AutoPulse NXT Resuscitation System recallZ-1533-2025 · ZOLL Circulation, Inc. · initiated Mar 4, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ZOLL Circulation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
AutoPulse NXT Resuscitation System may not provide adequate chest compressions or may stop compressions due to code FC1060.
FDA root cause category: Process change control
What the company is doing
On March 4, 2025, Zoll Circulation issued a "Medical Device Product Correction Notice" to affected consignees. Zoll asked consignees to take the following actions: 1. Stop using your affected platform(s) immediately. Your patients may not receive high-quality compressions. 2.Complete the Customer Acknowledgment/ No-Charge AP NXT Platform Repair Request Form (page 3)and email it to [email protected] within 3 days. Please feel free to call me at (408) 419 - 2955 if you have any questions. 3.Return your affected platform(s) to ZOLL as soon as you receive the loaner AP NXT Platform(s) from ZOLL.
Recall details
- Recall number
- Z-1533-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 4, 2025
- Posted by the FDA
- Apr 11, 2025
- Recalling firm
- ZOLL Circulation, Inc.
- Quantity
- 91
- Product code
- DRM
- FDA submissions
- K221700
- Distribution
- US: AL, CA, CT, FL, IL, KS, LA, MN, MO, MT, NC, ND, NM, NY, OH, OK< PA, TX, WI OUS: Canada, Hong Kong,
Product description
Brand Name: AutoPulse NXT Resuscitation System Product Name: AutoPulse NXT Platform Model/Catalog Number: Model 200 (Catalog Number: 8700-001070-1) Software Version: Version 2.0.1 Product Description: The AutoPulse Resuscitation System Model 200 is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. Component: N/A
Code information (models, lots, serial numbers)
UDI: (01) 00849111003384/Serial number: 1179 1168 1178 1184 1188 1197 1109 1110 1126 1166 1169 1152 1199 1202 1111 1185 1198 1203 1206 1208 1211 1212 1172 1182 1130 1128 1129 1189 1183 1131 1134 1191 1186 1192 1170 1173 1175 1180 1174 1176 1187 1177 1116 1165 1118 1119 1193 1138 1139 1145 1146 1147 1148 1149 1150 1153 1155 1159 1161 1162 1163 1164 1181 1141 1108 1190 1195 1120 1121 1122 1123 1107 1112 1113 1114 1115 1137 1140 1196
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AutoPulse NXT Resuscitation System Model 200Zoll Circulation, Inc. · Compressor, Cardiac, External1 recall130 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page