Optima MR360, NMRI system recallZ-1241-2025 · GE Healthcare (China) Co., Ltd. · initiated Jan 31, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare (China) Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
FDA root cause category: Software design
What the company is doing
On January 31, 2025, the firm notified customers through Urgent Medical Device Correction letters. Customers were provided with instructions for continued use of their MR systems. All potential users should be made aware of this safety notification and recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-1241-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 31, 2025
- Posted by the FDA
- Feb 28, 2025
- Recalling firm
- GE Healthcare (China) Co., Ltd.
- Quantity
- 33
- Product code
- LNH
- FDA submissions
- K123417
- Distribution
- Domestic US distribution nationwide. International distribution worldwide.
Product description
Optima MR360, NMRI system
Code information (models, lots, serial numbers)
A. UDI-DI 00840682113762 System ID 0910274038 B. No UDI-DI System IDs 082427040134 082427040185 082427060043 082427070088 082427070093 082427070099 082427100174 082427100178 082427100197 082427120140 082427140095 082427220046 082427270041 082427310088 082427310092 083027288790310 0856270061 AE1373MR01 DZ1339MR03 FI1432MR01 H4678MR04 MRR9282 MRR9877 PL2059MR01 PL2433MR01 RU3338MR03 SA1056MR03 SA2191MR04 SA2253MR01 SA2352MR01 TN1008MR01 TN1241MR01 34352MRS01 (Update 12/29/25 - Additional System IDs 83027809699615, LV4026MR01) (Update 1/28/26 - Additional System ID 82427190104)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA SIGNA ExplorerGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging2 recalls0 reports
Other devices under the same FDA submission
2 devices cleared under K123417 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 1.5T EXPRESS COIL - 16E HEAD NECK ARRAYGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging1 recall0 reports
- NA 1.5T EXPRESS COIL - 9E ANTERIOR ARRAYGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96311).
- Z-1239-2025 · Open, ClassifiedSIGNA Creator, SIGNA Explorer, NMRI systems
- Z-1240-2025 · Open, ClassifiedBrivo MR355, NMRI system
- Z-1242-2025 · Open, ClassifiedSIGNA MR355, SIGNA MR360, NMRI system
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page