Skip to content

NAMIC recallZ-1428-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Feb 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

FDA root cause category: Process change control

What the company is doing

Medline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medline will provide labeling that states the following: "ATTENTION This kit contains a recalled item Medline Pressure Monitoring Lines VPN: 24614303 DO NOT USE THIS COMPONENT The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*" Consignees were directed to notify their customers of these actions.

Recall details

Recall number
Z-1428-2025
Status
Open, Classified
Initiated by the company
Feb 12, 2025
Posted by the FDA
Mar 21, 2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
4400 units
Product code
DRI
Distribution
Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia

Product description

namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit

Code information (models, lots, serial numbers)

UDI/DI 10193489067507 (each), 20193489067504 (case), Lot number 0000140528

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 96367).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page