Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial… recallZ-1117-2025 · Max Mobility LLC · initiated Dec 20, 2024
Class I Open, Classified All recalls posted in February 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Max Mobility LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to faulty circuit board in speed control remote use in conjunction with a wheelchair power assist, their is a potential for lost of control of device potentially resulting in minor and major injuries.
FDA root cause category: Component change control
What the company is doing
Initially Max Mobility notified consignees of their Urgent Medical Correction between the dates of 12/20/24 - 01/17/2025. Since this time, Max Mobility has expanded the recall due to recalled product that may have been inadvertently comingled with redesigned product and shipped after December 20, 2024. Starting 3/25/25, Max Mobility began reaching out to customer via phone, email and mail about the expansion. In addition, Max Mobility updated their press release on 4/7/2025 1. Click on the link or scan QR code to enter the portal within the letter 2. Complete requested portal information. 3. If you are a dealer and have distributed the affected product to someone else, you will be asked to select a transmission method. You may either (a) notify your end users directly and include a copy of this notice with your communication or (b) provide end user contact information to Max Mobility through the online portal and we will contact them for you. 4. After acknowledgement, instructions for obtaining replacement SpeedControl Dial units will be provided. You will need the associated client ID provided in ATTACHMENT 1 of this letter when requesting SpeedControl Dial replacement(s). 5. Once the replacement SpeedControl Dial has been issued to the end user, you will be asked to verify that the correction was completed via the field action portal and destroy or return the effected SpeedControl Dial to Max Mobility.
Recall details
- Recall number
- Z-1117-2025
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Dec 20, 2024
- Posted by the FDA
- Feb 11, 2025
- Recalling firm
- Max Mobility LLC
- Quantity
- unknown
- Product code
- ITI
- FDA submissions
- K151199
- Distribution
- US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Sweden, Canada, Australia, Japan, New Zealand, Colombia, Argentina, Hong Kong, Canada, Mexico, Bermuda, South Korea, Chile
Product description
Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCK use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
Code information (models, lots, serial numbers)
All serial numbers manufactured between August 17,2023 through November 21, 2024
Other devices under the same FDA submission
2 devices cleared under K151199 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SmartDrive® MX2+Max Mobility, LLC · Wheelchair, Powered858 reports
- SmartDrive® MX2Max Mobility, LLC · Wheelchair, Powered6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96086).
- Z-1116-2025 · Open, ClassifiedMax Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.