 114759676 114760170 114760175 114760194 114761508 114761848 114763148 114766360 114766383 114766657 114767883 114767934 114769536 114770179 114771634 114773802 114773955 114774901 114774964 114776074 114776554 114776839 114778915 114778977 114778978 114779079 114779120 114780167 114780203 114781891 114781928 114782011 114782129 50543752 50544902 50546667 50548249 50549852 50549859 50551782 50552111 50552198 50552288 50553027 50557068 50557934 114785081 114786680 114789099 114789104 114789855 114790023 114791394 114792393 114793083 114794133 114795516 114795895 114796002 114797786 114797889 114799793 114800923 114801428 114803602 114803974 114803986 114804216 114805636 114807692 114809476 114809544 114809638 114810168 114810723 114811894 114812667 114812696 114812765 114812811 114814529 114814534 114814568 114815577 114815952 50558238 50558526 50558700 50559109 50562744 50563189 50564026 50564194 50566420 50566794 50567064 50567141 50567653 50567986 50568498 50568672 50570191 50570656 50570729 50571126 50572452 50572856 50573595 50573690 114817469 114818163 114819499 114819501 114821244 114821271 114826088 114826599 114829727 114831572 114831597 114832463 114833434 114833484 114834090 114838471 114838483 114838658 114839554 114841293 114842103 114843611 114843639 114846081 114846099 114846772 114846908 114847293 114847726 114847739 114849021 114849578 114849646 114849940 114850710 114850762 114851267 114852952 114852976 114852999 114854470 50575657 50575800 50576463 50576891 50576894 50577037 50577172 50577353 50577811 50577826 50578887 50579151 50580888 50582085 50583032 50583944 50584236 50584610 50584612 50585187 50586190 50586279 50587065 50587265 50587520 50587803 50587992 50588751 50589008 50589158 50589512 50589562 50590209 50590369 50590914 50592322 50592980 50593348 114855932 114855946 114856586 114858129 114858139 114859427 114861347 114861366 114862076 114862092 114863559 114863633 114864721 114865813 114867229 114868571 114869371 114869389 114870408 114870529 114870929 114872984 114873426 114875194 114876344 114876516 114877977 114878878 114878891 114878960 114879095 114879273 114880438 114880548 114881326 114882758 114882857 114882987 114884606 114885413 114885484 114886142 114886289 114887557 114888454 114888466 114888624 114888639 114892529 114892698 114893247 50590521 50593123 50593502 50594653 50595038 50595326 50595335 50596796 50597567 50598766 50600355 50600527 50601464 50603220 50604486 50604760 50605696 50605804 50606341 50606788 50607005 50607172 50607875 50607980 50608056 50610764 114894663 114894764 114895982 114896012 114896239 114897925 114898119 114898844 114899329 114900510 114901658 114901711 114901937 114903703 114904331 114904364 114904465 114904619 114905736 114906113 114907519 114907533 114908046 114908741 114910131 114911338 114911343 114911569 114911674 114913231 114913821 114913886 114913897 114914379 114917013 114917019 114917026 114917333 114918978 114919072 114919125 114921040 114921074 114921091 114921137 114921147 114921479 114922331 114922632 114923056 114925052 114925064 114925207 114925616 114925947 114926004 114927979 114928716 114929063 114929862 50610682 50611947 50612081 50612539 50612543 50613184 50613392 50613627 50614438 50614764 50614895 50615297 50615361 50615556 50615618 50615636 50615819 50616699 50617560 50617816 50617836 50619935 50620124 50620148 50620798 50620987 50621178 50621316 50621542 50621648 50622585 50622849 50623425 50623599 50624075 50625171 50625858 50626932 50627105 50627221 50628536 50628547 50629444 50629460 50630680 50631091 114929950 114930339 114930446 114930493 114931898 114932877 114936620 114937046 114937772 114938258 114938480 114939361 114939770 114940655 114940660 114940677 114941805 114941966 114942451 114943713 114944251 114944536 114945202 114945472 114945523 114946069 114946555 114946668 114946693 114947874 114948174 114948204 114949010 114949030 114949270 114949490 114951403 114952607 114953626 114953641 114953999 114954525 114955381 114955440 114955456 114955508 114956489 114957068 114957089 114958233 114958334 114958511 114958564 114960112 114960612 114961723 114962110 114962490 114962495 114962661 114962756 114963687 114964249 114964264 114964454 114965798 114965902 114966110 114966275 114966288 114968375 50631381 50631803 50632113 50633069 50633127 50633824 50634222 50635497 50636438 50636876 50637592 50638421 50639471 50640062 50640671 50640745 50640892 50641150 50641307 50641380 50641901 50642122 50642280 50642986 50645083 50645162 50645802 50645883 50646040 50646642 50647438 50647476 50647642 50647917 50648785 50649884 50650266 114970174 114971506 114973566 114973795 114975245 114975552 114976681 114976702 114977402 114977624 114978861 114979061 114979122 114979135 114979488 114979856 114980376 114980392 114980700 114980775 114980889 114981034 114981444 114982199 114982222 114983099 114983812 114984332 114984738 114985641 114986019 114986354 114986398 114986458 114986483 114987777 114989398 114990013 114990053 114991879 114993297 114993481 114994186 114994316 114995770 114997307 114997744 114998518 114998736 114999080 114999330 114999750 115000869 115001082 115001533 115001641 115001664 115002538 115002818 115003046 115003221 115003503 115003526 115004319 115004555 115004576 115005192 50650888 50651443 50651615 50651675 50652135 50652562 50653209 50653322 50654355 50654533 50654831 50655144 50655612 50656179 50656208 50656249 50656438 50656455 50656485 50657616 50658281 50658503 50658652 50658770 50658989 50660131 50660477 50660600 50660785 50661462 50661531 50661619 50661866 50662101 50662194 50662503 50662548 50662584 50662621 50663728 50664550 50665647 50665904 50666146 50666212 50667087 50667098 50668423 50668526 50669042 50669542 50669639 50669743 50669897 50670100 50670217 50670602 50671036 50671193 50671376 50672042 50672207 50672406 50672510 50672659 50672829 50673185 50673250 50673339 50673419 50673662 50673923 50674546 115006393 115006896 115007097 115007131 115007900 115007967 115008319 115008441 115008637 115008667 115008895 115008946 115009340 115009383 115009433 115009472 115010372 115010510 115010886 115011308 115012259 115012701 115012805 115012945 115013988 115014280 115014371 115014430 115015046 115016117 115016168 115016223 115016232 115017989 115018011 115018290 115019062 115019413 115019459 115019548 115019916 115019924 115021097 115021238 115021250 115021302 115021743 115021748 115021768 115021985 115022038 115023889 115025069 115025494 115025548 115025561 115026275 115026296 115026322 115026330 115026560 115027348 115027824 115027839 115027950 115028074 115028082 115028187 115029780 115029810 115029840 115029861 115030099 115032127 115032953 115032960 115033814 115033822 115033867 115033883 115034077 115034087 115034985 115034991 115035865 115035923 115035947 115037045 115037160 115037168 115037179 115037330 115037454 115037786 115037857 115038903 115039037 115039551 115039887 115039964 115041081 115043138 115043611 115044732 115045586 115045815 115045900 115046902 115047359 115047474 50674111 50674749 50674877 50675351 50676619 50677879 50679053 50679822 50680895 50681572 50681591 50682620 50682878 50683230 50683350 50683354 50683610 50683715 50684059 50684092 50685208 50685581 50686219 50686286 50687337 50687932 50688089 50688482 50688742 50688773 50688846 50688856 50688861 50689097 50689429 50690197 50691071 50691491 50691608 50692249 50692317 50692318 50692909 50693500 50693516 50693930 50696483 50697352 115048624 115048927 115049344 115049506 115049626 115049818 115051032 115052357 115052469 115053108 115053368 115053422 115054515 115054809 115055030 115055308 115055394 115055409 115056273 115056562 115057093 115057332 115058773 115058811 115059105 115060440 115060472 115060655 115060692 115060748 115061934 115061953 115062050 115063018 115063395 115063491 115064547 115065097 115065143 115065265 115065273 115065293 115065849 115066001 115066014 115066476 115066517 115066534 115066641 115067078 115068502 115068510 115068766 115068811 115069043 115069904 115070505 115070659 115070690 115070920 115072214 115072359 115072383 115072545 115074411 115074469 115074870 115074890 115075058 115075677 115076637 115076995 115077547 115078577 115078587 115078645 115078847 115078925 115080807 115081599 115082320 115082329 115082784 115082826 115082865 115085700 115086651 115086926 115086954 115086968 115086982 115087894 115088502 115088537 115088552 115088577 115088980 115089014 115089047 50696572 50696955 50697215 50698168 50698387 50698453 50699591 50700714 50701582 50701836 50702488 50702549 50702727 50702731 50703090 50703408 50703451 50703635 50703837 50704992 50706464 50706721 50707834 50707975 50708597 50709155 50709165 50709448 50709649 50709878 50711308 50711416 50711783 50712884 50714822 50716719 50717970 50718036 50718096 50718098 50718945 50719201 50719759 50720109 60000535 115089864 115090541 115092630 115092676 115092708 115093109 115093174 115093183 115094473 115094826 115096289 115096391 115096436 115096685 115096802 115099003 115100671 115101007 115103550 115104682 115106346 115106433 115106936 115106994 115107881 115108407 115108422 115109863 115110024 115110077 115110963 115112362 115112928 115115661 115115737 115115749 115115830 115115879 115116118 115117681 115117688 115117701 115119879 115122251 115123265 115123272 115123351 115123921 115123985 115127637 50717043 50717882 50720184 115132504 115132828 115133037 115137165 115139321 115146478 115152670 115161595 115164271 60001500 60001754 115177126 115184211 115184898 115195815 115196224 115200129 115203295 115204162 115205844 115206418 115208816 50765201 115211916 115213570 115220319 115314350 115322616 115322754 115325642 115335499 115350415 115350912 115364428 115364667 115021085 115025530 115035883 50680682 115055540 115091995 115099824 115119185 115125294 60000828 50748610 50779812 60003288 115280658 115303418 115304784 115333407 115342372 50847523 50858713 115355755 115363873

Other devices under the same FDA submission

287 devices cleared under K103691, K071638, K093321, K223697 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.