Synapse PACS Software Version 7.4.x recallZ-1407-2025 · FUJIFILM Healthcare Americas Corporation · initiated Feb 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from FUJIFILM Healthcare Americas Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The incorrect computed patient age is showing in VX for patients less than 3 months old.

FDA root cause category: Software change control

What the company is doing

On February 11, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. ACTIONS TO BE TAKEN BY THE COMPANY Fujifilm will issue corrections for all impacted 7.4.x versions. The following patch releases will be made available: - 7.4.020 for 7.4.0x versions (immediately available) - 7.4.120 for 7.4.1x versions (available February 28, 2025) - 7.4.210 for 7.4.2x versions (available February 28,2025) ACTIONS THAT CAN BE TAKEN BY THE END-USER IMMEDIATELY Regarding this issue, Synapse PACS VX can safely be used by the end-user as long as; 1] the end-user is aware of the issue and, 2] takes appropriate actions when rendering a diagnosis for patients under 18 years of age. The following workarounds can be used: 1) The image acquisition date and date of birth are displayed on the Image. The end-user can use the dates to verify the age or, 2) Open the Series Viewer by double clicking on the thumbnail displayed on top the Synapse VX Image viewing window or the worklist. The Series Viewer Window will render the image on the server using the Server-side rendering (Zero Viewer component) which will calculate the age correctly. Alternatively, the user can switch to the Synapse PACS Zero Viewer/Thin Viewer (Server-side Rendering). The Synapse PACS Zero Viewer/Thin Viewer calculates the age correctly. ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1) Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 2) Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to schedule a patch update. Fujifilm regrets any inconvenience this action may cause and appreciates your understanding and cooperation. Please contact our field action coordinator, Jay Wigley, via email at [email protected], if you have any further questions regarding this field action. Please contact Technical Assistance Center (TAC) at 1-888-385-4622 or 1-888-FU

Recall details

Recall number
Z-1407-2025
Status
Open, Classified
Initiated by the company
Feb 11, 2025
Posted by the FDA
Mar 20, 2025
Recalling firm
FUJIFILM Healthcare Americas Corporation
Quantity
104 units (69 US, 35 OUS)
Product code
LLZ
FDA submissions
K160108, K190232
Distribution
Domestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, OR, TN, TX, VT, WA, WI, WV, WY. International: Australia, Belgium, Brazil, France, Germany, Portugal, UK

Product description

Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.

Code information (models, lots, serial numbers)

Model numbers: Synapse PACS 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200 Software Version Numbers: Synapse PACS 7.4.000, UDI: (01)00854904006008(10)0704000; Synapse PACS 7.4.001, UDI: (01)00854904006008(10)0704001; Synapse PACS 7.4.010, UDI: (01)00854904006008(10)0704010; Synapse PACS 7.4.100, UDI: (01)00854904006008(10)0704100; Synapse PACS 7.4.110, UDI: (01)00854904006008(10)0704110; Synapse PACS 7.4.200, UDI: (01)00854904006008(10)0704200.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.