DuraDiagnost recallZ-1553-2025 · Philips North America Llc · initiated Mar 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for collimator to fall as a result of incorrect installation.

FDA root cause category: Under Investigation by firm

What the company is doing

Consignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an inspection of devices by field service engineers. Engineers may also replace the collimator, if necessary. Consignees with questions are to call Philips Customer Care Solutions Center at 1-800-722-9377 from Monday through Friday (8:00 AM to 8:00 PM EST).

Recall details

Recall number
Z-1553-2025
Status
Open, Classified
Initiated by the company
Mar 10, 2025
Posted by the FDA
Apr 11, 2025
Recalling firm
Philips North America Llc
Quantity
83 units
Product code
KPR
FDA submissions
K141381, K201640
Distribution
Worldwide - US Nationwide distribution in the states of AK, FL, KS, MS, NH, NY, OR, PA, TX, WI and the countries of Argentina, Australia, Canada, Chile, China, Germany, Hungary, India, Indonesia, Italy, Kenya, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, Slovenia, South Africa, Spain, Thailand, United Kingdom, Vietnam.

Product description

Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

Code information (models, lots, serial numbers)

REF 712203; UDI: (01)00884838095205(21); Serial No. 220025, 210023, 220171, 230103, 230152.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 96435).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page