DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137 recallZ-1326-2025 · Beckman Coulter, Inc. · initiated Nov 13, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results
FDA root cause category: Nonconforming Material/Component
What the company is doing
An IMPORTANT PRODUCT NOTICE letter dated 11/6/24 was sent to customers. Contact your Beckman Coulter representative if your laboratory has experienced wash wheel dispense arm spin mixer motor errors that are not resolved by initializing. See the System Management Chapter section in the DxI 9000 Access Immunoassay Analyzer Instruction for Use (IFU) for steps to initialize the Analyzer. Your Beckman Coulter service representative will contact you to schedule a part replacement that resolves this issue. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com
Recall details
- Recall number
- Z-1326-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 13, 2024
- Posted by the FDA
- Mar 12, 2025
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 245 units
- Product code
- JJE
- FDA submissions
- K221225
- Distribution
- US Nationwide. Global Distribution.
Product description
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI-DI: 15099590732103; Serial Numbers: 300116 300117 300123 300124 300126 300128 300129 300130 300131 300132 300133 300134 300135 300136 300137 300138 300139 300141 300142 300143 300144 300145 300147 300148 300149 300150 300151 300152 300153 300154 300155 300156 300157 300158 300161 300162 300163 300164 300166 300167 300168 300169 300170 300171 300172 300173 300174 300176 300177 300178 300179 300181 300182 300183 300186 300188 300189 300190 300191 300192 300193 300194 300195 300197 300198 300199 300200 300201 300202 300203 300204 300205 300206 300207 300208 300209 300210 300211 300212 300213 300214 300215 300216 300217 300218 300219 300220 300221 300222 300223 300224 300225 300226 300227 300228 300229 300230 300231 300232 300233 300234 300235 300236 300237 300238 300239 300240 300241 300242 300243 300244 300245 300246 300247 300248 300250 300251 300252 300253 300254 300255 300256 300257 300258 300259 300260 300261 300263 300264 300266 300267 300268 300269 300270 300271 300272 300273 300274 300275 300276 300277 300279 300280 300281 300282 300283 300284 300285 300286 300287 300288 300289 300290 300291 300293 300294 300296 300297 300298 300299 300300 300301 300302 300303 300304 300305 300306 300307 300308 300309 300312 300313 300314 300315 300316 300317 300318 300319 300320 300321 300322 300324 300325 300326 300328 300329 300330 300331 300332 300333 300334 300336 300337 300338 300339 300340 300341 300343 300346 300349 300350 300351 300352 300353 300354 300355 300356 300358 300360 300361 300362 300366 300367 300368 300370 300371 300372 300373 300375 300376 300377 300378 300379 300380 300381 300382 300385 300388 300389 300391 300392 300393 300395 300396 300397 300398 300399 300400 300402 300403 300404 300405 300406 300407 300408
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page