Brand Name: RayStation- Intended as a Radiation Therapy… recallZ-1191-2025 · Raysearch Laboratories AB · initiated Jan 16, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
FDA root cause category: Under Investigation by firm
What the company is doing
Raysearch issued Field Safety Notice, Medical Device Correction #148655 RayStation version 4.5 to 2024B including some service packs via email on 1/17/25. Letter states reason for recall, health risk and action to take: For users of HU-to-mass density CT calibration curves If a user wants to evaluate a plan with respect to relative stopping power/WE range error rather than mass density errors, there are two options: Option 1. Use a HU-to-SPR CT calibration curve instead of a HU-to-mass density curve. " For evaluation of an existing case, the evaluation can be done on an anonymized and exported copy of the patient and plan. " Contact RaySearch Service if you need help with the creation of a HU-to-SPR curve from an existing HU-to-mass density curve. Option 2. Evaluate the plan using a higher (effective) mass density uncertainty that gives similar relative stopping power and WE range change as the nominal value. " The following values of Effective mass density errors that give similar result as an intended Stopping power error have been deduced from a limited set of patients: " For other values of intended Stopping power error, interpolate the approximate Effective mass density value from the table above. " The values are estimates based on a few patients and are expected to vary slightly depending on the mass densities along the beam path. " Note that the values are not symmetrical around 0. For users of HU-to-SPR CT calibration curves Be aware that the density uncertainty given in Robust evaluation, Compute perturbed dose and Robust optimization is used directly to scale stopping power and therefore WE range. Starting from RayStation v2025, scheduled for market release in April 2025 (subject to market clearance in some markets), the impact of the mass density uncertainty in Robust evaluation and Compute perturbed dose when using a HU-to-mass density CT calibration curve will be changed to harmonize with the other use cases. This means that as of RayStat
Recall details
- Recall number
- Z-1191-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 16, 2025
- Posted by the FDA
- Feb 19, 2025
- Recalling firm
- Raysearch Laboratories AB
- Quantity
- 32
- Product code
- MUJ
- FDA submissions
- K222312
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK.
Product description
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Version: RayStation 12A, 12A Service Pack 1 and 2 and 2023B
Code information (models, lots, serial numbers)
Lot Code: UDI: 0735000201054920220616, 0735000201067920221007, 0735000201073020230913 and 0735000201055620230630 GTIN: 07350002010549, 07350002010679, 07350002010730 and 07350002010556 Serial Numbers: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Revisions: RayStation 12A, 12A Service Pack 1 and 2 and 2023B Expiration Date: 2028-07-07
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- RayStation 12ARaysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment4 recalls0 reports
- RayStation 2023BRaysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment4 recalls0 reports
- RayStation 12A SP1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment4 recalls0 reports
- RayStation 12A SP2Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment4 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96156).
- Z-1185-2025 · Open, ClassifiedBrand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Serv…
- Z-1186-2025 · Open, ClassifiedBrand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113 and 9.0.1.142 Software Version: RaySta…
- Z-1187-2025 · Open, ClassifiedBrand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52 and 10.0.2.10 Software Version…
- Z-1188-2025 · Open, ClassifiedBrand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pa…
- Z-1189-2025 · Open, ClassifiedBrand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStatio…
- Z-1190-2025 · Open, ClassifiedBrand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Versio…
- Z-1192-2025 · Open, ClassifiedBrand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.1.1.2, 6.2.0.7, 6.3.0.6 Software Version: RayStation 6 Service Pack 1, 2 a…
- Z-1193-2025 · Open, ClassifiedBrand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.24 Software Version: RayStation 6
- Z-1194-2025 · Open, ClassifiedBrand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 5.0.2.35 and 5.0.3.17 Software Version: RayStation 5 Service Pack 2 and 3
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page