Skip to content

Beckman Coulter DxI 9000 Access Immunoassay Analyzer… recallZ-1115-2025 · Beckman Coulter, Inc. · initiated Jan 3, 2024

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.

FDA root cause category: Software design

What the company is doing

A field service engineer was dispatched to the customer site to perform the software update and to replace the PCB TSI-RSI board from 01/03/2024 through 12/26/2024.

Recall details

Recall number
Z-1115-2025
Status
Completed
Initiated by the company
Jan 3, 2024
Posted by the FDA
Feb 6, 2025
Recalling firm
Beckman Coulter, Inc.
Quantity
61 units
Product code
JJE
FDA submissions
K221225
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brazil, Croatia, France, Germany, Hungary, India, Ireland, Italy, New Zealand, Portugal, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland.

Product description

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

Code information (models, lots, serial numbers)

UDI/DI 15099590732103, serial numbers: 300208, 300201, 300184, 300185, 300169, 300225, 300175, 300172, 300221, 300209, 300211, 300212, 300213, 300214, 300215, 300210, 300174, 300226, 300204, 300216, 300217, 300170, 300220, 300176, 300188, 300199, 300200, 300177, 300222, 300224, 300173, 300181, 300202, 300191, 300223, 300171, 300186, 300190, 300203, 300178, 300195, 300189, 300198, 300227, 300228, 300168, 300194, 300205, 300206, 300207, 300197, 300218, 300179, 300182, 300192, 300219, 300193, 300183, 300180, 300187, 300196.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page