Brand Name: ZOLL recallZ-1316-2025 · ZOLL Medical Corporation · initiated Feb 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ZOLL Medical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
On February 13, 2025, URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. REQUIRED ACTIONS Customers who have AEDs should routinely: (1) Monitor device for self-test results. (2) For customers who experience a self-test failure, take action to resolve any alerts as soon as possible. Customers who have at-risk devices should immediately: (1) Communicate this notification to G5 AED users. (2) Ensure users have a program in place that periodically monitors the status of these devices, in accordance with G5 User s Guide. (3) EC 501 is likely due to exposure to humidity outside the device s published specification. Ensure devices are stored within the device s published specifications: " Temperature: 0C to 50C (32F to 122F) " Humidity: 10% to 95% (non-condensing) (4) Respond to ZOLL via the customer notification form acknowledging that you have received this notice. We have notified the appropriate regulatory agencies of this correction and expect it to be classified as a recall. We apologize for any inconvenience this may cause you and thank you in advance for assistance in implementing this correction. Ensuring these devices are rescue ready is our highest priority. Our 24/7 technical support numbers 1 (800) 348-9011 or 1 (978) 421-9460 and are available to assist users with any aspect of this notice. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax with the fax number being 1-800-FDA-0178.
Recall details
- Recall number
- Z-1316-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 13, 2025
- Posted by the FDA
- Mar 12, 2025
- Recalling firm
- ZOLL Medical Corporation
- Quantity
- 2268 units
- Product code
- MKJ
- FDA submissions
- P160033
- Distribution
- Domestic: Nationwide Distribution International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.
Product description
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
Code information (models, lots, serial numbers)
Show the full code information (32k characters)
Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO; UDI-DI: (1) no UDI, (2) 00812394021154, (3) no UDI, (4) no UDI, (5) 00812394022243, (6) 00812394021192, (7) 00812394021390; Serial Numbers: (1) D00000177884, D00000177917, D00000177931, D00000177956, D00000178271, D00000178311, D00000178395, D00000178542, M00000002041, M00000008225, M00000008290, M00000008336, (2) D00000003482, D00000003546, D00000003557, D00000003600, D00000003625, D00000003783, D00000003978, D00000003983, D00000003985, D00000004036, D00000004037, D00000004040, D00000004043, D00000004089, D00000004098, D00000004100, D00000004105, D00000004107, D00000004109, D00000004111, D00000004116, D00000004118, D00000004119, D00000004120, D00000004123, D00000004129, D00000004131, D00000004132, D00000004133, D00000004138, D00000004140, D00000004141, D00000004142, D00000004143, D00000004145, D00000004146, D00000004147, D00000004148, D00000004150, D00000004155, D00000004157, D00000004159, D00000004160, D00000004161, D00000004162, D00000004163, D00000004164, D00000004165, D00000004166, D00000004167, D00000004168, D00000004169, D00000004170, D00000004171, D00000004172, D00000004173, D00000004174, D00000004175, D00000004176, D00000004177, D00000004178, D00000004179, D00000004180, D00000004181, D00000004182, D00000004183, D00000004184, D00000004185, D00000004186, D00000004187, D00000004188, D00000004189, D00000004190, D00000004191, D00000004192, D00000004193, D00000004194, D00000004196, D00000004199, D00000004201, D00000004203, D00000004204, D00000004205, D00000004206, D00000004207, D00000004208, D00000004209, D00000004211, D00000004213, D00000004214, D00000004215, D00000004216, D00000004217, D00000004219, D00000004221, D00000004224, D00000004225, D00000004227, D00000004229, D00000004230, D00000004233, D00000004235, D00000004238, D00000004240, D00000004241, D00000004242, D00000004243, D00000004246, D00000004247, D00000004248, D00000004249, D00000004257, D00000004258, D00000004259, D00000004261, D00000004262, D00000004266, D00000004267, D00000004268, D00000004269, D00000004270, D00000004271, D00000004272, D00000004273, D00000004274, D00000004276, D00000004277, D00000004278, D00000004279, D00000004280, D00000004281, D00000004282, D00000004283, D00000004284, D00000004285, D00000004286, D00000004287, D00000004288, D00000004289, D00000004290, D00000004291, D00000004292, D00000004293, D00000004294, D00000004295, D00000004297, D00000004298, D00000004299, D00000004302, D00000004305, D00000004311, D00000004312, D00000004314, D00000004316, D00000004319, D00000004320, D00000004322, D00000004325, D00000004326, D00000004328, D00000004331, D00000004332, D00000004333, D00000004334, D00000004335, D00000004336, D00000004337, D00000004338, D00000004339, D00000004340, D00000004341, D00000004343, D00000004344, D00000004345, D00000004347, D00000004348, D00000004349, D00000004350, D00000004351, D00000004353, D00000004355, D00000004356, D00000004358, D00000004359, D00000004361, D00000004362, D00000004363, D00000004364, D00000004366, D00000004367, D00000004368, D00000004369, D00000004371, D00000004372, D00000004373, D00000004374, D00000004375, D00000004376, D00000004379, D00000004381, D00000004384, D00000004389, D00000004393, D00000004403, D00000004419, D00000004425, D00000004431, D00000004434, D00000004435, D00000004437, D00000004438, D00000004448, D00000004451, D00000004455, D00000004457, D00000004458, D00000004460, D00000004466, D00000004475, D00000004481, D00000004488, D00000004501, D00000004517, D00000004519, D00000004528, D00000004532, D00000004533, D00000004536, D00000004537, D00000004539, D00000004541, D00000004544, D00000004545, D00000004547, D00000004548, D00000004549, D00000004550, D00000004552, D00000004553, D00000004554, D00000004557, D00000004558, D00000004559, D00000004561, D00000004562, D00000004565, D00000004567, D00000004568, D00000004569, D00000004570, D00000004571, D00000004574, D00000004576, D00000004577, D00000004578, D00000004583, D00000004584, D00000004586, D00000004587, D00000004590, D00000004592, D00000004593, D00000004599, D00000004600, D00000004603, D00000004606, D00000004609, D00000004611, D00000004612, D00000004613, D00000004615, D00000004616, D00000004617, D00000004618, D00000004619, D00000004620, D00000004621, D00000004622, D00000004623, D00000004624, D00000004625, D00000004626, D00000004627, D00000004628, D00000004629, D00000004630, D00000004631, D00000004632, D00000004633, D00000004635, D00000004636, D00000004637, D00000004639, D00000004641, D00000004643, D00000004644, D00000004755, D00000004757, D00000004759, D00000004760, D00000004768, D00000004811, D00000005416, D00000005419, D00000005423, D00000005424, D00000005425, D00000005426, D00000005427, D00000005428, D00000005429, D00000005430, D00000006873, D00000006874, D00000006875, D00000006876, D00000006877, D00000006968, D00000007022, D00000007043, D00000007106, D00000007108, D00000007138, D00000007141, D00000007143, D00000007145, D00000007146, D00000007153, D00000007158, D00000007160, D00000007162, D00000007166, D00000007168, D00000007265, D00000007288, D00000007331, D00000007332, D00000007335, D00000007341, D00000007349, D00000007354, D00000007369, D00000007370, D00000007372, D00000007389, D00000008242, D00000008243, D00000008246, D00000008252, D00000008325, D00000008818, D00000008819, D00000008820, D00000008821, D00000008822, D00000008823, D00000008828, D00000008832, D00000008833, D00000008835, D00000008836, D00000008839, D00000008848, D00000008851, D00000008852, D00000008853, D00000008854, D00000008855, D00000008856, D00000008857, D00000008858, D00000008859, D00000008862, D00000008863, D00000008864, D00000008867, D00000008873, D00000008876, D00000008877, D00000008878, D00000008880, D00000008883, D00000008884, D00000008885, D00000008888, D00000008889, D00000008890, D00000008891, D00000008892, D00000008893, D00000008896, D00000008898, D00000008899, D00000008901, D00000008902, D00000008903, D00000008904, D00000008906, D00000008910, D00000008915, D00000008920, D00000008943, D00000008972, D00000008974, D00000008977, D00000008979, D00000008981, D00000008982, D00000008984, D00000008986, D00000008987, D00000008988, D00000008990, D00000008991, D00000008997, D00000009158, D00000009160, D00000009161, D00000009162, D00000009164, D00000009165, D00000009198, D00000009200, D00000009206, D00000009208, D00000009209, D00000009210, D00000009211, D00000009212, D00000009224, D00000009226, D00000009228, D00000009229, D00000009246, D00000009250, D00000010138, D00000010163, D00000010165, D00000010166, D00000010167, D00000010176, D00000010177, D00000010179, D00000010180, D00000010182, D00000010183, D00000010185, D00000010186, D00000010287, D00000010290, D00000010293, D00000010294, D00000010295, D00000010296, D00000010297, D00000010299, D00000010300, D00000010301, D00000010302, D00000010312, D00000010313, D00000010314, D00000010315, D00000010316, D00000010317, D00000010319, D00000010352, D00000010353, D00000010354, D00000010361, D00000011113, D00000011121, D00000011130, D00000011138, D00000011139, D00000012025, D00000012681, D00000012726, D00000013788, D00000013806, D00000013810, D00000013812, D00000013813, D00000013815, D00000013831, D00000013834, D00000013836, D00000013839, D00000013841, D00000013843, D00000013844, D00000013845, D00000013892, D00000013913, D00000013914, D00000013915, D00000013917, D00000013921, D00000013924, D00000013925, D00000013928, D00000013929, D00000013930, D00000013931, D00000013932, D00000013933, D00000013934, D00000013935, D00000013936, D00000013937, D00000013938, D00000013941, D00000013942, D00000013943, D00000013944, D00000013945, D00000013959, D00000013962, D00000013963, D00000013964, D00000013965, D00000013966, D00000013967, D00000013968, D00000013969, D00000013970, D00000013971, D00000013972, D00000013973, D00000013974, D00000013975, D00000013976, D00000013977, D00000013978, D00000013979, D00000013980, D00000013981, D00000013982, D00000013983, D00000013984, D00000013985, D00000013987, D00000013988, D00000013990, D00000013991, D00000013994, D00000013995, D00000013997, D00000013998, D00000013999, D00000014000, D00000014001, D00000014003, D00000014006, D00000014008, D00000014010, D00000016171, D00000016184, D00000016190, D00000016204, D00000016221, D00000016238, D00000016239, D00000016240, D00000016241, D00000016242, D00000016243, D00000016244, D00000016246, D00000016249, D00000016252, D00000016254, D00000016256, D00000016259, D00000016317, D00000016319, D00000016323, D00000016324, D00000016325, D00000016326, D00000016328, D00000016329, D00000016330, D00000016332, D00000016334, D00000016456, D00000016468, D00000016806, D00000019311, D00000019326, D00000019335, D00000019337, D00000019340, D00000019341, D00000019342, D00000019345, D00000019347, D00000019351, D00000019352, D00000019354, D00000019355, D00000019358, D00000019361, D00000019362, D00000019366, D00000019367, D00000019368, D00000019372, D00000019373, D00000019380, D00000019381, D00000019399, D00000019401, D00000019403, D00000019405, D00000019407, D00000019414, D00000019427, D00000019429, D00000019433, D00000019440, D00000019443, D00000019447, D00000022047, D00000022048, D00000022049, D00000022118, D00000022139, D00000022142, D00000022143, D00000022144, D00000022145, D00000022146, D00000022148, D00000022149, D00000022150, D00000022151, D00000022152, D00000022153, D00000022154, D00000022155, D00000022156, D00000022157, D00000022158, D00000022159, D00000022160, D00000022161, D00000022162, D00000022163, D00000022164, D00000022165, D00000022166, D00000022167, D00000022168, D00000022169, D00000022170, D00000022171, D00000022172, D00000022173, D00000022174, D00000022175, D00000022176, D00000022177, D00000022178, D00000022179, D00000022180, D00000022181, D00000022182, D00000022183, D00000022184, D00000022185, D00000022186, D00000022187, D00000022188, D00000022189, D00000022190, D00000022191, D00000022192, D00000022193, D00000022194, D00000022195, D00000022196, D00000022197, D00000022198, D00000022199, D00000022200, D00000022201, D00000022202, D00000022203, D00000022204, D00000022205, D00000022206, D00000022207, D00000022208, D00000022209, D00000022210, D00000022212, D00000022213, D00000022214, D00000022215, D00000022216, D00000022217, D00000022218, D00000022219, D00000022220, D00000022221, D00000022223, D00000022224, D00000022225, D00000022228, D00000022229, D00000022230, D00000022231, D00000022232, D00000022233, D00000022234, D00000022235, D00000022236, D00000022237, D00000022238, D00000022239, D00000022243, D00000022245, D00000022246, D00000022247, D00000022248, D00000022254, D00000024969, D00000025372, D00000025433, D00000025434, D00000025442, D00000025446, D00000025447, D00000025448, D00000025449, D00000025450, D00000025451, D00000025452, D00000025454, D00000025456, D00000025458, D00000025460, D00000025462, D00000025463, D00000025464, D00000025465, D00000025467, D00000025468, D00000025470, D00000025471, D00000025472, D00000025473, D00000025475, D00000025476, D00000025478, D00000025479, D00000025481, D00000025483, D00000025484, D00000025485, D00000025489, D00000025491, D00000025492, D00000025493, D00000025494, D00000025495, D00000025496, D00000025497, D00000025500, D00000025501, D00000025503, D00000025505, D00000025507, D00000025508, D00000025518, D00000025524, D00000025525, D00000025529, D00000025534, D00000026391, D00000026525, D00000026526, D00000026527, D00000026528, D00000026621, D00000026622, D00000026627, D00000026633, D00000026634, D00000026635, D00000026636, D00000026637, D00000026638, D00000026639, D00000026640, D00000026642, D00000026645, D00000026646, D00000026647, D00000026649, D00000026650, D00000026652, D00000026654, D00000026659, D00000026660, D00000026663, D00000026713, D00000028976, D00000029066, D00000029111, D00000029124, D00000029130, D00000029145, D00000029151, D00000029163, D00000029180, D00000029184, D00000029187, D00000029202, D00000029203, D00000029205, D00000029214, D00000029217, D00000029223, D00000029224, D00000029225, D00000029230, D00000029241, D00000029245, D00000033045, D00000033130, D00000033235, D00000033737, D00000033742, D00000033762, D00000033842, D00000033844, D00000033875, D00000033888, D00000033917, D00000033924, D00000033928, D00000033932, D00000033933, D00000033934, D00000033936, D00000033938, D00000033945, D00000033947, D00000039682, D00000039722, D00000039740, D00000039741, D00000039743, D00000039750, D00000039752, D00000039753, D00000039762, D00000039788, D00000039847, D00000039850, D00000039852, D00000039854, D00000039859, D00000039986, D00000041341, D00000041385, D00000041386, D00000041413, D00000041414, D00000041417, D00000041419, D00000041423, D00000041475, D00000041476, D00000041804, D00000041885, D00000041907, D00000041917, D00000041919, D00000041921, D00000042006, D00000042030, D00000042042, D00000042043, D00000042050, D00000042055, D00000042057, D00000042069, D00000042134, D00000042144, D00000042173, D00000042175, D00000042192, D00000042194, D00000042199, D00000042207, D00000042209, D00000044591, D00000044593, D00000044598, D00000044600, D00000044602, D00000044609, D00000044610, D00000044612, D00000044617, D00000044621, D00000044659, D00000044662, D00000044668, D00000044670, D00000044735, D00000044737, D00000044738, D00000044739, D00000044741, D00000044743, D00000044746, D00000045483, D00000045484, D00000045604, D00000045608, D00000045610, D00000045611, D00000045618, D00000045623, D00000045629, D00000045631, D00000045632, D00000045640, D00000046175, D00000046189, D00000046190, D00000046192, D00000046195, D00000046196, D00000046234, D00000046255, D00000046256, D00000046333, D00000046405, D00000046518, D00000046839, D00000046851, D00000046869, D00000046870, D00000046871, D00000046874, D00000046875, D00000046876, D00000046877, D00000046878, D00000046879, D00000046880, D00000046881, D00000046882, D00000046883, D00000046884, D00000046885, D00000046886, D00000046887, D00000046890, D00000046903, D00000046914, D00000046916, D00000046919, D00000046952, D00000047175, D00000047210, D00000047249, D00000047250, D00000047421, D00000047851, D00000047938, D00000048060, D00000048184, D00000048187, D00000048195, D00000048202, D00000048355, D00000049835, D00000050256, D00000050579, D00000050596, D00000050606, D00000050611, D00000050612, D00000050624, D00000050627, D00000050635, D00000050639, D00000050646, D00000050751, D00000050757, D00000050779, D00000050782, D00000050785, D00000050795, D00000050804, D00000050805, D00000050809, D00000050811, D00000050815, D00000050824, D00000050825, D00000050828, D00000050830, D00000050831, D00000050833, D00000050835, D00000050837, D00000050838, D00000050841, D00000050842, D00000050844, D00000050846, D00000050851, D00000050852, D00000050860, D00000050861, D00000050864, D00000050867, D00000050873, D00000050875, D00000050880, D00000050883, D00000050885, D00000050889, D00000050893, D00000050894, D00000050896, D00000050899, D00000050902, D00000050903, D00000050909, D00000050913, D00000050915, D00000052399, D00000052401, D00000052402, D00000052406, D00000052407, D00000052408, D00000052412, D00000054178, D00000054186, D00000054192, D00000054196, D00000054229, D00000054230, D00000054232, D00000054235, D00000054237, D00000054239, D00000054241, D00000054244, D00000054245, D00000054246, D00000054249, D00000054253, D00000054254, D00000054255, D00000054256, D00000054257, D00000054259, D00000054261, D00000054266, D00000054269, D00000054270, D00000054271, D00000054272, D00000054273, D00000054275, D00000054277, D00000054293, D00000054295, D00000054302, D00000054304, D00000054305, D00000054335, D00000054336, D00000054350, D00000054352, D00000054354, D00000054356, D00000054358, D00000054360, D00000054361, D00000054364, D00000054365, D00000054368, D00000054370, D00000054372, D00000054409, D00000054410, D00000054413, D00000054414, D00000054946, D00000054949, D00000054972, D00000055122, D00000055224, D00000055229, D00000055237, D00000055238, D00000055384, D00000055385, D00000055446, D00000055448, D00000056108, D00000056111, D00000056130, D00000056236, D00000056260, D00000056998, D00000056999, D00000057001, D00000057009, D00000057011, D00000057012, D00000057052, D00000057096, D00000057117, D00000057120, D00000057121, D00000057124, D00000057125, D00000057133, D00000057136, D00000057142, D00000057185, D00000057486, D00000057487, D00000057511, D00000057512, D00000057656, D00000057668, D00000057672, D00000057674, D00000057675, D00000057687, D00000057688, D00000057689, D00000057690, D00000057691, D00000057692, D00000057693, D00000057694, D00000057695, D00000057696, D00000057704, D00000057705, D00000057706, D00000057708, D00000057709, D00000057710, D00000057711, D00000057712, D00000057713, D00000057723, D00000057724, D00000057725, D00000057726, D00000057727, D00000057736, D00000057738, D00000057739, D00000057740, D00000057741, D00000057742, D00000057743, D00000057744, D00000057745, D00000057746, D00000057747, D00000057748, D00000057749, D00000057750, D00000057751, D00000057752, D00000057753, D00000057754, D00000057755, D00000057756, D00000057757, D00000057758, D00000057762, D00000057763, D00000057765, D00000057768, D00000057770, D00000057771, D00000057772, D00000058901, D00000058923, D00000058931, D00000059038, D00000059044, D00000059339, D00000059341, D00000059345, D00000059349, D00000059356, D00000059359, D00000059363, D00000059378, D00000059379, D00000059763, D00000059942, D00000060028, D00000060042, D00000060045, D00000060047, D00000060049, D00000060063, D00000060065, D00000060067, D00000060075, D00000060083, D00000060089, D00000060114, D00000060141, D00000061013, D00000061020, D00000065476, D00000065479, D00000065485, D00000065486, D00000065611, D00000065616, D00000065622, D00000065638, D00000065651, D00000065664, D00000065678, D00000065699, D00000068770, D00000068774, D00000068805, D00000069043, D00000069048, D00000069049, D00000069053, D00000069498, D00000069539, D00000069540, D00000069541, D00000069542, D00000069543, D00000069544, D00000069546, D00000069547, D00000069550, D00000069571, D00000069573, D00000069574, D00000069575, D00000069576, D00000069577, D00000069578, D00000069579, D00000069580, D00000069581, D00000069582, D00000069640, D00000069644, D00000069645, D00000069649, D00000069651, D00000069670, D00000069671, D00000069673, D00000069675, D00000071936, D00000071937, D00000071938, D00000071948, D00000071950, D00000071975, D00000071979, D00000071981, D00000071985, D00000072034, D00000072035, D00000073639, D00000073658, D00000073671, D00000073679, D00000073693, D00000073777, D00000074412, D00000075866, D00000075880, D00000075881, D00000075882, D00000075893, D00000076470, D00000076474, D00000076480, D00000076487, D00000076532, D00000076541, D00000076840, D00000076842, D00000076951, D00000076965, D00000077007, D00000079161, D00000079168, D00000079169, D00000079179, D00000079180, D00000079243, D00000079244, D00000079245, D00000079246, D00000079247, D00000079248, D00000080680, D00000080684, D00000080689, D00000080691, D00000080692, D00000080693, D00000080694, D00000080695, D00000080697, D00000080699, D00000080700, D00000080701, D00000080708, D00000080710, D00000080713, D00000080715, D00000080716, D00000080717, D00000080718, D00000080719, D00000080720, D00000080722, D00000080723, D00000080724, D00000080725, D00000080726, D00000080727, D00000080728, D00000080729, D00000080731, D00000080733, D00000080736, D00000080737, D00000080738, D00000080739, D00000080740, D00000080741, D00000080743, D00000080744, D00000080746, D00000080747, D00000080748, D00000080749, D00000080750, D00000080757, D00000080769, D00000080773, D00000080775, D00000080779, D00000082330, D00000082332, D00000082407, D00000082421, D00000082431, D00000082442, D00000082447, D00000083161, D00000083165, D00000083167, D00000083192, D00000083198, D00000083225, D00000083232, D00000083260, D00000083276, D00000083333, D00000083338, D00000085964, D00000085967, D00000086037, D00000086038, D00000086039, D00000086040, D00000086279, D00000086288, D00000086291, D00000086333, D00000086336, D00000086339, D00000086345, D00000086357, D00000086358, D00000086360, D00000086364, D00000086365, D00000086367, D00000086368, D00000086370, D00000086383, D00000086392, D00000086411, D00000086420, D00000086421, D00000086426, D00000086428, D00000086429, D00000086431, D00000086432, D00000086437, D00000088358, D00000088368, D00000088370, D00000088380, D00000088392, D00000088395, D00000088403, D00000088420, D00000088422, D00000088424, D00000088425, D00000088429, D00000088433, D00000088438, D00000088440, D00000088441, D00000088444, D00000088445, D00000088448, D00000088450, D00000088451, D00000088453, D00000088454, D00000088466, D00000088469, D00000088474, D00000088475, D00000088481, D00000088482, D00000088484, D00000088488, D00000088489, D00000088493, D00000088494, D00000088500, D00000088508, D00000088510, D00000089087, D00000089104, D00000089115, D00000089145, D00000089148, D00000089324, D00000089328, D00000089346, D00000089353, D00000089359, D00000089579, D00000089585, D00000089590, D00000089592, D00000089593, D00000089594, D00000089595, D00000089596, D00000089597, D00000089603, D00000089604, D00000089605, D00000089606, D00000089608, D00000089609, D00000089611, D00000089614, D00000089619, D00000089620, D00000089621, D00000089622, D00000089624, D00000089625, D00000089626, D00000089627, D00000089635, D00000089652, D00000089658, D00000089679, D00000089681, D00000089683, D00000089685, D00000089686, D00000093097, D00000093126, D00000093134, D00000093159, D00000093182, D00000093609, D00000093628, D00000093675, D00000094019, D00000094020, D00000094024, D00000094026, D00000094028, D00000094030, D00000094033, D00000094034, D00000094036, D00000094037, D00000094042, D00000094044, D00000094047, D00000094048, D00000094049, D00000094052, D00000094055, D00000094056, D00000094060, D00000094062, D00000094065, D00000094068, D00000094069, D00000094070, D00000094074, D00000094076, D00000094079, D00000094081, D00000094084, D00000094085, D00000094087, D00000094092, D00000094096, D00000095576, D00000095585, D00000095700, D00000095709, D00000095712, D00000095935, D00000096068, D00000096086, D00000096089, D00000096092, D00000096103, D00000096112, D00000096117, D00000096122, D00000096124, D00000096125, D00000096126, D00000096128, D00000096129, D00000096131, D00000096132, D00000096135, D00000096136, D00000096137, D00000096138, D00000096141, D00000096146, D00000096150, D00000096152, D00000096153, D00000096160, D00000096163, D00000096170, D00000096176, D00000096180, D00000096181, D00000096183, D00000096184, D00000097093, D00000097097, D00000097098, D00000097100, D00000097102, D00000097103, D00000097104, D00000097106, D00000097107, D00000097108, D00000097109, D00000097110, D00000097112, D00000097113, D00000097115, D00000097116, D00000097691, D00000097714, D00000097725, D00000097731, D00000097806, D00000097849, D00000101642, D00000101679, D00000102118, D00000102771, D00000102818, D00000102830, D00000102839, D00000102842, D00000103506, D00000103510, D00000103515, D00000103573, D00000103578, D00000104113, D00000104141, D00000104143, D00000104155, D00000104221, D00000104391, D00000104487, D00000104509, D00000104561, D00000104563, D00000104571, D00000104574, D00000104576, D00000104586, D00000104587, D00000106128, D00000106403, D00000106408, D00000106420, D00000106442, D00000106451, D00000106452, D00000106473, D00000106497, D00000106499, D00000106503, D00000106505, D00000106508, D00000106992, D00000107059, D00000107081, D00000107096, D00000107099, D00000109821, D00000109827, D00000109828, D00000109857, D00000109860, D00000109861, D00000109886, D00000109888, D00000109889, D00000109891, D00000112162, D00000112172, D00000112180, D00000112189, D00000112193, D00000117371, D00000117440, D00000118125, D00000118133, D00000118136, D00000118138, D00000118139, D00000118140, D00000118151, D00000118154, D00000118155, D00000118157, D00000118158, D00000118161, D00000118163, D00000118170, D00000118325, D00000118408, D00000119564, D00000119643, D00000119645, D00000119730, D00000119741, D00000119756, D00000119759, D00000119787, D00000119788, D00000119792, D00000119796, D00000120317, D00000120360, D00000121388, D00000121436, D00000121466, D00000121468, D00000121473, D00000121478, D00000121483, D00000121544, D00000121553, D00000121561, D00000121562, D00000121564, D00000121566, D00000122677, D00000122679, D00000122726, D00000122727, D00000122749, D00000122753, D00000122761, D00000122768, D00000122775, D00000122782, D00000126484, D00000126498, D00000126589, D00000127128, D00000127133, D00000127151, D00000127153, D00000127165, D00000127183, D00000127194, D00000130143, D00000130189, D00000130195, D00000130206, D00000130224, D00000130227, D00000132431, D00000132433, D00000132434, D00000132435, D00000132437, D00000135640, D00000135711, D00000135713, D00000135715, D00000135716, D00000136551, D00000136574, D00000136575, D00000136587, D00000136588, D00000136602, D00000136606, D00000136614, D00000136615, D00000136636, D00000136642, D00000136650, D00000136662, D00000137249, D00000137281, D00000137325, D00000137371, D00000137374, D00000140722, D00000140812, D00000140822, D00000140826, D00000140838, D00000140874, D00000140876, D00000140944, D00000144477, D00000144499, D00000144517, D00000144525, D00000144530, D00000144535, D00000144553, D00000146191, D00000146193, D00000146196, D00000146198, D00000146247, D00000146251, D00000146622, D00000146632, D00000146644, D00000146645, D00000146655, D00000146661, D00000146669, D00000146683, D00000146685, D00000146696, D00000150374, D00000150392, D00000150409, D00000150410, D00000150427, D00000150430, D00000150435, D00000150436, D00000150440, D00000150442, D00000150444, D00000150452, D00000150457, D00000150461, D00000150462, D00000150464, D00000150467, D00000150496, D00000150499, D00000150505, D00000150508, D00000150517, D00000150519, D00000151298, D00000151299, D00000151308, D00000151309, D00000151310, D00000151311, D00000151315, D00000151316, D00000151317, D00000151318, D00000151321, D00000151322, D00000151323, D00000151324, D00000151325, D00000151327, D00000151331, D00000151358, D00000151360, D00000151365, D00000151366, D00000151369, D00000151370, D00000151384, D00000151385, D00000151386, D00000151388, D00000151628, D00000151631, D00000151673, D00000152424, D00000152425, D00000152427, D00000152431, D00000152474, D00000152476, D00000152480, D00000154025, D00000155104, D00000155110, D00000155118, D00000155127, D00000155140, D00000155144, D00000155154, D00000155156, D00000155179, D00000155183, D00000155205, D00000155215, D00000155224, D00000155228, D00000155230, D00000155233, D00000155236, D00000155237, D00000155239, D00000155247, D00000155250, D00000155292, D00000157281, D00000157316, D00000157433, D00000157435, D00000157437, D00000158883, D00000158952, D00000158955, D00000158963, D00000158981, D00000158994, D00000159012, D00000159267, D00000159280, D00000159287, D00000159335, D00000159348, D00000159871, D00000160277, D00000160279, D00000160310, D00000160312, D00000160314, D00000160321, D00000160323, D00000160326, D00000160328, D00000160330, D00000160333, D00000160336, D00000160339, D00000160340, D00000160341, D00000160342, D00000160343, D00000160344, D00000160346, D00000160347, D00000160351, D00000160354, D00000160357, D00000160359, D00000160362, D00000160365, D00000160366, D00000160368, D00000160369, D00000160371, D00000160372, D00000160373, D00000160374, D00000160375, D00000160376, D00000160377, D00000160379, D00000160382, D00000160383, D00000160384, D00000160385, D00000160386, D00000160387, D00000160388, D00000160390, D00000160391, D00000160392, D00000160393, D00000160394, D00000160396, D00000160398, D00000160400, D00000160408, D00000161910, D00000162538, D00000162540, D00000162552, D00000162593, D00000162613, D00000163743, D00000164161, D00000164186, D00000164187, D00000164191, D00000164194, D00000164203, D00000164204, D00000164205, D00000164207, D00000164209, D00000164211, D00000164212, D00000164216, D00000164217, D00000164218, D00000164221, D00000164238, D00000164256, D00000164270, D00000164271, D00000164273, D00000164277, D00000164284, D00000164292, D00000164295, D00000164296, D00000164297, D00000164298, D00000164300, D00000164303, D00000164315, D00000164337, D00000164348, D00000164359, D00000164365, D00000164375, D00000164377, D00000164378, D00000164381, D00000164382, D00000164385, D00000164386, D00000164389, D00000164391, D00000164393, D00000165174, D00000165181, D00000165274, D00000165278, D00000165284, D00000165286, D00000165302, D00000165308, D00000165326, D00000165327, D00000166878, D00000166954, D00000166963, D00000166966, D00000166969, D00000166972, D00000166974, D00000166975, D00000166976, D00000166977, D00000166978, D00000166989, D00000166995, D00000167017, D00000167039, D00000167041, D00000167045, D00000167046, D00000167047, D00000167048, D00000167053, D00000167054, D00000167056, D00000169463, D00000169485, D00000169570, D00000169571, D00000169593, D00000169597, D00000169599, D00000169611, D00000169616, D00000169637, D00000173112, D00000173153, D00000173156, D00000173160, D00000173162, D00000173164, D00000173165, D00000173167, D00000173168, D00000173171, D00000173172, D00000173176, D00000173179, D00000173182, D00000173192, D00000173196, D00000173199, D00000173200, D00000173221, D00000173224, D00000173225, D00000173226, D00000173227, D00000173229, D00000173233, D00000173234, D00000173235, D00000173236, D00000173237, D00000173239, D00000173240, D00000173246, D00000173248, D00000173252, D00000173255, D00000173258, D00000173265, D00000173266, D00000173843, D00000173893, D00000173906, D00000173949, D00000175374, D00000175387, D00000177037, D00000177038, D00000178032, D00000178040, D00000178050, D00000178062, D00000178067, D00000178069, D00000178070, D00000178071, D00000178076, D00000178080, D00000178091, D00000178093, D00000178100, D00000178101, D00000178104, D00000178108, D00000178116, D00000178118, D00000178128, D00000178131, D00000178132, D00000178136, D00000184822, D00000184845, D00000184889, D00000185922, D00000186276, D00000191653, D00000191661, D00000191727, D00000191729, D00000191732, D00000191736, D00000191747, D00000191766, D00000191771, D00000193156, D00000193161, D00000193163, D00000193164, D00000193165, D00000196513, D00000196516, D00000204614, D00000204647, D00000204650, D00000211731, D00000211742, D00000211770, D00000211771, D00000211775, D00000217516, D00000217539, D00000217544, D00000217557, D00000217570, D00000217572, D00000217581, D00000217602, D00000217611, D00000217612, D00000217621, D00000217651, D00000223579, D00000240888, D00000240890, D00000251653, D00000251670, D00000251672, D00000251756, D00000251768, D00000251780, D00000251783, D00000251787, D00000251790, D00000251791, D00000270189, D00000270190, D00000270225, D00000270228, D00000270230, D00000270231, D00000270233, D00000270235, D00000270243, D00000270245, D00000270252, D00000270271, D00000270272, D00000270276, M00000014302, M00000014596, M00000014610, M00000014631, M00000014632, M00000014636, M00000014638, M00000014652, M00000014667, M00000014740, M00000014830, M00000014844, M00000014886, M00000014893, M00000014899, M00000014927, M00000014963, M00000014965, M00000014966, M00000014967, M00000014968, M00000014970, M00000014973, M00000014975, M00000014976, M00000014983, M00000014991, M00000015008, M00000015013, M00000015015, M00000015017, M00000015018, M00000015020, M00000015021, M00000015023, M00000015028, M00000015030, M00000015031, M00000015032, M00000015034, M00000015037, M00000015046, M00000015049, M00000015261, M00000015346, M00000015388, M00000015430, M00000015436, M00000015469, M00000015483, M00000015515, M00000015521, M00000015522, M00000015552, M00000015570, M00000015679, M00000015691, (3) D00000237326, M00000009615, (4) D00000183040, D00000183084, (5) D00000243580, D00000243582, D00000243594, D00000243601, D00000243602, D00000243604, D00000243605, D00000243606, D00000243608, D00000243610, D00000243617, D00000243629, D00000243637, D00000243640, D00000243641, D00000243644, D00000243645, D00000243647, D00000243650, D00000243661,
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80AZoll Medical Corporation · Automated External Defibrillators (Non-Wearable)2 recalls143 reports
- Powerheart® G5 TSO Kit, Auto, US EN/LA ESP G5A-80A-TSOZoll Medical Corporation · Automated External Defibrillators (Non-Wearable)2 recalls1 report
- Powerheart G5 Automatic AED G5A-19AZoll Medical Corporation · Automated External Defibrillators (Non-Wearable)1 recall1 report
- Powerheart® G5 Automatic AED G5A-01AZoll Medical Corporation · Automated External Defibrillators (Non-Wearable)1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96198).
- Z-1317-2025 · Open, ClassifiedBrand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external …
- Z-1318-2025 · Open, ClassifiedBrand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80C, (2) G5A-80C-TSO. Software Version: NA Product Description: The Powerheart G5 automated external …
- Z-1319-2025 · Open, ClassifiedBrand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-01C, (2) G5A-02C, (3) G5A-05C, (4) G5A-08C, (5) G5A-10C, (6) G5A-13C, (7) G5A-15C, (8) G5A-22C, (9) G…
- Z-1320-2025 · Open, ClassifiedBrand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C; Software Version: NA Product Description: The Powerheart …
- Z-1321-2025 · Open, ClassifiedBrand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) i…
- Z-1322-2025 · Open, ClassifiedBrand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-00A-TSO, (2) G5S-01A, (3) G5S-02A, (4) G5S-05A, (5) G5S-08A, (6) G5S-10A, (7) G5S-11A, (8) G5S-80A-TS…
- Z-1323-2025 · Open, ClassifiedBrand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80A; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) i…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page