Skip to content

Medline General Surgery Tray Kits recallZ-0991-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Dec 31, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The catheter subassembly in the Integrated Arterial Catheter was manufactured with excess material on the catheter hub.

FDA root cause category: Process design

What the company is doing

A MEDICAL DEVICE RECALL notification letter dated 12/31/24 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-24-278-FGX1 Recall Code: RECALL CODE 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-0991-2025
Status
Open, Classified
Initiated by the company
Dec 31, 2024
Posted by the FDA
Jan 29, 2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
942 kits
Product code
LRO
FDA submissions
K213481
Distribution
US Nationwide.

Product description

Medline General Surgery Tray Kits: 1) ARTERIAL LINE INSERTION KIT, Item Number ART1185B; 2) ARTERIAL LINE BUNDLE, Item Number ART255; 3) ARTERIAL LINE TRAY, Item Number ART350; 4) ARTERIAL LINE BUNDLE, Item Number ART355; 5) ARTERIAL LINE BUNDLE, Item Number ART355 ; 6) ARTERIAL LINE INSERTION TRAY, Item Number ART475; 7) ARTERIAL LINE BUNDLE - OR, Item Number ART535A; 8) ARTERIAL LINE TRAY, Item Number ART545A; 9) ARTERIAL LINE INSERTION KIT, Item Number ART775C; 10) ARTERIAL LINE TRAY, Item Number ART840; 11) ARTERIAL LINE KIT, Item Number ART890B

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

1) ART1185B UDI-DI: 10653160379225(each), 00653160379228(case), Lot Number: 2024102590; 2) ART255 UDI-DI: 10653160367598(each), 00653160367591(case), Lot Number: 2024082690; 3) ART255 UDI-DI: 10653160367598(each), 00653160367591(case), Lot Number: 2024101590; 4) ART350 UDI-DI: 10653160362050(each), 00653160362053(case), Lot Number: 2024093090; 5) ART355 UDI-DI: 10653160361817(each), 00653160361810(case), Lot Number: 2024081290; 6) ART355 UDI-DI: 10653160361817(each), 00653160361810(case), Lot Number: 2024112590; 7) ART475 UDI-DI: 10653160367659(each), 00653160367652(case), Lot Number: 2024100790; 8) ART475 UDI-DI: 10653160367659(each), 00653160367652(case), Lot Number: 2024073090; 9) ART475 UDI-DI: 10653160367659(each), 00653160367652(case), Lot Number: 2024081590; 10) ART475 UDI-DI: 10653160367659(each), 00653160367652(case), Lot Number: 2024090990; 11) ART535A UDI-DI: 10653160367604(each), 00653160367607(case), Lot Number: 2024100890; 12) ART535A UDI-DI: 10653160367604(each), 00653160367607(case), Lot Number: 2024080190; 13) ART545A UDI-DI: 10653160375135(each), 00653160375138(case), Lot Number: 2024102890; 14) ART775C UDI-DI: 10653160366102(each), 00653160366105(case), Lot Number: 2024083090; 15) ART840 UDI-DI: 10653160366966(each), 00653160366969(case), Lot Number: 2024082390; 16) ART840 UDI-DI: 10653160366966(each), 00653160366969(case), Lot Number: 2024093090; 17) ART890B UDI-DI: 10653160360834(each), 00653160360837(case), Lot Number: 2024100990;

Devices named in this recall

9 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

314 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96093).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page