EsophyX Z recallZ-1640-2025 · Merit Medical Systems, Inc. · initiated Mar 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Update to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners.
FDA root cause category: Under Investigation by firm
What the company is doing
On March 14, 2025, Merit Medical issued a "Urgent Medical Device Recall Notice via E-Mail, followed by mail to affected consignees. Merit asked consignees to take the following actions: 1. Read and understand this communication and the updated IFU referenced in the attached poster. 2. There is no need to return product to Merit. 3. Post the attached poster on or near the affected products so users are aware of the updated IFU. 4. Discard the current IFU with the product at point of use. 5. Ensure that applicable personnel within your organization are made aware of this communication. 6. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 7. If you require paper IFUs, please contact [email protected]. 8. Please acknowledging you are aware of the information communicated and have completed the actions required of you. 9. If you require additional product training regarding the IFU updates, please contact your Merit Sales Representative to schedule.
Recall details
- Recall number
- Z-1640-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 14, 2025
- Posted by the FDA
- Apr 21, 2025
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 1655 units
- Product code
- ODE
- FDA submissions
- K240879
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Brazil, Egypt, Israel, Saudi Arabia, Spain, Thailand, Turkey, United Arab Emirates.
Product description
Merit Medical, REF: R2007, EsophyX Z+, (1) Fastener Delivery Device, SterileEO, RxOnly
Code information (models, lots, serial numbers)
All Lots/ UDI: 00810275011089
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EsophyX Z+ R2007Endogastric Solutions, Inc. · Endoscopic Suture/Plication System, Gastroesophageal Reflux Disease (Gerd)1 recall1 death69 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96482).
- Z-1641-2025 · Open, ClassifiedMerit Medical , REF: R2275, EsophyX Z+ with SerosaFuse Implantable Fasteners, (1) Fastener Delivery Device, (2) 7.5mm Cartridges, STERILEEO,RxOnly
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.