daVinci 5 recallZ-1325-2025 · Intuitive Surgical, Inc. · initiated Feb 26, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
FDA root cause category: Device Design
What the company is doing
On February 26, 2025, Intuitive Surgical issued a Urgent: Medical Device Correction to affected consignees via E-Mail. Intuitive asked consignees to take the following actions: 1. Read and understand this notification. 2. Share this notification with all personnel at your site who perform and support da Vinci procedures. 3. Inform affected personnel when the correction has been completed. 4. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 5. Please retain a copy of this letter and the acknowledgement form for your files. 6. Please inform Intuitive of any Adverse Events* or quality problems concerning the use of the subject devices via the standard complaint process 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 8. Complete and submit the report Online Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178. An Intuitive representative will schedule a visit to replace the affected foot tray assembly.
Recall details
- Recall number
- Z-1325-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 26, 2025
- Posted by the FDA
- Mar 12, 2025
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 439 units
- Product code
- NAY
- FDA submissions
- K232610
- Distribution
- Worldwide distribution - US Nationwide and the countries of Alaska, Arizona, Arkansas, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the country of South Korea.
Product description
Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer to HHE for additional details. Component: N/A
Code information (models, lots, serial numbers)
UDI: 00886874119747/ All da Vinci dV5 consoles and foot trays assemblies manufactured prior to 12/20/2024.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- daVinci 5 380730Intuitive Surgical, Inc. · System, surgical, computer controlled instrument4 recalls579 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.