CircuitGuard Bacterial/Viral Filter with Elbow… recallZ-1269-2025 · Kung Shin Plastics Co. Ltd. · initiated Jan 9, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Kung Shin Plastics Co. Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.

FDA root cause category: Process design

What the company is doing

On January 9, 2025, the firm sent a recall notice to their sole distributor in the US, followed by a more detailed notice on January 10. Customers were instructed to locate and immediately discontinue use of affected products. Affected product should be quarantined and returned. If products of affected batches have been sold, the distributor should immediately notify the buyers to stop using them, quarantine them, and destroy them.

Recall details

Recall number
Z-1269-2025
Status
Open, Classified
Initiated by the company
Jan 9, 2025
Posted by the FDA
Mar 4, 2025
Recalling firm
Kung Shin Plastics Co. Ltd.
Quantity
30000 ea/ 300 cases
Product code
CAH
FDA submissions
K032878
Distribution
US Nationwide distribution in the state of IL.

Product description

CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)

Code information (models, lots, serial numbers)

UDI (01)00810050853682(10)C2N24002(11)240828(17)290827 Lot C2N24002