Medline medical convenience kits labeled as recallZ-1505-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Feb 24, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific Medline Kits were manufactured using Intubation ORAL/NASAL Endotracheal Tubes which were subsequently recalled by Smiths Medical.
FDA root cause category: Process change control
What the company is doing
Medline issued a SMITHS MEDICAL URGENT MEDICAL DEVICE CORRECTION notice on 02/24/2025 via email and letter. The notice explained the issue, potential risk, and requested the following: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately.2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form.3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-25-041-FGX1Recall Code: RECALL CODE4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. On 04/07/2025 Medline issued an updated notice requesting the destruction of all affected units.
Recall details
- Recall number
- Z-1505-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 24, 2025
- Posted by the FDA
- Apr 10, 2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 1550 units
- Product code
- LRO
- FDA submissions
- K213481
- Distribution
- California
Product description
Medline medical convenience kits labeled as: 1) NEONATAL INTUBATION TRAY, REF ACC010371C; 2) NEWBORN RESUSCITATION KIT, REF ACC010381A; 3) NEWBORN RESUSCITATION KIT, REF ACC010381B; 4) NEWBORN RESUSCITATION KIT, REF ACC010381C; 5) PEDS GREY-PINK-RED, REF ACC010395; 6) PEDS PURPLE, REF ACC010397; 7) NEONATAL INTUBATION TRAY, REF ACC010541A; 8) NEONATAL INTUBATION TRAY, REF ACC010541B; 9) NEONATAL INTUBATION TRAY, REF ACC010541C; 10) PEDIATRICS INTUBATION TRAY, REF DYNJAA252.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
1) REF ACC010371C: UDI/DI 10193489268188 (ea) 40193489268189 (case), Lot Numbers: 20DDA921, 20EDC042, 20IDC153, 21ADB113, 21CDB435, 21HDB557, 21JDA180, 21LDB244, 22HDB092, 22IDA205, 22JDA908, 22KDA049, 22LDA351, 23EDC611; 2) REF ACC010381A: UDI/DI 10193489800944 (ea) 40193489800945 (case), Lot Numbers: 20BDB854, 20FDA703; 3) REF ACC010381B: UDI/DI 10193489457803 (ea) 401934894578049 (case), Lot Numbers: 20LDA314, 20LDA466, 21ADB162, 21BDB508; 4) REF ACC010381C: UDI/DI 10193489922127 (ea) 40193489922128 (case), Lot Numbers: 21IDA608, 21JDB271, 21KDA979, 21LDA450, 22BDA818, 22FDB822, 22KDB696, 23ADA938, 23CDA005, 23CDB015, 23EDA977, 23FDB762, 23IDB196, 23LDA440, 23LDB579, 24CDA637, 24GDB084, 24IDB423, 24JDB726, 24KDB730; 5) REF ACC010395: UDI/DI 10193489218152 (ea) 40193489218153 (case), Lot Numbers: 20BDC901; 6) REF ACC010397: UDI/DI 10193489218169 (ea) 40193489218160 (case), Lot Numbers: 20CDA453, 20CDA459, 20EDB330; 7) REF ACC010541A: UDI/DI 10195327036980 (ea) 40195327036981 (case), Lot Numbers: 22ADA311; 8) REF ACC010541B: UDI/DI 10195327153649 (ea) 40195327153640 (case), Lot Numbers: 22EDB187; 9) REF ACC010541C: UDI/DI 10198459155765 (ea) 40198459155766 (case), Lot Numbers: 24JDC035; 10) REF DYNJAA252: UDI/DI 10193489297980 (ea) 40193489297981 (case), Lot Numbers: 20EBC867, 20EBD635, 20FBJ920, 22FBF130, 22IBP405
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. ACC010381CMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010395Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010397Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010371CMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA252Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. ACC010541AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. ACC010541BMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. ACC010541CMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
Other devices under the same FDA submission
321 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Centurion ART690Centurion Medical Products Corporation · General Surgery Tray4 recalls0 reports
- Centurion CVI4505Centurion Medical Products Corporation · General Surgery Tray3 recalls0 reports
- MEDLINE ECVC6355Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE ECVC1045Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ART255Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- Centurion CVI3880Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- MEDLINE ECVC1375Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE DYNDV2520Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ST11670ACenturion Medical Products Corporation2 recalls0 reports
- Medline Industries, Inc. ECVC6370Centurion Medical Products Corporation · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC6840Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE ECVC6265Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page