Hot Dog PATIENT WARMING, Underbody Warming Mattress +… recallZ-1198-2025 · Augustine Temperature Management, LLC · initiated Dec 16, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Augustine Temperature Management, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

FDA root cause category: Device Design

What the company is doing

Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.

Recall details

Recall number
Z-1198-2025
Status
Open, Classified
Initiated by the company
Dec 16, 2024
Posted by the FDA
Feb 20, 2025
Recalling firm
Augustine Temperature Management, LLC
Quantity
1 unit
Product code
DWJ
Distribution
US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

Product description

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia

Code information (models, lots, serial numbers)

UDI-DI UDI-DI 00855913001824, Serial Numbers: 1304283

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.