Blueprint Mixed Reality recallZ-1540-2025 · Tornier S.A.S. · initiated Mar 5, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Tornier S.A.S. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its intended use. The most commonly reported complaints were related to impaired software functioning, impaired instrumentation functioning, and general difficulties using the product.
FDA root cause category: Device Design
What the company is doing
Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 03/12/2025 via email. The notice explained the issue and requested the following: "1. Immediately check your internal inventory to locate the product listed on the attached business reply form, remove them from their point of use, and isolate/quarantine the unit(s) to prevent accidental use. 2. Return the enclosed business reply form by email to confirm receipt of this notification/document product segregation. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers." For questions or concerns, please contact [email protected]. On 03/27/2025, Stryker issued an updated notice to include and additional product via email.
Recall details
- Recall number
- Z-1540-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 5, 2025
- Posted by the FDA
- Apr 8, 2025
- Recalling firm
- Tornier S.A.S.
- Quantity
- 474 units
- Product code
- OLO
- FDA submissions
- K222510
- Distribution
- US: TX, MN, AK, PA, MA, TN, ID, and France, Canada, and Ireland
Product description
stryker Blueprint Mixed Reality Depth Stop Pin, Catalog Number MRUE206; Total Shoulder Arthroplasty
Code information (models, lots, serial numbers)
All lots manufactured through Oct 2024, Distributed 23-Aug-2023 through 31-Dec-2024, GTIN 03700434022988
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Blueprint Mixed Reality MRUE206Tornier · Orthopedic stereotaxic instrument1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96437).
- Z-1534-2025 · Open, Classifiedstryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide PPSU + Stainless Steel, Catalog Number MRUE200, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number …
- Z-1535-2025 · Open, Classifiedstryker Blueprint Mixed Reality Glenoid Digitizer, Catalog Number MRUE201, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthro…
- Z-1536-2025 · Open, Classifiedstryker Blueprint Mixed Reality Coracoid Clamp, Catalog Number MRUE202, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthropla…
- Z-1537-2025 · Open, Classifiedstryker Blueprint Mixed Reality Instrument Check Block, Catalog Number MRUE203, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder A…
- Z-1538-2025 · Open, Classifiedstryker Blueprint Mixed Reality Box Base, Catalog Number MRUE204, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthroplasty
- Z-1539-2025 · Open, Classifiedstryker Blueprint Mixed Reality Glenoid Box Lid, Catalog Number MRUE205, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthropl…
- Z-1542-2025 · Open, Classifiedstryker HoloBlueprint Application, Catalog Number MRUE001; Total Shoulder Arthroplasty
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page