Skip to content

Helix Elite Molecular Standards: Inactivated Standard recallZ-1454-2025 · Microbiologics Inc · initiated Feb 20, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The RSV target may give a late Ct value and could potentially not pass QC.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An Urgent Medical Device Recall notification letter dated 2/20/25 was sent to customers. INSTRUCTIONS: 1. REVIEW lab procedures to understand how this information affects your usage 2. DISCARD affected product. 3. COMPLETE the response form provided. 4. RETURN the response form to [email protected] 5. KEEP this letter for your records. 6. NOTIFY other facilities or customers to whom the product may have been distributed or transferred. 7. CONTACT Microbiologics if a replacement kit is needed. Microbiologics is dedicated to ensuring the success of your quality control processes. Please contact our Recall Support team at 320.229.7080 or [email protected] with questions or concerns. Collect calls may be made.

Recall details

Recall number
Z-1454-2025
Status
Open, Classified
Initiated by the company
Feb 20, 2025
Posted by the FDA
Mar 25, 2025
Recalling firm
Microbiologics Inc
Quantity
28 units
Product code
OHQ
Distribution
Worldwide - US Nationwide distribution in the states of KY, MT, NJ, NY, WI and the countries of Ireland, Germany, UK.

Product description

Helix Elite Inactivated Standard: Inactivated Influenza A/B and Respiratory Syncytial Virus, Multi-analyte, unassayed quality control material, Catalog Number HE0044N

Code information (models, lots, serial numbers)

UDI-DI: 10845357042184; Lot Number: HE0044-131

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page