regard Item Number: 830106006, LD01114F - Newborn Kit… recallZ-1047-2025 · ROi CPS LLC · initiated Dec 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ROi CPS LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 12/26/2024, the recalling firm issued a notice dated 12/26/2024 via email to their sole consignee notifying them of the recall of the convenience kits. The email also contained the notice issued by the manufacturer of the affected device component. The recalling firm's notice explained the reason for recall, the health risk involved, and the actions to be taken by the consignee, which included: (1) CHECK: Please review your inventory to determine if the manufacturer item numbers and lot numbers listed in the above table are in your inventory. (2) QUARANTINE: Please quarantine the kits affected by this notice and do not further distribute to end user facilities. (3) NOTIFY: Immediately notify any customers to whom you have distributed or forwarded product affected by this medical device correction notice. Provide a copy of this notice to all customers who have received impacted product and require a response. Customer responses will be requested to determine effectiveness of this recall. (4) END USERS: End users are instructed to discontinue use of the affected kits and quarantine. Kits are to be returned to the distribution location. (5) REPLY: Reply to this notice by completing the reply form attached and emailing to [email protected]. (6) CORRECTION: ROi CPS, LLC personnel will conduct correction activities at the distribution location to ensure removal of the recalled Mercury Medical item. The enclosed reply form is for the consignee to report they have read and understand the attached notice. The consignee is to indicate whether they have forwarded the notice to all customers who have been shipped the affected product, whether or not they have the consignee has the affected product, and that they have completed the "Quantity on Hand" portion of the reply form.
Recall details
- Recall number
- Z-1047-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 26, 2024
- Posted by the FDA
- Feb 4, 2025
- Recalling firm
- ROi CPS LLC
- Quantity
- 414 kits
- Product code
- MLS
- Distribution
- Distribution was made to MO. There was no government/military/foreign distribution.
Product description
regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
Code information (models, lots, serial numbers)
Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Regard™ LD01114FRoi Cps, LLC · Labor And Delivery Kit1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page