DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137 recallZ-1388-2025 · Beckman Coulter, Inc. · initiated Feb 5, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
FDA root cause category: Process design
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/30/25 was sent to customers. Affected DxI 9000 Access Immunoassay Analyzers could experience an increase in the following errors: IncubateWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) WashWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) This can cause the analyzer to enter the red, or Stopped, system status and any in-process tests are cancelled. There could be a delay in reporting patient test results to the physician. If the error occurs: 1. Confirm that the DxI 9000 Access Immunoassay Analyzer is in either the Paused or Stopped system state. 2. Select the System state button in the lower left corner of the user interface. 3. Select Initialize. When initialization is complete, the analyzer returns to the green, or Running, state. If your laboratory has experienced the issue and it is not resolved by initializing the analyzer, contact your Beckman Coulter representative and schedule a service visit. Beckman Coulter has identified the root cause of this issue and is no longer distributing affected motors for the Incubate and Wash Wheel assemblies. A Beckman Coulter service representative will schedule a visit to replace any affected motors on your DxI 9000 Access Immunoassay Analyzer. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center; From our website: http://www.beckmancoulter.com
Recall details
- Recall number
- Z-1388-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 5, 2025
- Posted by the FDA
- Mar 17, 2025
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 206 units
- Product code
- JJE
- FDA submissions
- K221225
- Distribution
- US Nationwide. Global Distribution.
Product description
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.
Code information (models, lots, serial numbers)
UDI-DI: 15099590732103; Serial Numbers: 300116-300117, 300123-300124, 300126, 300128-300139, 300141-300145, 300147- 300158, 300160-300164, 300166-300179, 300181-300183, 300186, 300188-300195, 300197-300207, 300209-300250, 300252- 300256, 300258-300259, 300261, 300263- 300264, 300266-300277, 300279-300290, 300291, 300293-300299, 300300-300309, 300312-300352
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page