Maquet VH-3010 Power Supply. Used to deliver power to the… recallZ-1577-2025 · Maquet Cardiovascular, LLC · initiated Mar 10, 2025
Class II Open, Classified All recalls posted in April 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Cardiovascular, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE REMOVAL notification dated March 2025 was mailed to consignees. The notification instructs consignees to examine their inventory for impacted serial numbers. For power supplies whose serial numbers are not listed in the provided Appendix A of recalled devices, devices can continue to be used. If power supplies with affected serial numbers are identified as being in stock, the associated devices should not be used and returned to Getinge. Returns are facilitated by contacting Getinge Customer Service at 888-800-2874 between the hours of 6:00 AM and 5:00 PM PST to request a return authorization number and shipping instructions. Consignees are to forward the provided notification to all users of the device within the facility and forwarded to any locations where devices were further distributed. Consignees with any questions about the recall are to contact their Getinge representative.
Recall details
- Recall number
- Z-1577-2025
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Mar 10, 2025
- Posted by the FDA
- Apr 16, 2025
- Recalling firm
- Maquet Cardiovascular, LLC
- Quantity
- 110 units
- Product code
- GEI
- FDA submissions
- K101274
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.
Product description
Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model No. VH-3010; UDI-DI: 0060756770826; Serial No. H24080065G, H24080074G, H24080012G, H24080022G, H24080184G, H24080185G, H24080051G, H24080117G, H24080069G, H24080016G, H24080079G, H24080041G, H24080056G, H24080030G, H24080068G, H24080034G, H24080058G, H24080026G, H24080033G, H24080020G, H24080032G, H24080053G, H24080066G, H24080077G, H24080144G, H24080123G, H24080042G, H24080070G, H24080008G, H24080003G, H24080174G, H24080064G, H24080027G, H24080044G, H24080047G, H24080049G, H24080061G, H24080078G, H24080010G, H24080138G, H24080080G, H24080006G, H24080015G, H24080038G, H24080085G, H24080084G, H24080031G, H24080101G, H24080148G, H24080040G, H24080055G, H24080025G, H24080052G, H24080048G, H24080093G, H24080126G, H24080073G, H24080007G, H23060082G, H24080075G, H24080113G, H24080127G, H24080011G, H24080005G, H24080013G, H24080023G, H24080057G, H24080110G, H24080009G, H24080067G, H24080029G, H24080043G, H24080183G, H24080165G, H24080017G, H24080001G, H24080037G, H24080060G, H24080072G, H24080076G, H24080045G, H24080024G, H24080028G, H24080114G, H24080136G, H24080062G, H24080018G, H24080054G, H24080063G, H24080112G, H24080118G, H24080071G, H24080182G, H24080002G, H24080004G, H24080014G, H24080019G, H24080021G, H24080035G, H24080036G, H24080039G, H24080046G, H24080050G, H24080059G, H24080081G, H24080082G, H24080083G, H24080125G, H24080186G, H24080189G.
Other devices under the same FDA submission
3 devices cleared under K101274 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- Hemopro 2 w/Vasoshield VH-4001Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories4 recalls105 reports
- HemoPro 2 Adaptor C-VH-4020Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories79 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page