GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers recallZ-0894-2025 · Ge Medical Systems Israel, Functional Imaging · initiated Dec 18, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ge Medical Systems Israel, Functional Imaging or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

FDA root cause category: Package design/selection

What the company is doing

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at [email protected] to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri

Recall details

Recall number
Z-0894-2025
Status
Open, Classified
Initiated by the company
Dec 18, 2024
Posted by the FDA
Jan 10, 2025
Recalling firm
Ge Medical Systems Israel, Functional Imaging
Quantity
17 units
Product code
KPS
FDA submissions
K093514
Distribution
Worldwide

Product description

GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

1) H3100XF: UDI/DI N/A, System ID NT1011, Serial Number T2AG12009/ 88227HL4; UDI/DI N/A, System ID NT2002, Serial Number T2AG12011/ 89154HL9; UDI/DI N/A, System ID NT1009, Serial Number T2AG12013/ 90088HL6; UDI/DI N/A, System ID NT1003, Serial Number T2BG11001/ 79573HL2; UDI/DI N/A, System ID NT1004, Serial Number T2AG12003/ 79574HL0; UDI/DI N/A, System ID NT1005, Serial Number 84775HL6; UDI/DI N/A, System ID NT1020, Serial Number T2AG12012/ 89666HL2; UDI/DI N/A, System ID NT1016, Serial Number T2AG13020/ 96237HL3; UDI/DI N/A, System ID NT0002, Serial Number 106491HL4; UDI/DI N/A, System ID NT2004, Serial Number 105724HL9; UDI/DI N/A, System ID NT1012, Serial Number T2AG13016/ 93885HL2; UDI/DI N/A, System ID NT1012, Serial Number T2AG13017/ 93885HL2; UDI/DI N/A, System ID NT1013, Serial Number 94165HL8; UDI/DI N/A, System ID NT1010, Serial Number T2AG12008/ 87697HL9; UDI/DI N/A, System ID NT3001, Serial Number 85748HL2; UDI/DI N/A, System ID NT1017, Serial Number T2AG13027/ 104213HL4; UDI/DI N/A, System ID NT1015, Serial Number T2AG13018/ 94292HL0.

Other devices under the same FDA submission

6 devices cleared under K093514 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96079).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page