Medline Convenience Kits for Various Procedures recallZ-1182-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Jan 9, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits contain recalled components due to reports of reduced skin barrier wear time. This refers to a shorter duration that a skin barrier can effectively adhere to and protect the skin before it needs to be replaced.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/9/25 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-282-FGX1 Recall Code: Recall Code 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1182-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 9, 2025
- Posted by the FDA
- Feb 19, 2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 70 kits
- Product code
- CAI
- Distribution
- US States: CA, NY
Product description
Medline Convenience Kits for Various Procedures: 1) ADULT INTUBATION TRAY, Model Number DYNJAA254A; 2) ADULT MAIN DISP. INTUB PART A, Model Number DYNJAA262A
Code information (models, lots, serial numbers)
1) DYNJAA254A, UDI-DI 10195327686260 (each) 40195327686261 (case), Lot Number 24KBR956, 24LBA655; 2) DYNJAA262A, UDI-DI 10195327208172 (each) 40195327208173 (case), Lot Number 24LBD397
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJAA262AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYNJAA254AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96188).
- Z-1183-2025 · Open, ClassifiedMedline Convenience Kits for Various Procedures: 1) CLOVIS RESP/INTUB DRAWER 2, Model Number ACC010809; 2) RESP/INTUBATION DRAWER #2, Model Number ACC010813
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page