GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers recallZ-0894-2025 · Ge Medical Systems Israel, Functional Imaging · initiated Dec 18, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ge Medical Systems Israel, Functional Imaging or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
FDA root cause category: Package design/selection
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at [email protected] to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri
Recall details
- Recall number
- Z-0894-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 18, 2024
- Posted by the FDA
- Jan 10, 2025
- Recalling firm
- Ge Medical Systems Israel, Functional Imaging
- Quantity
- 17 units
- Product code
- KPS
- FDA submissions
- K093514
- Distribution
- Worldwide
Product description
GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
1) H3100XF: UDI/DI N/A, System ID NT1011, Serial Number T2AG12009/ 88227HL4; UDI/DI N/A, System ID NT2002, Serial Number T2AG12011/ 89154HL9; UDI/DI N/A, System ID NT1009, Serial Number T2AG12013/ 90088HL6; UDI/DI N/A, System ID NT1003, Serial Number T2BG11001/ 79573HL2; UDI/DI N/A, System ID NT1004, Serial Number T2AG12003/ 79574HL0; UDI/DI N/A, System ID NT1005, Serial Number 84775HL6; UDI/DI N/A, System ID NT1020, Serial Number T2AG12012/ 89666HL2; UDI/DI N/A, System ID NT1016, Serial Number T2AG13020/ 96237HL3; UDI/DI N/A, System ID NT0002, Serial Number 106491HL4; UDI/DI N/A, System ID NT2004, Serial Number 105724HL9; UDI/DI N/A, System ID NT1012, Serial Number T2AG13016/ 93885HL2; UDI/DI N/A, System ID NT1012, Serial Number T2AG13017/ 93885HL2; UDI/DI N/A, System ID NT1013, Serial Number 94165HL8; UDI/DI N/A, System ID NT1010, Serial Number T2AG12008/ 87697HL9; UDI/DI N/A, System ID NT3001, Serial Number 85748HL2; UDI/DI N/A, System ID NT1017, Serial Number T2AG13027/ 104213HL4; UDI/DI N/A, System ID NT1015, Serial Number T2AG13018/ 94292HL0.
Other devices under the same FDA submission
6 devices cleared under K093514 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Discovery 670 PRO H3100AHG E Medical Systems Israel Ltd. · System, tomography, computed, emission2 recalls0 reports
- Discovery 670 DRG E Medical Systems Israel Ltd. · System, x-ray, tomography, computed2 recalls0 reports
- Discovery 670 ES H3100AGG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- Discovery NMCT 670 ESG E Medical Systems Israel Ltd. · System, x-ray, tomography, computed1 recall0 reports
- Discovery 670 DRG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- NMCT 870 DRG E Medical Systems Israel Ltd. · System, x-ray, tomography, computed1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96079).
- Z-0889-2025 · Open, ClassifiedGE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomography, Computed, Emission
- Z-0890-2025 · Open, ClassifiedGE Healthcare Discovery NM 630, Model/Catalog Numbers: 1) 5376204-30-2, 2) H2401CF, 3) H2401CH, 4) H2401JE, 5) H3100JY; System, Tomography, Computed, Emission
- Z-0891-2025 · Open, ClassifiedGE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2) H2401BC, 3) H2401CE, 4) H2401CG, 5) H3100AG, 6) H3100JW, 7) H3100NC; System, Tomography, Co…
- Z-0892-2025 · Open, ClassifiedGE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2) H2401AW; 3) H3100AN; 4) H3100BT; 5) H3100PX; System, Tomography, Computed, Emission
- Z-0893-2025 · Open, ClassifiedGE Healthcare Discovery NM/CT 670 PRO, Model/Catalog Numbers: 1) 5376204-70-54-2; 2) H2401AK; 3) H2402AC; 4) H3100AH; 5) H3100XB; System, Tomography, Computed, Emission
- Z-0895-2025 · Open, ClassifiedGE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page