Skip to content

CVAC Aspiration System, REF: CVC127020-1, and User Manual… recallZ-1396-2025 · Calyxo, Inc. · initiated Feb 19, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Calyxo, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.

FDA root cause category: Device Design

What the company is doing

On 2/19/2025, correction notices were mailed to distributors and healthcare providers who were asked to do the following: 1) Notify all urologists utilizing the affected aspiration system in your institution that additional instructions are required. Please provide each of them with a copy of the additional instructions. 2) As another safety step, a firm representative will be contacting all users from your institution directly to offer training on these additional instructions. If you prefer to coordinate the training, please specifically state that preference in the acknowledgement. 3) Please acknowledge receipt of the notice, additional instructions, and updated training instructions MA00100.A by emailing [email protected], or calling 833-214-3354 and providing your name, title, institution and address.

Recall details

Recall number
Z-1396-2025
Status
Open, Classified
Initiated by the company
Feb 19, 2025
Posted by the FDA
Mar 21, 2025
Recalling firm
Calyxo, Inc.
Quantity
11,246
Product code
FED
FDA submissions
K233472
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.

Product description

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Code information (models, lots, serial numbers)

CVC127020-1 UDI-DI: 00860005357710, L00018 Rev C

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates