BCI recallZ-1405-2025 · Smiths Medical ASD, Inc. · initiated Feb 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
FDA root cause category: Process control
What the company is doing
On February 14, 2025, customers were notified via Urgent Medical Device Correction letters. Consignees were instructed to locate and quarantine all affected product within their institution and destroy them. The recalling firm will contact customers who have returned completed response forms identifying affected products within their possession to discuss credit.
Recall details
- Recall number
- Z-1405-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 14, 2025
- Posted by the FDA
- Mar 20, 2025
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 175,994 US; 45,223 OUS
- Product code
- CCK
- FDA submissions
- K941982
- Distribution
- US Nationwide distribution. International distribution pending.
Product description
CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
UDI-DI 30610586038444 Lots WW8214 022980-1-22 022404-1-17 021860-1-10 021110-1-7 020609-1-2 016288-1-1 022980-1-24 022404-1-18 021860-1-14 021110-1-8 020609-1-3 017774-1-1 022980-1-23 022404-1-19 021860-1-13 021110-1-9 020609-1-4 018139-1-1 022980-1-26 022404-1-20 021860-1-15 021110-1-10 020609-1-5 018184-1-1 022980-1-27 022404-1-21 021860-1-17 021110-1-11 020609-1-6 018252-1-1 022980-1-28 022404-1-22 021860-1-16 021110-1-12 020609-1-7 018458-1-1 022980-1-25 022404-1-23 021860-1-18 021110-1-13 020609-1-8 018457-1-1 022980-1-32 022404-1-24 021860-1-20 021110-1-14 020609-1-9 018614-1-1 022980-1-30 022404-1-25 021860-1-21 021110-1-16 020609-1-10 018641-1-1 022980-1-29 022404-1-26 021860-1-19 021110-1-15 020609-1-11 018789-1-1 022980-1-31 022404-2-2 021860-2-1 021110-1-17 020609-1-12 018789-2-1 022980-1-33 022404-1-27 021860-2-2 021110-1-18 020609-1-13 018789-3-1 023592-1-3 022404-2-1 021860-1-22 021110-1-19 020609-1-14 018789-4-1 023592-1-2 022404-2-3 021860-2-5 021110-1-20 020609-1-15 018936-1-1 023592-1-1 022404-2-4 021860-2-3 021110-1-21 020609-1-16 018936-2-1 023592-1-5 022404-2-5 021860-2-4 021110-1-22 020609-1-19 018936-3-1 023592-1-6 022404-2-6 021860-2-6 021110-1-23 020609-1-20 018936-4-1 023592-1-4 022404-2-7 021860-2-8 021110-1-24 020609-1-17 019179-1-1 023592-1-7 022404-2-8 021860-2-7 021110-1-25 020609-1-21 019254-1-1 023592-1-9 022404-2-9 021860-2-10 021110-1-26 020609-1-18 019323-1-1 023592-1-8 022404-2-10 021860-2-11 021110-1-27 020609-1-23 019383-1-1 023592-1-12 022404-2-11 021860-2-9 021110-1-28 020609-1-22 019409-1-1 023592-1-11 022404-2-12 021860-2-12 021110-1-29 020609-1-24 019410-1-1 023592-1-10 022404-2-13 021860-2-13 021110-1-30 020609-1-26 019549-1-1 023592-1-15 022980-1-3 021860-2-14 021110-1-33 020609-1-27 019613-1-1 023592-1-16 022980-1-1 021860-2-15 021110-1-31 020609-1-25 019671-1-1 023592-1-17 022980-1-2 021860-2-16 021110-1-34 020609-1-28 019747-1-1 023592-1-18 022980-1-4 021860-2-17 021110-1-35 020609-1-30 019782-1-1 023592-1-19 022980-1-5 021860-2-18 021110-1-32 020609-1-31 019810-1-1 023592-1-20 022980-1-6 022404-1-2 021110-1-36 020609-1-29 019920-1-1 9755313 022980-1-7 022404-1-1 021110-1-37 020609-1-34 019974-1-1 9755312 022980-1-8 022404-1-3 021110-1-38 020609-1-32 020000-1-1 9757103 022980-1-10 022404-1-4 021110-1-39 020609-1-33 020022-1-1 9757119 022980-1-11 022404-1-5 021110-1-40 020609-1-35 020022-2-1 9755311 022980-1-9 022404-1-6 021860-1-3 020609-1-36 020074-1-1 9759068 022980-1-12 022404-1-7 021860-1-4 020609-1-37 020126-1-1 9757614 022980-1-13 022404-1-8 021860-1-2 020609-1-38 020127-1-1 9761695 022980-1-15 022404-1-9 021860-1-1 020609-1-39 020159-1-1 9760404 022980-1-14 022404-1-10 021860-1-5 020609-1-40 020160-1-1 9762385 022980-1-16 022404-1-11 021860-1-7 021110-1-1 020227-1-1 9759947 022980-1-17 022404-1-12 021860-1-6 021110-1-2 020323-1-1 9759328 022980-1-18 022404-1-13 021860-1-11 021110-1-3 020346-1-1 9762393 022980-1-19 022404-1-14 021860-1-8 021110-1-4 020386-1-1 9763712 022980-1-20 022404-1-15 021860-1-9 021110-1-5 020435-1-1 9763703 022980-1-21 022404-1-16 021860-1-12 021110-1-6 020466-1-1 020609-1-1 020561-1-1 020551-1-1 020505-1-1 /
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BCI WW8214Icu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96238).
- Z-1404-2025 · Open, ClassifiedCAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW1100
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page