Definium recallZ-1045-2025 · GE Medical Systems, LLC · initiated Jan 28, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
FDA root cause category: Radiation Control for Health and Safety Act
What the company is doing
Firm sent an "IMPORTANT ELECTRONIC PRODUCT RADIATION WARNING" email/letter to its consignees. The email/letter described the product, the problem and the actions to be taken. The consignees were instructed to do the following: You can continue to use the device as intended for clinical use, by following the instructions below: 1.If you perform an exposure and observe an on-screen message that the AEC termination limit has been reached, please acknowledge the message. 2.Take any relevant actions to adjust techniques or patient positioning for the next exposure to avoid reaching the termination limit for that exposure, if appropriate. Please ensure that all potential users in your facility are made aware of this notification and the recommended actions. Please retain this document for your records. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-1045-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 28, 2025
- Posted by the FDA
- Feb 3, 2025
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 3149 (1697 US; 1452 OUS)
- Product code
- KPR
- FDA submissions
- K172869
- Distribution
- Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Product description
Discovery XR656 HD. X-Ray imaging system.
Code information (models, lots, serial numbers)
Definium Tempo (GTIN: 00195278118356) Definium Tempo Pro (GTIN: 00195278070265)
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Definium Tempo ProGE Hualun Medical Systems Co., Ltd · Solid state x-ray imager (flat panel/digital imager)1 recall0 reports
- Definium TempoGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96208).
- Z-1043-2025 · Open, ClassifiedDefinium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
- Z-1044-2025 · Open, ClassifiedDiscovery XR656HD. X-Ray imaging system.
- Z-1046-2025 · Open, ClassifiedOptima XR646 HD. X-Ray imaging system.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page