Skip to content

Intact Vascular Tack Endovascular System, to treat vascular… recallZ-1070-2025 · Spectranetics Corporation · initiated Jan 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spectranetics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.

FDA root cause category: Device Design

What the company is doing

On 1/10/2025, Philips mailed recall notices to customers informing them of the following: Firm will no longer distributed the device. Check your product inventory and quarantine any affected devices to prevent use. Do not open or use any products that have been identified within your inventory. Circulate this notice to all users of the device, or to any organization where the affected product may have been transferred, so they are aware of the product Recall. Complete and return the response form via email to [email protected]

Recall details

Recall number
Z-1070-2025
Status
Open, Classified
Initiated by the company
Jan 10, 2025
Posted by the FDA
Feb 14, 2025
Recalling firm
Spectranetics Corporation
Quantity
2,939
Product code
QCT
FDA submissions
P180034, P190027
Distribution
Distribution US nationwide and countries of: NL, DE, NL, AU, IT, CH, GB, and AT

Product description

Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062

Code information (models, lots, serial numbers)

UDI-DI: REF/UDI-DI/Batch: 154150041/00863328000189/332905 - 349049, 154150042/328641, REF: 154090042/322101, 156135061/00863328000103/332824 - 336121, 206135061/00850003494043/331281 - 336152, 206135062/328643 - 345567, 156135062/271330 - 283066, 206080062/321083 - 349035, 156080062/302251 - 349034

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page