DxI 9000 Access Immunoassay Analyzer, Catalog Number… recallZ-1387-2025 · Beckman Coulter, Inc. · initiated Feb 6, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.
FDA root cause category: Device Design
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter was sent to customers beginning on 2/6/25. Beckman Coulter has determined that a DxI 9000 Access Immunoassay Analyzer containing tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. If the analyzer experiences the issue described above, it can lead to the analyzer entering the red, or Stopped, system status due to Reagent Storage and Tip Stocker "axis is out of tolerance" errors. Any in process tests that require addition of reagent at the time of the error are cancelled. There could be a delay in reporting patient test results to the physician. If the error occurs: 1. Confirm that the DxI 9000 Access Immunoassay Analyzer is in either the Paused or Stopped system state. 2. Select the System state button in the lower left corner of the user interface. 3. Select Initialize. When initialization is complete, the analyzer returns to the green, or Running, state. If your laboratory has experienced the issue and it is not resolved by initializing the analyzer, contact your Beckman Coulter representative and schedule a service visit. Beckman Coulter has identified the root cause of this issue and is no longer distributing affected TSI-RSI boards. A Beckman Coulter service representative will schedule a visit to replace any affected TSI-RSI boards on your DxI 9000 Access Immunoassay Analyzer. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. I
Recall details
- Recall number
- Z-1387-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 6, 2025
- Posted by the FDA
- Mar 17, 2025
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 80 units
- Product code
- JJE
- FDA submissions
- K221225
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, TX, WA.
Product description
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278
Code information (models, lots, serial numbers)
UDI-DI: 15099590732103; Serial Numbers: 300158 300161 300164 300167 300169 300175 300178 300179 300182 300183 300185 300189 300194 300197 300201 300202 300204 300210 300216 300217 300218 300220 300224 300225 300233 300234 300236 300237 300239 300240 300241 300243 300244 300246 300247 300252 300255 300256 300260 300267 300270 300272 300273 300275 300280 300288 300290 300295 300296 300299 300300 300304 300313 300314 300317 300318 300321 300322 300323 300324 300333 300334 300336 300338 300339 300341 300351 300362 300364 300368 300370 300373 300374 300375 300379 300380 300385 300386 300388 300389
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page