DxI 9000 Access Immunoassay Analyzer, catalog number C11137 recallZ-1529-2025 · Beckman Coulter, Inc. · initiated Jul 12, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.

FDA root cause category: Device Design

What the company is doing

The firm sent notice of "Urgent Medical Device Recall" to affected consignees (all OUS), dated July 3, 2024. Customers were informed of the potential for barcode read failure error, delay of reporting patient test results, and possibility for a small amount of smoke to be emitted. Consignees were advised that no action was necessary if their laboratory has not observed any RSM Barcode Read Failure error messages. If you do observe any such error messages, or notice smoking or burning smell coming from the Dxl 9000 analyzer, please contact your Beckman Coulter service representative. Beckman Coulter will no longer ship these analyzers with the affected ground screw. Your Beckman Coulter service representative will contact you to schedule a part replacement that resolves the issue.

Recall details

Recall number
Z-1529-2025
Status
Open, Classified
Initiated by the company
Jul 12, 2024
Posted by the FDA
Apr 4, 2025
Recalling firm
Beckman Coulter, Inc.
Quantity
30 OUS
Product code
JJE
FDA submissions
K221225
Distribution
OUS (Foreign) distribution to countries of: Australia, Croatia, Czechia, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Africa, Spain, and United Kingdom of Great Britain and Northern Ireland.

Product description

DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids

Code information (models, lots, serial numbers)

UDI-DI 15099590732103 Serial/Lot Numbers: 300116, 300117, 300118, 300123, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300137, 300138, 300139, 300141, 300142, 300143, 300145, 300147, 300148, 300149, 300150, 300151, 300152, 300153, 300154, 300155, and 300156