CRANI PACK, Model/Catalog Number WECR41D, medical procedure… recallZ-1406-2025 · American Contract Systems Inc · initiated Feb 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.

FDA root cause category: Device Design

What the company is doing

Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION notice to its consignees on 02/01/2025 via email. The notice explained the issue, risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of the Codman Integrated Bipolar Cord and Tubing Set (Manufacturer's Item Number COD9190002RPB, ACS Item Number 347473). " Add warning labels to affected kits which contain the component to ensure the component is discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected Procedure tray and product case in a prominently visible location to end -users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. You are also responsible for the reconciliation of these customers. Please do not have your customers respond directly to ACS. " Complete the attached Recall Response Form and return as soon as possible. Report any quality problems or adverse events associated with the products listed in this notification to [email protected].

Recall details

Recall number
Z-1406-2025
Status
Open, Classified
Initiated by the company
Feb 4, 2025
Posted by the FDA
Mar 20, 2025
Recalling firm
American Contract Systems Inc
Quantity
91 units
Product code
OJG
Distribution
US: MI

Product description

CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit.

Code information (models, lots, serial numbers)

Lot Code: UDI: 191072216984, Lot Cocdes: 71-052536, Exp: 10/30/2025; 71-051491, Exp: 07/16/2025; LOT: 71-051609, Exp: 07/30/2025; 71-051767, Exp: 08/14/2025; 71-053065, Exp: 12/10/2025

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.