Belmont Medical 3-Spike Disposable Set used with the… recallZ-1099-2025 · Belmont Instrument LLC · initiated Jan 2, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Belmont Instrument LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set
FDA root cause category: Process control
What the company is doing
Belmont Medical Technologies issued Urgent Medical Device Correction letter via FedEx to all affected Medical User facilities and US distributors on January 2, 2025. Additionally, email with the notification letter is being emailed to all purchasing agents requesting forwarding the information to the appropriate departments in their organization. Foreign distributors will be notified via email once translations of the communications are completed. Letter states reason for recall, health risk and action to take: 1. Locate the Disposable Sets (P/N 903-00006) in your possession and check their lot numbers against the lot numbers listed above. 2. If the affected Disposable Sets were consumed, please complete the attached form and return the completed form to Belmont Medical Technologies per the instructions provided on the form. No further action is required. 3. If the affected sets are in inventory, please ensure users of these devices are made aware of this notification and are provided with instructions on how to inspect and test the disposable sets prior to use. See Appendix 1. units in inventory from the identified lot numbers should be inspected immediately prior to use. Any unit identified as having cracks or leaks during priming should be discarded. Lot number of the set should be noted and reported to Belmont Medical Technologies. Actions Taken by Belmont Medical Technologies: 1. We have corrected the issue and are in the process of manufacturing products that are free of the defect listed above. Due to the full manufacturing cycle, which includes sterilization, there will be delays in availability of defect free disposable sets in the coming weeks. 2. We will replace any product found defective as soon as the product becomes available. For any questions regarding this action, technical assistance, or to report an adverse event or product complaint, please contact Belmont Medical Technologies using the following information: 1-855-387-4547 TECHSUPPORT@BELMO
Recall details
- Recall number
- Z-1099-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 2, 2025
- Posted by the FDA
- Feb 4, 2025
- Recalling firm
- Belmont Instrument LLC
- Quantity
- 1,029 cases (12,348 sets)
- Product code
- FRN
- FDA submissions
- K242735
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, China, Japan, Malaysia, Poland, Switzerland, Taiwan (Province of China), Ukraine.
Product description
Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or alternate care environments to provide warmed blood and fluids to any patients e10 kg requiring warmed infusion. Part Number:903-00006
Code information (models, lots, serial numbers)
UDI-DI: 00896128002022 Lot Numbers: 20240910, 20241002, 20241006, 20241014
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- 3-Spike Disposable Set (1 Piece) 903-00006PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall1 death97 reports
Other devices under the same FDA submission
7 devices cleared under K242735 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Rapid Infuser RI-2, 120V, 750 ml/min 903-00039Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid132 reports
- Rapid Infuser RI-2, 120V, 1000ml/min 903-00037Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 death76 reports
- 3.0 Liter Disposable Set 903-00032Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall1 report
- 3.0 Liter Reservoir (1 Piece) 903-00018PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid14 reports
- 3-Spike Disposable Set 903-00006Belmont Instrument Corporation · Pump, Infusion8 reports
- Dual Patient Line (1 Piece) 903-00004PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid5 reports
- 3.0 Liter Reservoir (6 Count) 903-00018Belmont Instrument Corporation · Pump, Infusion1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page