Brand Name: Cerene Cryotherapy Device recallZ-1584-2025 · Channel Medsystems, Inc. · initiated Mar 20, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Channel Medsystems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device
FDA root cause category: Under Investigation by firm
What the company is doing
On 03/20/2025, the firm emailed customer an "URGENT MEDICAL DEVICE CORRECTION" Letter informing them that the evaporator gasket inside the handle of the device may not seal properly, which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device. Customers are instructed to: 1. Check your internal inventory to determine if you have any affected devices. 2. Stop use of affected devices. 3. Vent the affected devices by following the attached instructions and then discard the devices according to your facility s procedures. 4. Complete and return the customer acknowledgement form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore, please complete the form even if you no longer have any of the affected devices in your physical inventory. 5. Circulate this letter to all other relevant parties, as needed. For questions, contact Channel Medsystems at [email protected] or at (510)338-9301 Monday thru Friday 8:00 AM to 5:00 PM Pacific Time
Recall details
- Recall number
- Z-1584-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 20, 2025
- Posted by the FDA
- Apr 16, 2025
- Recalling firm
- Channel Medsystems, Inc.
- Quantity
- 281
- Product code
- MNB
- FDA submissions
- P180032
- Distribution
- U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX.
Product description
Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A
Code information (models, lots, serial numbers)
Lot Code: Model No: FGS-7000 UDI-DI: 00850008595035 Lot Numbers: 108638953, 108645585, 108645586, 108645587, 108866674, 108866675, 109493291, 109493292 and 109895874
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Cerene Cryotherapy Device FGS-7000Channel Medsystems, Inc. · Device, Thermal Ablation, Endometrial2 recalls30 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page